Final Considerations in Ventilator Setup


Final Considerations in Ventilator Setup


Learning Objectives


On completion of this chapter, the reader will be able to do the following:



Recommend fractional inspired oxygen concentration (FiO2) settings when initiating mechanical ventilation.


Discuss the pros and cons of using the sigh function during mechanical ventilation.


Compare the use of sigh with the concept of a recruitment maneuver in acute respiratory distress syndrome.


List the actions necessary for final ventilator setup.


Explain the concept of using extrinsic positive end-expiratory pressure (PEEP) in patients with airflow obstruction and air trapping who have trouble triggering a breath during mechanical ventilation.


Calculate the desired FIO2 setting given the current partial pressure of arterial oxygen (PaO2) and FIO2 values.


List the essential capabilities of an adult intensive care unit (ICU) ventilator.


Provide initial ventilator settings from the guidelines for patient management for any of the following patient conditions: chronic obstructive pulmonary disease (COPD), neuromuscular disorders, acute asthma episodes, closed head injuries, adult respiratory distress syndrome (ARDS), and acute cardiogenic pulmonary edema.


Key Terms


• Barotrauma


• Cushing response


• Humidity deficit


• Isothermic saturation boundary


• Pulsus paradoxus


• Relative humidity


• Status asthmaticus (acute severe asthma)


Several issues must be considered after decisions about the type of ventilator to be used, mode selection, and settings for pressure and volume have been made. These issues include selecting appropriate ventilator settings for the fractional concentration of inspired oxygen (FIO2), sensitivity, sigh breaths, alarms, and monitors, in addition to concerns regarding humidification of inspired gases. Only after these issues have been addressed can mechanical ventilation be initiated.


This chapter provides a summary of these issues and also addresses the initial settings for patients with specific pathological conditions, such as chronic obstructive pulmonary disease (COPD), asthma, neuromuscular diseases, and acute respiratory distress syndrome (ARDS).


Selection of Additional Parameters and Final Ventilator Setup


Selection of Fractional Concentration of Inspired Oxygen


The goal of selecting a specific FIO2 for a patient is to achieve a clinically acceptable arterial oxygen tension (e.g., 60-100 mm Hg). To accomplish this goal, a baseline arterial blood gas (ABG) should be performed. If the patient’s partial pressure of arterial oxygen (PaO2) is within the desired range before beginning ventilatory support, the FIO2 that the patient is receiving at the time of the baseline ABG can be used when mechanical ventilation is initiated. If the PaO2 is not within the desired range, the following equation can be used to estimate FIO2:


Desired FIO2=[PaO2(desired)×FIO2(known)]PaO2(known)


image

This relationship is based on the assumption that the patient’s cardiopulmonary function will not radically change from the time of the baseline ABG to the time when mechanical ventilation is initiated.1,2 Some changes will obviously occur because the application of positive pressure ventilation (PPV) can affect a patient’s cardiopulmonary status.


If a baseline ABG is not available, it is advisable to select a high initial FIO2 setting ≥0.50) for patients with presumed severe hypoxemia. This can provide a way of restoring normal oxygenation and replacing tissue oxygen storage when oxygen debt and lactic acid accumulation have occurred. Many practitioners start with an FIO2 of 1.0 and then reduce it as quickly as possible. Extended use of 100% O2 is not recommended because it can quickly result in absorption atelectasis and, in the long term, can lead to oxygen toxicity. It is important to state, however, that 100% oxygen should not be withheld if the patient is seriously ill and requires a high FIO2. Indeed, any procedure that places the patient at risk of developing hypoxemia should be performed with the patient breathing 100% O2. For example, administering 100% O2 before and after suctioning and also during bronchoscopy is a common practice.


Titrating the FIO2 using pulse oximetry and ABG findings can minimize the risk of administering too much oxygen.35 The FIO2 can be adjusted after ventilation is started, based initially on the pulse oximetry saturation (SpO2).6 An SpO2 greater than 92% (PaO2 ≥ 60 mm Hg) is a common and acceptable goal. Within 10 to 20 minutes of beginning ventilation, an ABG sample should be collected to assess the adequacy of ventilation and oxygenation. Appropriate ventilator changes based on ABG results are reviewed in Chapters 12 and 13.


