Single-Photon Emission Computed Tomography Artifacts

Chapter 5 Single-Photon Emission Computed Tomography Artifacts





TECHNICAL ARTIFACTS



Flood Field Nonuniformity


Flood field nonuniformity will result in “ring” artifacts in reconstructed SPECT images. These relatively photon-deficient rings may be apparent in tomographic slices and in severe cases may also appear in polar coordinate maps (Fig. 5-1). Because patients are frequently positioned differently within the camera field for rest and stress scans, flood field artifacts may occur in differents of the myocardium, mimicking reversible or partially reversible defects. Therefore, routine acquisition and inspection of intrinsic and extrinsic flood fields acquired according to vendor recommendations are absolutely critical to avoid such artifacts. Flood fields are routinely acquired the first thing in the morning each working day. However, if ring artifacts appear in clinical SPECT images, it may be necessary to reacquire flood field images in the middle of the day.




Center of Rotation and Camera-Held Alignment Errors


If the camera center of rotation (COR) is incorrect, filtered backprojection during SPECT reconstruction will result in image misregistration and apparent misalignment of the myocardial walls (Fig. 5-2). Technically, this is similar to the error created by cardiac motion. However, unlike motion artifacts, those due to COR error are usually more systematic and predictable. The severity of the apparent defect is directly proportional to the magnitude of the COR error. An error similar to that produced by the wrong center of rotation is produced when the detector is not aligned perpendicular to the radius of rotation.



To avoid these artifacts, camera COR checks should be performed according to the camera manufacturer’s guidelines. The resultant plots should be inspected regularly by a knowledgeable technologist and physician before the camera is used for patient studies. With the observation of a myocardial defect suspected of representing such an artifact, the COR determination should be repeated immediately. Likewise for variable-angle multiheaded detector systems, camera head alignment should be assessed periodically, particularly after camera repair and whenever the detectors are impacted by objects such as stretchers.



Errors in Selecting Oblique Cardiac Axes and Subsequent Polar Map Reconstruction


If the long axis of the left ventricle (LV) is defined incorrectly on either the transaxial or midventricular vertical long-axis slice, the geometry of the heart in subsequently reconstructed orthogonal tomographic slices can be distorted. Consequently, the apparent regional count density can be altered in polar maps, resulting in apparent perfusion defects. These may be accentuated by quantitative analyses in which patient data are compared to normal files (Fig. 5-3). In polar map reconstruction, such errors most often occur in basal myocardial regions at the periphery of the bull’s-eye plot, owing to foreshortening of one of the myocardial walls (Fig. 5-4). Also, the apex, which often demonstrates physiologic thinning and decreased count density, is displaced from the exact center of the polar plot. The displaced apex may also consequently result in an artifact. Misregistration of perfusion defects in short-axis slices and polar maps is not infrequently encountered in patients with sizable severe infarcts.






Arrhythmias and Gating Errors


If arrhythmia is present or if the heart rate changes during SPECT acquisition, inordinately short or long cardiac cycles will be rejected during an 8- or 16-frame/gated acquisition. If the degree of the regular beat rejection varies during the SPECT acquisition, the number of cardiac cycles acquired for each projection image may vary if each projection image is acquired for the same length of time. Therefore, projection images will vary in count density. When viewed in endless loop cinematic format, the projection images will appear to “flash.” The most serious effect of arrhythmia is a decrease in image count density due to these “discarded” cardiac cycles. However, only with severe arrhythmias associated with atrial fibrillation and frequent premature ventricular contractions are clinically significant perfusion artifacts encountered. Although this problem may be overcome by prolonging SPECT acquisition until a prescribed number of regular cardiac cycles have been acquired, such longer acquisition times increase the probability of patient motion due to discomfort or anxiety. Recently, some manufacturers have established a “9th bin” wherein rejected cardiac cycles are stored temporarily and excluded from reconstruction of the gated tomograms, but added back into the summed, nongated images. Alternately, other manufacturers allow simultaneous acquisition of gated and nongated images so that arrhythmic beat rejection will not result in low count-density ungated images.


