Principles and Physics: Imaging Artifacts and Pitfalls

6


Principles and Physics


Imaging Artifacts and Pitfalls



Ultrasound travels as transverse waves in the direction of propagation, with compression and expansion of tissue occurring in the direction of travel. Because both light and ultrasound are wave phenomena, wave physics have a large role in image generation. Understanding wave physics is essential to correctly interpret artifacts that do not represent pathologic structures. Furthermore, it is essential to understand the limitations of ultrasound so the underlying anatomy and pathology are assessed correctly.


Ultrasound imaging consists of transduction, beamforming, and image presentation. To create the best possible image, each of these processes should be optimized by adjusting the transmission frequency, gain, image brightness, and contrast. The transducer should use a transmission frequency appropriate for the sector length. Higher frequencies will result in the best resolution of a given object of interest, but they are attenuated more than lower frequencies, limiting their penetration. Gain may be thought of as a volume control. Adjusting the overall and sector time gain compensation does not affect the magnitude of the transmitted ultrasound but changes the manner of its display. System processing, such as “smoothing,” can average out small structures like small plaques or vegetations. Finally, image brightness and contrast should be optimized for faint structures. Altogether these adjustments are known as optimizing signal-to-noise ratio.



image Reflection and Multipath


The ultrasound beamformer assumes a direct line of travel to and from a target. Specular targets, such as prosthetic valves, can cause ultrasound waves to “veer off.” This mirroring can cause ultrasound energy to be reflected in other directions. In Figure 6-1, there are two targets; the ultrasound beam is directed toward target 1. Most of the energy is reflected from this target to the transducer and is accurately imaged. A portion of the energy is reflected from target 1 to target 2 and returns to the transducer after following a circuitous path. Since “time-of-flight” is equivalent to depth, the longer length of the multipath transmission will be interpreted as a deeper target along the original axis (i.e., in the axis of target 1), resulting in artifact production.




image Refraction


Refraction artifact is similar to a multipath artifact. In a multipath artifact, the echo lines are reflected between two targets. In a refraction artifact, the ultrasound beam is bent by a refractor ( Fig. 6-2). Ultrasound scanlines are all directed in a straight path. During their transit, however, some may be refracted (steered off at an angle). These refracted beams may then be reflected back toward the transducer. The machine measures both the refracted and non-refracted echoes, but assumes that all the returned echoes traveled along a straight path. Thus, two images are seen: the true image of the target formed by the straight path echoes, and a false double image seen in line with the refracted beams.




image Ringing, Rattling, and Reverberation


Structures of interest are usually imaged by assuming an ultrasound pulse is reflected once by an object of interest. The dimensions of the object are generally determined by the duration between the emitted and received signal. Certain targets of high acoustical impedance, however, may trap ultrasound waves and result in a to-and-fro reflection of ultrasound energy within it. This reverberation may be seen as either a mirror artifact or as a linear reverberation (“ring-down”). For example, when one attempts to image a beaker, one beaker wall will be closer to the transducer and one farther away. If ultrasound energy is reflected once from each beaker wall, the object will be accurately imaged, but it is possible for an ultrasound wave to continue to be reflected between the beaker walls, thus delaying return of this ultrasound energy to the transducer. Since the ultrasound transducer has no means of differentiating a singly reflected wave from a wave that has been reflected multiple times, a copy of the image appears in the far field ( Fig. 6-3 and image Video 6-1). Alternatively, ultrasound energy may be trapped within a very thin object. This trapping of ultrasound energy in this small space may result in a continuous return of ultrasound energy to the transducer, which will be displayed as an echogenic line called linear reverberation or “ring-down.” This vibration or rattling of the target causes bright targets to appear but limits interrogation in the far field beyond the reverberating target.

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Jun 12, 2016 | Posted by in CARDIOLOGY | Comments Off on Principles and Physics: Imaging Artifacts and Pitfalls

Full access? Get Clinical Tree

Get Clinical Tree app for offline access