The equation for obtaining a desired FIO2 shown earlier in this section can also be used to adjust the FIO2. When an FIO2 greater than 0.50 is required to maintain oxygenation, positive end-expiratory pressure (PEEP) may be indicated (see Chapter 13). An FIO2 of 0.50 or greater increases the risk of oxygen toxicity and intrapulmonary shunting that occurs with oxygen induced atelectasis.


Sensitivity Setting


Ventilator sensitivity is normally set so that patients can easily flow- or pressure-trigger a breath (see Chapter 3). Flow triggering is set in a range of 1 to 10 L/min below the base flow, depending on the selected ventilator. Pressure sensitivity is commonly set between −1 and −2 cm H2O.


Many clinicians prefer using flow triggering because it provides a slightly faster response time compared with pressure triggering for two main reasons. First, the exhalation valve does not have to close during flow triggering. With pressure triggering, the circuit has to close and a patient’s inspiratory effort has to drop circuit pressure to the trigger setting. Second, there is a flow of gas in the circuit during exhalation when flow triggering is selected. This flow requires that the inspiratory flow control valve remains open. This provides almost immediate flow on demand for the patient. With pressure triggering, the circuit pressure has to drop before the inspiratory valve opens and flow goes to the patient7 (Key Point 7-1).



image Key Point 7-1


Flow triggering has a slightly faster response time compared with pressure triggering.


If auto-PEEP (intrinsic PEEP, or PEEPI) is present, patients may have trouble triggering a breath. Indeed, not every patient effort will trigger a breath, and when auto-PEEP is very high, patients might not be able to trigger a breath at all. It can therefore be particularly difficult to adjust the ventilator sensitivity so that it senses a patient’s effort when auto-PEEP is present. Furthermore, the cause of the problem often goes unsolved unless auto-PEEP is detected and measured (Box 7-1). When auto-PEEP occurs in mechanically ventilated, spontaneously breathing patients with airflow obstruction (e.g., in COPD), setting the extrinsic PEEP (PEEPE) level to equal about 80% of the patient’s auto-PEEP level may allow the ventilator to sense the patient’s inspiratory efforts.



Box 7-1


Definitions of Positive End-Expiratory Pressure (PEEP)


PEEP = Positive end-expiratory pressure; airway pressure greater than zero at the end of exhalation


Extrinsic PEEP (PEEPE) = the level of PEEP set by the operator on the ventilator


Auto-PEEP (Intrinsic PEEP, or PEEPI) = the amount of pressure in the lungs at the end of exhalation when expiration is incomplete (i.e., expiratory flow is still occurring) and no PEEPE is present (PEEPE is excluded from this value)


Intrinsic PEEP can occur in three situations: (1) strong active expiration, often with normal or even with low lung volumes (e.g., Valsalva maneuver); (2) high minute ventilation (>20 L/min), where expiratory time (TE) is too short to allow exhalation to functional residual capacity; or (3) expiratory flow limitation due to increased airway resistance, as may occur in patients with chronic obstructive pulmonary disease on mechanical ventilation or with small endotracheal tubes or obstructed (clogged) expiratory filters.


Total PEEP = PEEPE + auto-PEEP


Figure 7-1, A and B, helps illustrate this problem. Imagine that you are trying to sip water through a straw from a glass in which the water level is 10 cm below your mouth. You would have to generate at least −10 cm H2O to draw the water up to your mouth. A similar situation occurs in ventilated patients with air trapping who are trying to trigger a breath. The patient must create a pressure gradient between the alveolus and mouth by decreasing alveolar pressure (Palv) to zero or lower so that mouth pressure (PM) is greater than Palv. This gradient allows air to flow into the lungs. For example, if +10 cm H2O of PEEPI is present, the patient would have to generate an effort equal to −10 cm H2O to achieve a Palv of zero. Then the patient must generate an additional −1 to −2 cm H2O to trigger inspiratory flow.