If the accepted cardiac cycles vary in length, and because each cycle begins at the R-wave used for gating, relative shorter beats will not fill the entire 8 or 16 “bins” of the cardiac cycle. When all of the acquired cardiac cycles are summed to create an 8- or 16-frame gated image, this will result in data “dropout” in the last few “bins” of the cardiac cycle.


Displays of cardiac image count density from individual planar projection images is now possible using some commercially available software programs (Fig. 5-6). Errors in LV volume and ejection fraction may result from gating errors, but these are beyond the scope of this chapter.1




PATIENT-RELATED ARTIFACTS




Breast Attenuation


Because breasts vary in size, position, configuration, and density, breast attenuation artifacts are extremely variable in appearance. In addressing the characteristics of breast attenuation, it is always necessary to consider the position and configuration of the breasts with the patient in the supine position, because patients are imaged in this position with most commercially available SPECT systems. In women of average body habitus, the left breast overlies the anterolateral wall of the heart. In women with large, pendulous breasts, the breasts lie adjacent to the lateral chest wall and more often result in a lateral attenuation artifact. In some women with very large, pendulous breasts, the attenuation artifact may create apparent inferior or inferolateral perfusion defects. In women with very large breasts, the breast tissue may overlie the entire LV. In this instance, the resulting attenuation artifact may be diffuse and less discrete or may primarily involve the apex. In addition to breast position, the caudal angulation of the heart within the thorax will affect the appearance of the attenuation artifact.23 Thus, in summary, the severity of breast attenuation artifacts is not necessarily directly proportional to breast size or chest circumference but may vary considerably according to the position, configuration, and density of the breast, body habitus of the patient, and orientation of the heart within the thorax. Additionally, when women are imaged upright or in a semi-upright, “reclining” position, the breasts are usually more pendulous than in the supine position. Therefore, apical and inferior breast attenuation artifacts are more frequent when women are imaged in these positions.


In some laboratories, a binder is used to flatten the breasts against the chest wall. Although this technique can be quite useful to decrease the thickness of the breast, there is often a degree of uncertainty about the exact position of the breast beneath the binder. Moreover, precise repositioning of the breast beneath the binder in stress and rest studies is difficult. Therefore, in most laboratories, breast binders are not used. For similar reasons, taping the breast upward in an attempt to avoid its overlying the heart is associated with uncertainty in reproducing the breast position for stress and rest SPECT acquisitions.


The interpreting physician must also be cognizant of factors that may further alter the position, configuration, and density of the breast. A bra positions the breast anteriorly and increases the amount of soft-tissue attenuation by thickening the breast. For this reason, it is recommended that SPECT imaging be performed with the bra off. Breast implants may be denser than normal breast tissue and may accentuate attenuation artifacts. Therefore, women who are candidates for SPECT should always be questioned about a history of breast implants or augmentation. Patients with previous left mastectomies may wear a breast prosthesis. Such prostheses may be quite dense, and women who have them must always be instructed to remove them before myocardial perfusion imaging.


Quantitative analysis of SPECT myocardial perfusion imaging is helpful, particularly to the novice, in differentiating normal, physiologic variations in radiotracer distribution from true perfusion defects. However, gender-matched normal files are derived from an average population, and normal limits are established on the basis of variations in regional count density within this population. However, this type of analysis does not take into account the marked variability in body habitus observed in the patient population referred for myocardial perfusion imaging. Therefore, quantitative analysis, instead of aiding the physician in identifying soft-tissue attenuation artifacts, may actually decrease the specificity of SPECT testing by incorrectly identifying such artifacts as abnormalities.


Attenuation correction for SPECT is now commercially available but is still a topic of intense interest and research. The methods for accomplishing it have used either scatter correction or attenuation correction using a separate radionuclide or x-ray transmission image. This approach has been demonstrated to significantly reduce breast and other attenuation artifacts in phantom and patient studies.718 However, whereas attenuation correction may minimize attenuation artifacts created by overlying breast tissue, breast artifacts are frequently not totally eliminated.


