Initial Patient Assessment


Initial Patient Assessment


Learning Objectives


On completion of this chapter, the reader will be able to do the following:



Understand the importance of performing an operational verification procedure.


State the recommended times when an oxygen analyzer is used to measure the fractional inspired oxygen concentration (FIO2) during mechanical ventilation.


Identify various pathophysiologic conditions that alter a patient’s transairway pressure, peak pressure, and plateau pressure.


Use pressure-time and flow-time curves obtained during pressure-controlled continuous mandatory ventilation (PC-CMV) to determine the plateau pressure.


Identify a system leak from a volume-time curve.


Use physical examination and radiographic data to determine whether pneumonia, acute respiratory distress syndrome (ARDS), flail chest, pneumothorax, asthma, pleural effusion, or emphysema is present.


Determine whether a lung compliance problem or an airway resistance problem is present using the ventilator flow sheet and time, volume, peak inspiratory pressure (PIP), and plateau pressure data.


Evaluate a static pressure–volume curve for static compliance and dynamic compliance to determine changes in compliance or resistance.


Estimate a patient’s alveolar ventilation based on ideal body weight, tidal volume, and respiratory rate.


10 Detect a cuff leak by listening to breath sounds.


11 Recognize inappropriate endotracheal tube cuff pressures and an inappropriate tube size and recommend measures to correct these problems.


12 Evaluate flow sheet information about a patient on pressure control ventilation and recommend methods for determining whether compliance and airway resistance have changed.


13 Explain the technique for measuring endotracheal tube cuff pressure using a manometer, syringe, and three-way stopcock.


14 Describe two methods that can be used to remedy a cut pilot tube (pilot balloon line) without changing the endotracheal tube.


Key Terms


• Ascites


• Inflection point


• Operational verification procedure


• Patient-ventilator system check


• Upper inflection point


• Ventilator flow sheet


A patient’s color, respiratory rate, breathing pattern, use of accessory muscles, chest movement, breath sounds, work of breathing (WOB), and level of consciousness can provide valuable information about the patient’s physiological status. These observations, along with information derived from ventilator displays and hemodynamic monitoring, are among the first assessments the clinician records for a patient who is undergoing mechanically ventilation.1


This chapter reviews assessment and documentation of patient-ventilator interactions after a patient has been placed on a mechanical ventilator. The first step in this process involves the respiratory therapist verifying the physician’s orders (Box 8-1).2 Once the physician’s orders have been verified, the respiratory therapist ensures that the designated ventilator has passed an operational verification procedure (OVP). The OVP process usually is described in the respiratory therapy department’s policies and procedures manual.



Box 8-1


Physician’s Orders for Mechanical Ventilation


The orders written by the physician for mechanical ventilation settings vary between institutions. In some cases the physician’s orders may be very specific, with little flexibility for respiratory therapist involvement. More commonly, the physician orders simply request a particular protocol for mechanical ventilation, which is then followed by the respiratory therapist, nurse, and other staff members involved with the care of the patient.2


The orders or protocol should include at least one (and preferably both) of the following:



It is important that the respiratory therapy department maintains records showing the OVP history for each ventilator. In addition, a label or form should be attached to each ventilator showing when the OVP was performed, by whom, and whether the ventilator passed the multiple-part test. Newer microprocessor-controlled ventilators perform a series of automated self-tests when the operator initiates the self-test process. This self-test record may be part of the OVP.


The equipment evaluation process should also involve checking the integrity of the ventilator circuit and humidifier system and ensuring that related equipment has been correctly attached and tested to make sure the system is free of leaks (Box 8-2).2



Documentation of the Patient-Ventilator System


In addition to documentation of the OVP, patient information and ventilator settings should be documented regularly when a patient is receiving ventilatory support. These data can be recorded on a computer software program with specific entry fields or kept as a paper record. Regardless of the form it takes, the document often is called a ventilator flow sheet.