The straw-sipping problem also could be solved by raising the water level closer to the mouth by filling the glass. Similarly, the problem that patients with auto-PEEP have with triggering a breath can be solved by increasing pressure at the mouth (PEEP) until it equals Palv (i.e., the pressure gradient between the mouth and the alveolus is reduced). This reduction is accomplished by applying PEEP with the ventilator (see Fig. 7-1, C). PEEP can be added until most of the airways are no longer collapsed, and then patients only have to generate enough pressure to trigger the ventilator based on the sensitivity setting. Note that this technique will not be effective if the auto-PEEP is a result of a high minute ventilation (VE) and if there is insufficient expiratory time [TE].4


An easy way to estimate the amount of PEEPE to add, if auto-PEEP cannot be measured, is to increase PEEPE until peak inspiratory pressure (PIP) begins to increase. This increase in PIP is an indication that more pressure and volume have been added to the lung. Another technique of estimating the amount of PEEPE to add is to observe whether activity of the accessory muscles of breathing (e.g., sternocleidomastoids) decreases as PEEPE is added (Case Study 7-1). Still another technique involves comparing the number of triggered breaths with the number of patient efforts. As the level of set PEEP is increased, the number of triggered breaths should match the patient’s efforts. Chapters 13 and 17 provide additional information about the complications associated with auto-PEEP, its causes, and methods to reduce auto-PEEP.



image Case Study 7-1


Auto-PEEP and Triggering


A 60-year-old man with COPD is receiving PSV. He appears to be having difficulty triggering the breaths. Auto-PEEP is measured at +8 cm H2O and no PEEPE is being used. Sensitivity is set at −1 cm H2O. How much of an effort (in centimeters H2O) must the patient generate to trigger a breath?


See Appendix A for the answer.


It is important to mention that sensitivity also can be influenced by the type of humidifier system being used. If the humidifier is located between the patient and the point at which the ventilator detects triggering, the patient has to work harder to trigger a breath. When the trigger device is located proximal to the patient’s airway, this is less of a problem.3


Humidification


A spontaneously breathing individual’s inspired air is typically conditioned down to the fourth or fifth generation of subsegmental bronchi (i.e., the isothermic saturation boundary) (Fig. 7-2).8 Under normal circumstances, conditioning of inspired air occurs as air passes through the nose and upper airway. Because these are bypassed during invasive ventilation, a humidity source must be added to the ventilator circuit.



The humidification system used during mechanical ventilation should provide at least 30 mg H2O/L of absolute humidity at a temperature range of about 31° to 35° C for all available flows up to a image of 20 to 30 L/min.9,10 Some clinicians prefer a delivered temperature range of 35° to 37° C.5


Heated Humidifiers

Humidity can be provided by a variety of heated humidification systems. Devices in this category include the following types of humidifiers: passover, vapor phase, wick, and active heat and moisture exchanger.1012 Refilling heated humidifiers is best accomplished by using closed-feed system. With a closed-feed system, the water level in the reservoir is either maintained manually by adding water from a bag through a fill port or by a float-feed system that keeps the water level fairly constant. Notice that the latter system helps maintain a constant water level and even temperature. Both types avoid the need to open the ventilator circuit to refill the device and thus reduce the risk of potential contamination.


Heated humidifiers typically include a servo-controlled heater with a temperature probe that is placed close to the patient airway. These devices are typically equipped with a temperature display and a temperature alarm. The high-temperature alarm is set at 37° to 38° C, so that inspired gas does not exceed 37° C. A minimum alarm setting of 30° C is appropriate.8,9,13


Whenever the temperature in the patient circuit is less than the temperature of the gas leaving the humidifier, condensate accumulates in the circuit. Notice that condensate accumulation (rain-out) will increase as the room temperature becomes cooler.14,15 (Using heated wire circuits on the inspiratory and expiratory lines of the circuit can significantly reduce the amount of rain-out.)


If the temperature of the gas in the patient circuit is higher than the humidifier, the relative humidity in the circuit decreases (Critical Care Concept 7-1).8 This can occur when using heated wire circuits. Anytime a deficit exists between the amount of humidity provided and the amount needed by the patient, drying of secretions can occur. Thus, to assess whether a humidity deficit is present, the therapist should check the patient’s secretions. For example, thick secretions that are hard to suction or the presence of bronchial casts and mucous plugs are signs of drying of the airways (Table 7-1).



image Critical Care Concept 7-1


Changes in Relative Humidity


Gas leaves a heated humidifier at a temperature of 34° C and 100% relative humidity. The absolute humidity is 37 mg/L. The gas enters a heated wire circuit that is heated to 37° C at the proximal airway. What is the absolute humidity of the gas that is 100% saturated at normal body temperature? What is the humidity deficit (i.e., the difference between what is provided by the humidifier and the amount of humidity required by the patient)? What happens to the relative humidity of the gas as it leaves the humidifier and enters the circuit?