Certain breast attenuation artifacts are unique to quantitative analysis in cardiac SPECT. The most common is an apparent inferior perfusion defect incorrectly identified by quantitative analysis in women who have had a left mastectomy. After removal of the breast, there is little or no anterior soft-tissue attenuation. In such cases, the anterior and inferior myocardial count densities are nearly the same as they are for males. Since the normal female file anticipates that the inferior wall will be more intense than the anterior wall, and in the patient study the inferior wall has an identical or lower intensity than the anterior wall, the inferior wall will be identified as abnormal (Fig. 5-7). This quantitative error can be circumvented by comparison of the patient’s data to the normal male file instead. Similar inferior artifacts have been observed in quantitative plots in women with very small breasts, and it might be argued that data from such patients should be compared to normal male limits. However, small breasts may be quite dense, and the assumption that they do not produce photon attenuation may be incorrect. Therefore, in most laboratories, the normal male file is used routinely only for women who have undergone left mastectomy.



The position of the breast may vary from stress to rest if a woman wears different clothing at the time of the two SPECT acquisitions. Breast position varies considerably with a bra on and off. The position of the breast may also vary depending upon the degree of elevation of the left arm, which is preferably positioned above the head for a SPECT acquisition. A shifting breast attenuation artifact can mimic a reversible perfusion defect (i.e., ischemia). For example, if during stress SPECT acquisition, the breast lies high over the anterior chest wall, an anterior attenuation artifact may result. However, if during the resting acquisition, the breast lies more laterally and inferiorly, the artifact will involve the inferolateral wall of the LV. The resulting attenuation artifacts are therefore different in the stress and rest images, so artifactual defect reversibility as well as artifactual “reverse distribution” may result. In this example, the anterior defect will appear to be reversible, whereas there will appear to be “reverse distribution” in the inferolateral wall (Fig. 5-8).



The astute physician and technologist can recognize breast attenuation artifacts by several means. From the following list, the observer should learn to anticipate artifacts:






5. In reconstructed SPECT images, an area of markedly decreased photon density may be observed anterior or lateral to the heart in patients with very large or dense breasts. Adjacent to this photopenic defect are frequently observed accentuated streak artifacts due to nonisotropic attenuation.3 The position of the breast is best evaluated from transaxial or horizontal long-axis slices. When such a large area of markedly decreased count density is noted, an attenuation artifact that may mimic a myocardial perfusion defect often occurs in the adjacent myocardium.

Gated myocardial perfusion SPECT is helpful in differentiating breast attenuation artifacts from scar, since artifacts will demonstrate normal wall motion and wall thickening, whereas infarcts, if they are transmural and sizable, will be hypokinetic with decreased wall thickening (Fig. 5-9).4,5 However, patients with nontransmural myocardial infarctions might have normal wall motion, depending on the thickness of the infarct zone. For artifacts created by a breast that shifts in position from stress to rest, gating is less helpful to differentiate artifact from ischemia as a cause of the reversible or partially reversible defect.




Lateral Chest-Wall Fat Attenuation


In obese patients, there may be a considerable accumulation of fat in the lateral chest wall. When such a patient lies supine on the imaging palette, the thickness of soft tissue in the lateral chest wall is further accentuated. Although this tissue thickness may be the same as or greater than that of the breast, chest-wall fat is usually more uniformly distributed. Therefore, the resultant attenuation artifact is usually more diffuse, often involving the entire lateral wall of the LV.


Like breast attenuation artifacts, apparent perfusion defects due to lateral chest-wall fat should be anticipated before the actual inspection of tomographic slices, through awareness of the following points:



Stay updated, free articles. Join our Telegram channel

Jun 11, 2016 | Posted by in CARDIOLOGY | Comments Off on Single-Photon Emission Computed Tomography Artifacts

Full access? Get Clinical Tree

Get Clinical Tree app for offline access