The frequency of patient-ventilator system checks depends on the institution’s policy. They generally are performed every 1 to 4 hours.3 In addition to this schedule, patient-ventilator system checks are performed:



It is important to recognize that these patient-ventilator system checks represent a documented evaluation of the ventilator’s function and the patient’s response to ventilatory support. The American Association for Respiratory Care (AARC) has developed clinical practice guidelines for recording patient-ventilator system checks.2 Several points regarding these guidelines should be emphasized:



Figure 8-1 shows the information that may be included on a ventilator flow sheet.2 The top of the form contains basic patient information:




Spaces are also provided for current information and measurements of patient and ventilator parameters. These may include the following:



• Date


• Time


• Mode of ventilation


• Minute ventilation (image)


• Respiratory rate (f)


• Tidal volume (VT)


• Peak inspiratory pressure (PIP)


• Plateau pressure (Pplateau)


• Static compliance (Cs)


• Airway resistance (Raw)


• Fractional inspired oxygen (FIO2)


• Temperatures of inspired gases


• Inspiratory-to-expiratory (I : E) ratio


• Continuous positive airway pressure (CPAP) or positive end-expiratory pressure (PEEP)


• Inspiratory and end-expiratory positive airway pressures (IPAP/EPAP)


• Arterial blood gases (ABGs)


• Alveolar-to-arterial partial pressure of oxygen (P[A-a]O2) or ratio of arterial oxygen partial pressure to fractional inspired oxygen (PaO2/FIO2)


• Vital capacity (VC)


• Maximum inspiratory pressure (MIP)


• Vital signs


• Alarm settings


Volumes, pressures, temperature, vital signs, and FIO2 are measured during each patient-ventilator system check. Please note that although FIO2 may be measured intermittently for adults, it should be continuously monitored for infants receiving mechanical ventilation.2 Most of the current intensive care unit (ICU) ventilators have oxygen analyzers that provide continuous monitoring of FIO2. Alarms should be regularly checked to ensure that they have been set appropriately. ABGs, shunt, P(A-a)O2, and PaO2/FIO2 are determined when the patient’s condition or the ventilator settings change significantly.


Despite the importance of regular, accurate documentation, many respiratory therapy departments do not follow the AARC’s clinical practice guideline for patient-ventilator system checks or a similar model.3,4 Case Study 8-1 provides a clinical scenario illustrating the important of maintaining accurate patient-ventilator records.5



image Case Study 8-1


The Importance of Documentation


A 38-year-old woman is intubated for respiratory failure secondary to severe pneumonia. After 24 hours her status improves. The endotracheal tube is kept in place to allow suctioning, because she has large amounts of secretions. On the third day after intubation, her respiratory status declines. She has a cardiac arrest and is resuscitated but suffers brain injury as a result. She dies several weeks later.


The family hires an attorney. At issue is the fact that in the medical record, the respiratory therapist’s notes with the ventilator flow sheet indicate that the patient had been suctioned about every 2 hours for large amounts of thick, yellow secretions. However, during the 8 hours before the arrest, nowhere did the notes state that the patient had been suctioned. The therapist states that he had suctioned the patient but had not recorded it in the chart. Was the therapist negligent and did his actions lead to the wrongful death of the patient?


See Appendix A for the answer.


The First 30 Minutes


Immediately after the patient is connected to a mechanical ventilator, the clinician should perform auscultation of the patient’s chest to confirm adequate volume delivery and proper placement of the ET. The patient’s vital signs are checked, making particular note of heart rate and blood pressure, because ventilation may affect these parameters (Key Point 8-1). The alarms (e.g., apnea, low pressure, low VT, and high pressure limit) are activated.6 An arterial blood sample is obtained about 15 minutes after mechanical ventilation is initiated for evaluation of the effectiveness of ventilation and oxygenation.3,7 If not already done, a chest radiograph is obtained to confirm proper placement of the ET. Box 8-3 lists other clinical laboratory tests a physician might order to assess the patient’s status when mechanical ventilation is initiated.3



image Key Point 8-1


Positive-pressure ventilation can reduce venous return to the heart, cardiac output, and blood pressure. (See Chapter 16 for more information on the cardiovascular effects of mechanical ventilation.)



Once the patient assessment shows the individual is stable, the respiratory therapist then performs the first ventilator check.


Mode


The mode of ventilation is recorded in the appropriate space on the ventilator flow sheet. It may be recorded as follows:



Sensitivity


If a patient-triggered mode is used (i.e., inspiration is initiated by patient effort), the pressure or flow required to trigger the ventilation should be checked; no more than −1 or −2 cm H2O should be required. If the ventilator is flow triggered, the sensitivity should be set so that the ventilator will trigger at a flow change of 2 to 3 L/min. The ventilator should be checked for auto-triggering, and the patient’s ability to trigger a breath should be assessed. Adjustments are made as needed.