Without a heated wire circuit, the humidifier may need to be heated to as much as 50° C for the gas temperature to approximate body temperature (37° C) by the time it reaches the patient’s upper airway. As the highly saturated and warm gas passes through the ventilator circuit, ambient air surrounding the circuit tubing cools this gas and condensate forms in the circuit. Placing water traps at gravity-dependent parts of the circuit to catch excessive rain-out can help alleviate this problem. Water traps should be emptied regularly in a manner that protects the practitioner from any aerosolized spray that may be produced when the trap is opened. Some traps have spring-loaded caps that seal the circuit when they are unscrewed. Others have suction ports from which excess water can be suctioned. Traps that remain sealed during emptying help avoid interruption in ventilation during the process (Key Point 7-2). Maintaining a seal prevents breaking the circuit and thus reduces the risk of introducing contaminants.



image Key Point 7-2


Condensate in the circuit tubing can potentially be a source of accidental lavage when the patient is turned. This water should be directed away from the patient and never allowed to enter the patient’s airway.


Heat-Moisture Exchangers

Heat-moisture exchangers (HMEs), or artificial noses, can also be used for humidification in patients receiving mechanical ventilation. However, there are some circumstances when HMEs should not be used (Box 7-2).9,13,16 HMEs can provide 10 to 14 mg/L of water at tidal volumes (VT) of 500 to 1000 mL. More efficient hygroscopic heat and moisture exchangers (HHME) can provide 22 to 34 mg/L at similar volumes.8 Because a net heat and water loss occurs when HMEs are used for extended periods, the patient should be assessed for signs of drying secretions.



Most HMEs have a resistance to flow of between 2.5 and 3.5 cm H2O/L/min.8 During extended use, HMEs can accumulate moisture and secretions, resulting in an increased resistance to flow. This increased resistance can cause gas trapping (i.e., auto-PEEP) and increase expiratory work of breathing (WOB). If more than four HMEs are used during a 24-hour period because of secretion buildup, it is probably advisable to change to a heated humidifier that provides 100% RH at 31° to 35° C.17


It is also important to recognize that HMEs add mechanical dead space (VDmech) to the ventilator circuit. The dead space for most HMEs ranges from about 50 to 100 mL. This is an important consideration when HMEs are used on patients with low VT, such as infants, children, and adult patients with VT of 400 mL or less (Key Point 7-3).



image Key Point 7-3


Passive humidifiers (heat-moisture exchangers [HMEs]) placed at the endotracheal tube should not be used simultaneously with heated humidifiers. Water produced by the heated humidifier can occlude the filter and significantly reduce airflow to the patient.8


Heat-moisture exchangers should be taken out of line during delivery of an aerosolized medication. It should be kept in mind, however, that circuit disconnection increases the risk of circuit contamination. An alternate approach is to use a meter dose inhaler (MDI) with an MDI adapter placed between the HME and the endotracheal tube (ET). If a spacer is used with the MDI on the inspiratory line, the HME must still be removed. Another solution is to use a circuit adapter that does not require the HME to be removed during aerosol treatments (Fig. 7-3).



Although some manufacturers recommend changing HMEs every 24 hours, replacement may be required only every 2 to 3 days if the HME is not partially obstructed with secretions.3,8,18 Practitioners have reported using HMEs for up to 5 days without difficulties.19 However, if secretions appear thick after two consecutive suctioning procedures, the HME should be removed and the patient switched to a heated humidification system.8 For critically ill patients requiring more than 5 days of ventilation, it is probably better to use a heated humidification system that will optimize humidification and help prevent secretion retention. Long-term use (>7 days) of HMEs for the critically ill patient can increase the rate of ET occlusion. On the other hand, patients in long-term care facilities with tracheostomy tubes in place can use artificial noses for more extended periods of time without difficulty, as long as secretions do not present a problem.19


Alarms


Audible and visible alarm systems are designed to alert the clinician of potential dangers related to the patient-ventilator interaction. This section reviews the most commonly used ventilator alarms and how they are set by most clinicians.20,21 Box 7-3 shows the various levels of alarms and gives some examples of what causes them to become activated.