As discussed previously, when auto-PEEP is present, the patient has more difficulty triggering breaths. The presence of auto-PEEP should be suspected if the patient is using his accessory muscles of inspiration or demonstrates labored breathing. Ventilator graphics that show failure of the expiratory flow to return to zero before the next breath also is an indicator of auto-PEEP (see Chapter 9).8


If auto-PEEP is present, a number of strategies can be used to reduce its effects, including increasing the flow (reducing the inspiratory time [TI]), reducing VT, or reducing the rate (i.e., reducing image), suctioning the patient, or changing modes to allow for more spontaneous breaths. It is important to recognize that it might not always be possible to eliminate auto-PEEP, particularly in patients with increased flow resistance and airway closure. The addition of extrinsic PEEP (PEEPE) in these patients may make triggering easier (see Fig. 7-1). PEEPE is increased progressively during VC-CMV until the patient’s use of accessory muscles diminishes or until PIP and Pplateau begin to rise. (Additional information about treatment of auto-PEEP can be found in Chapter 17.)


Tidal Volume, Rate, and Minute Ventilation


Usually, VT, f, and image are displayed digitally on the front panel of the ventilator in the data display window. Most ventilators display the set VT (VTset) and the exhaled VT (VTexh). Newer microprocessor-controlled ventilators provide excellent, reliable flow and pressure monitoring. If this information is not available, VT can be measured using a handheld bedside pulmonary function device or other volume-measuring device (e.g., respirometer) and a watch or clock with a sweep second hand. Although this technique generally is used only with older ventilators, it can also be used to verify digital readouts if a question arises about the machine’s reliability (Fig. 8-2 and Box 8-4).




Correcting Tubing Compliance


Accurate reporting of volumes requires correction for volume loss within the patient circuit due to the effects of tubing compliance (also called compressible volume). Tubing compliance (CT) for most ventilator circuits ranges from about 1.5 to 2.5 mL/cm H2O.9 Box 8-5 presents a sample calculation of volume loss associated with CT (also see Chapter 6).



Box 8-5


Calculation of Compressible Volume (Volume Lost to Tubing Compliance)


If the positive inspiratory pressure (PIP) is 25 cm H2O, the tidal volume (VT) at the exhalation port is 600 mL during performance of the CT calculation, and tubing compliance (CT) is 3 mL/cm H2O. The volume lost to the tube is calculated as follows:


PIP×CT


image

25cm H2O×3mL/cm H2O=75mL


image

The volume actually delivered to the patient’s lungs is found by subtracting 75 mL from the VT measured at the exhalation port:


60075=525mL


image

Most microprocessor-controlled ventilators (e.g., Puritan Bennett 840 [Covidien-Nellcor and Puritan Bennett, Boulder, Colo.] Servoi [Maquet Inc, Wayne, N.J.]) do not require calculation of tubing compliance because these machines automatically compensate for this factor. As discussed previously, ventilators that perform this function typically increase the delivered volume so that the set volume is the amount provided to the patient (i.e., the digital readout indicates the VT actually delivered to the patient’s airway). With older ventilators, the operator can set the value for the circuit’s CT and the ventilator will add volume to the set VT to compensate for volume loss due to tubing compliance.


Alveolar Ventilation


Monitoring of alveolar ventilation (image) has declined in popularity in recent years because many acute care facilities do not include this variable on the ventilator flow sheet. Unfortunately, its importance often is overlooked when using low VT strategies. In-line heat and moisture exchangers (HMEs), closed suction systems, and other circuit adapters and equipment can add mechanical dead space (VDmech) to the ventilator circuit and affect the VD/VT. Knowledge of the effect of dead space on alveolar volume delivery can be particularly important in infants, children, and smaller adults with ARDS.


Two factors must be considered in determining alveolar ventilation:



Dead Space


Normal anatomic dead space (VDanat) is about 1 mL/lb IBW.* Bypassing the upper airway with an artificial airway reduces VDanat by about one half. Using a Y-connector, additional flex tubing between the Y-connector and the ET, or a HME adds mechanical dead space.


Added Mechanical Dead Space


Because HMEs or other adapters attached to the ET (Fig. 8-3) add to mechanical dead space (VDmech), the volume of these devices, along with the VDanat, must be subtracted from the VT to determine actual alveolar ventilation (Box 8-6). For example, if a 150-lb adult has a VT of 500 mL and the added VDmech is 100 mL, the alveolar ventilation for each breath would be:


VTVDmechVDanat=500mL100mL150mL=250mL


image

(Key Point 8-2)




Box 8-6


Added Mechanical Dead Space


Some patient situations require the addition of a small amount of mechanical dead space. For example, in Fig. 7-3 a small circuit has been added at the Y-connector to allow for intermittent use of a metered-dose inhaler (MDI) without removing the heat and heat-moisture exchanger (HME) or disconnecting the circuit. This adds VDmech to the ventilator circuit.