Low-pressure alarms are usually set about 5 to 10 cm H2O below PIP. These alarms are useful for detecting patient disconnections and leaks in the system. High-pressure alarms are set about 10 cm H2O above PIP and usually end inspiration when activated. High-pressure alarms can be activated when the patient coughs, if secretions increase, compliance drops, or there are kinks in the ET or circuit tubing. Low PEEP/continuous positive airway pressure (CPAP) alarms are usually set about 2 to 5 cm H2O below the PEEP level. Activation of these latter alarms usually indicates the presence of a leak in the patient-ventilator circuit.


Apnea alarms are used to monitor mandatory or spontaneous breaths. An apnea period of 20 seconds is the highest accepted maximum. In some situations, apnea alarms are set so the patient will not miss two consecutive machine breaths (apnea time > total cycle time [TCT] and < [TCT × 2]). Apnea settings provide full ventilatory support for the patient if apnea occurs and should be set appropriately (e.g., VT 5-8 mL/kg ideal body weight [IBW], rate 10 to 20 breaths/min with a high percentage of oxygen [80-100%].)


Most ventilators also have an alarm or indicator that alerts the operator when the inspiratory time (TI) is more than half the set TCT. Some ventilators, such as the Servoi (Maquet Inc. Wayne, N.J.), will automatically end inspiration if the TE is so short that the patient does not have time to exhale. The shortest possible TE is 20% of any cycle time unless the patient is receiving bilevel PAP (bilevel positive airway pressure) and can be activated or inactivated.


Low-source gas alarms alert the operator that the available high-pressure gas source is not functioning. This alarm is critical for newer microprocessor ventilators that rely on high-pressure gas to function, particularly for ventilators that do not have a built-in compressor (Key Point 7-4)



image Key Point 7-4


Low-source gas alarms cannot be silenced if gas is critical to ventilator operation.


Most ventilators also include alarms for low VT, low and high image, low and high respiratory rates (f), and low and high oxygen FIO2. Alarms should not be set so sensitively that they are constantly being triggered. The following suggestions can be used as a guide:



Other alarms are available for detecting low battery levels, if the ventilator is inoperative, ventilator circuit malfunction, exhalation valve leaks, and inappropriately set parameters. For example, a set parameter (e.g., VT) may be outside the range of the ventilator.


Unfortunately, because there are so many alarms and warning indicators on ICU equipment, many clinicians can become desensitized to audible alarms causing the clinicians to respond slowly, or not at all to these alerts.


Action During Ventilator Alarm Situations


When a ventilator malfunctions during use, the clinician must first ensure that the patient is being ventilated. When in doubt, the practitioner should disconnect the patient from the ventilator, start manual ventilation using a manual resuscitation bag, silence the alarms, and call for help. If the practitioner cannot immediately correct the problem, it may be necessary to replace the ventilator (Box 7-4). The operating manuals provided with ventilators usually have troubleshooting sections to solve most problems and can be consulted when time permits. If a ventilator problem cannot be resolved by the in-house biomedical technician support team, it will be necessary to call the local maintenance representative for the company.



Box 7-4


Alarm Failure?


An intensive care unit (ICU) patient receiving mechanical ventilatory support is on an air-filled mattress, and there is a fan in his room to help cool him. He has a pleural drainage system with suction in place.


A nurse at the ICU station hears the ECG monitor alarm, which shows a pattern of asystole. She goes to the patient’s bedside, begins cardiopulmonary resuscitation, and a normal sinus rhythm is quickly restored.


The nurse notes that the patient had been disconnected from the ventilator and attributed the life-threatening event to this occurrence. The nurse also notes that when the ventilator alarm was sounding, it could barely be heard. When confronted with the nurse’s concerns about alarm failure, the respiratory therapist notices that the alarm’s volume adjustment is on the lowest setting and resets it to a higher volume.


This scenario occurs all too frequently in the clinical setting and represents a critical medical error that can be avoided. What would you suggest to prevent a recurrence of this situation?


Periodic Hyperinflation or Sighing


A sigh is a deep breath that occurs regularly as part of a normal breathing pattern. It is used occasionally during mechanical ventilation and related maneuvers (e.g., deep breaths or sighs are used before and after suctioning a patient (Box 7-5).2230



Box 7-5


History of Sighs and Ventilation


Bendixen and colleagues demonstrated that anesthetized and intubated surgical patients developed increased intrapulmonary shunting, decreased PaO2 values, and reduced compliance following mechanical ventilation. They attributed these findings to microatelectasis from constant low tidal volumes.22 When patients were given periodic deep breaths (sighs), these changes were reversed. Unfortunately, the effectiveness of periodic hyperinflation (sighing) continues to be debated because subsequent studies did not entirely support Bendixen’s findings.2329 The decrease in lung compliance (CL) and in PaO2 values seen in surgical patients may actually be due to a loss of functional residual capacity in the supine position and to the effects of anesthetics, muscle relaxants, and similar medications on diaphragm and intercostal muscles function. The decrease in CL and PaO2 can often be improved by the addition of low levels of PEEP.29,30


The sigh or deep breath was a popular idea during the 1960s. Ventilators developed in the 1970s and 1980s incorporated sigh breaths into their designs, although traditional sigh breaths had not been shown to be clinically beneficial. These ventilators were capable of providing one or more deep breaths at periodic timed intervals (i.e., three or four times per hour or once every 10 minutes), depending on the ventilator. Because a normal sigh in a spontaneously breathing, nonintubated person occurs about every 6 minutes, ventilator manufacturers designed their machines to deliver sighs at a similar frequency.28 Sigh volumes were set at one and a half to two times the regular low VT setting.28 (Interestingly, low VT settings [e.g., 5-7 mL/kg IBW] were popular at the time.)


Other investigators found that large VT (10-15 mL/kg) in anesthetized patients reduced atelectasis.31,32 As already discussed, using these higher volumes for patients with acute respiratory failure can cause alveolar overdistention and increase the risk of ventilator-induced lung injury. Mechanical ventilator sigh breaths are therefore not recommended with higher VT rates (VT > 7 mL/kg IBW) or in the presence of plateau pressures greater than 30 cm H2O.


Mild hypoxemia sometimes occurs in patients receiving pressure-supported ventilation (PSV) with low volumes (4-6 mL/kg). Studies of the use of sigh breaths in these patients may be worth examining.29 However, sigh breaths are not indicated for these patients and may be harmful to spontaneously breathing patients receiving CPAP for the treatment of hypoxemia.33


With the advent of low VT strategy in patients with ARDS, another ventilator strategy called lung recruitment has been successfully used in selected patients with ARDS.34,35 The recruitment maneuver is not unlike sigh breaths. The recruitment maneuver, which is used to expand collapsed areas of the lung, involves using a sustained high pressure of 35 to 45 cm H2O for 40 to 60 seconds.3537 Interestingly, the sigh breaths used by Bendixen and colleagues in 1963,22 more than 40 years ago, were as follows:



These sustained high-pressure maneuvers are not unlike the recruitment maneuvers that are used in the management of patients with ARDS (Key Point 7-5 and Box 7-629).



image Key Point 7-5


Sighs are probably not necessary when using tidal volumes greater than 7 mL/kg ideal body weight. Low levels of PEEP (3-5 cm H2O) can also serve to reduce atelectasis formation. When low tidal volumes are used, such as in patients with acute lung injury or ARDS, a recruitment maneuver may be an effective method to avoid atelectasis.



Sighs or deep breaths may be appropriate in the following situations:



Final Considerations in Ventilator Equipment Setup


Before initiating mechanical ventilation, the respiratory therapist should perform a final check of the equipment to be used. This check should include the following steps:



Check ventilator and circuit function to ensure they are operating correctly and no significant leaks are present.


Fill the humidifier with sterile water, and set the humidifier temperature so that the final gas temperature at the airway will be approximately 31° to 35° C, or place an HME in line.


Place a temperature monitoring device near the patient connector when heated humidification is used.


Check the FIO2, set VT (or inspiratory pressure) and f.


Adjust the alarms.


Ensure that the patient is connected to an electrocardiographic monitor.


Have an emergency airway tray available in case the patient’s airway is removed or damaged.


Check that suctioning equipment is available and functioning.


Select a volume-monitoring device and an oxygen analyzer if one is not available with the ventilator.


10 Ensure that a manual resuscitation bag is available and easily accessible.

Stay updated, free articles. Join our Telegram channel

Aug 7, 2016 | Posted by in RESPIRATORY | Comments Off on Final Considerations in Ventilator Setup

Full access? Get Clinical Tree

Get Clinical Tree app for offline access