Excessive turbulence at the upper airway and Y-connector sometimes can produce a pressure spike at the beginning of a breath during pressure-targeted ventilation. This has been noted with CareFusion (Pulmonetics) LTV-1000 ventilator (CareFusion, Yorba Linda, Calif.) in a clinical situation.12 Auto-triggering occasionally occurs in this same ventilator. These two phenomena can be prevented in most cases by adding about 2 in. of corrugated tubing between the Y-connector and the endotracheal tube (or between the HME and the endotracheal tube) but must be accounted for as additional VDmech.



image Key Point 8-2


The volume of mechanical dead space can be easily measured for any device added to a ventilator circuit. The device is simply filled with water and then emptied into a graduated container; the volume measured is the volume of mechanical dead space for the device. (A piece of equipment [i.e., tubing] similar to that being placed on the patient should be used for this type of measurement.)


Final Alveolar Ventilation


During VC-CMV (A/C), image is calculated by multiplying the number of breaths counted for 1 minute by the corrected VT: image = (VT − VDanat − Added VDmech) × f. When a patient is on VC-SIMV, the mandatory rate and volume delivery must be calculated separately from the patient’s spontaneous rate and volume; these values then are added to determine the total minute ventilation (Box 8-7).



Monitoring Airway Pressures


All positive-pressure ventilators have a pressure monitor or display that continuously shows the upper airway pressure. The pressure monitor is probably most accurate when it shows pressures measured very near the patient’s upper airway because this eliminates the effects of the humidifying device and the breathing circuit on the measurement. Proximal pressure measurements are obtained by connecting the monitor tubing (usually a small-diameter plastic tube) to the Y-connector of the patient circuit. If the tubing is not visible on the circuit, pressures are detected through the main patient circuit and not at the upper airway. Although these values are not measured at the proximal airway, ICU ventilators are generally very accurate and clinically useful1 (Key Point 8-3).



image Key Point 8-3


Proximal pressure and flow monitor lines must be free of moisture and secretions to provide accurate readings.


Pressure monitoring allows the clinician to assess pressure delivery to the upper airway. It ensures that a minimal pressure is maintained (low-pressure limit) and that high-pressure limits are not exceeded during mechanical ventilation. Intermittent readings of PIP, Pplateau, set pressure (for pressure-targeted modes such as PC-CMV and pressure support [PS]), transairway pressure (PTA = PIP − Pplateau), mean airway pressure (image), and end-expiratory pressure (EEP) can provide valuable information about the patient’s condition.


Peak Inspiratory Pressure


PIP (or PPeak), which is the highest pressure observed during inspiration, can be used to calculate dynamic compliance (CD; also called dynamic characteristic). A fairly constant VT with an increasing Ppeak may indicate a reduction in lung compliance (CL) or an increase in airway resistance (Raw). Conversely, a declining PIP may indicate a leak or may be a sign of improvement in compliance or resistance.


Plateau Pressure


A Pplateau reading can be obtained by using the ventilator’s inspiratory pause control or by setting a pause time of about 0.5 to 1.5 seconds. Remember that the static pressure is read when no gas flow is occurring. Traditionally, Pplateau was determined by occluding the ventilator’s expiratory port at the end of inspiration and reading the pressure registered on the ventilator’s pressure manometer. After PIP is reached, the manometer needle or pressure indicator shows a drop of a few centimeters from peak and then briefly remains in a plateau (static) position before dropping to zero (Fig. 8-3). It is important to remember that Pplateau cannot be measured accurately if the patient makes active respiratory efforts, has a high f, or resists extending TI; therefore measurement of Pplateau is difficult, if not impossible, to obtain in spontaneously breathing individuals.


Pplateau most often is used to calculate CS, which reflects the elastic recoil of the alveolar walls and thoracic cage against the volume of air in the patient’s lungs. Current practice strongly recommends keeping Pplateau below 30 cm H2O (Key Point 8-4). Notice that during either pressure or volume ventilation, if a condition of no gas flow occurs near the end of inspiration (i.e., the flow-time waveform graphic shows a reading of zero), the corresponding pressure on the pressure-time curve is an indicator of Pplateau (Fig. 8-4 and Key Point 8-5).



image Key Point 8-4


Current practice strongly recommends keeping Pplateau below 30 cm H2O to avoid ventilator induced lung injury.


Stay updated, free articles. Join our Telegram channel

Aug 7, 2016 | Posted by in RESPIRATORY | Comments Off on Initial Patient Assessment

Full access? Get Clinical Tree

Get Clinical Tree app for offline access