Extracting Versus Abandoning Sterile Pacemaker and Defibrillator Leads




Nonfunctional or recalled cardiac implantable electronic device leads can be revised with either lead extraction (LE) or lead capping (LC). Factors that influence this decision and comparative outcomes of these strategies are unclear. We reviewed data from our institution to identify patients who received LE (n = 296) or LC (n = 192) from 2006 to 2012. Patients with infectious indications for lead removal were excluded. We compared unanticipated device-related procedures, defined as cardiac implantable electronic device procedures not for device upgrade or battery depletion, using a proportional hazards model adjusted for differences in baseline characteristics. Secondary outcomes were procedural complications, hospitalizations, and all-cause mortality. Patients who received LE were younger and more likely to have an operator with extraction experience (76% vs 26%, p <0.001). Leads removed by experienced extractors versus nonextractors had longer dwell times (4.2 ± 3.6 vs 0.9 ± 1.1 years, p <0.001). Over a median follow-up of 3.0 (interquartile range = 3.2) years, the adjusted risk of unanticipated device-related procedures was similar for LE versus LC (hazard ratio 1.04, 95% confidence interval 0.62 to 1.75). Complications, hospitalization rates, and mortality rates were also similar between the 2 groups. In conclusion, lead revision strategy is influenced by operator extraction experience and dwell time of leads. In our analysis, we found no difference in outcomes between the 2 strategies.


Cardiac implantable electronic devices (CIEDs) use has substantially increased over the past decade, driven by expanding indications. Because of the high rates of lead failures and recalls, and the aging of the population, clinicians often encounter the need to replace a sterile, nonfunctional, or recalled lead or to upgrade a device with resulting superfluous but functional leads. In these situations, the clinician has to decide between extracting the nonfunctional or superfluous lead or abandoning it and adding a new lead. Anecdotally, the decision is often driven by whether the clinician caring for the patient is an “experienced extractor” or not. The decision between extracting or abandoning the CIED lead is not easy as it requires careful assessment of short- and long-term trade-offs. On the one hand, lead extraction carries an incremental procedural risk of death or other major complications, but on the other hand, it is likely to decrease the long-term risks of vascular complications and the need for future extractions of older leads in the event of infection, for example. This study was therefore designed to identify the factors that influence the decision of lead extraction (LE) versus lead capping (LC) and examine the comparative outcomes of these 2 strategies.


Methods


This study was approved by the Internal Review Board of the University of Pittsburgh. All patients who underwent lead revision procedures at the hospitals of the University of Pittsburgh Medical Center (UPMC) from January 1, 2006, to December 31, 2012, constituted the study cohort. Patients with CIED-related infections were excluded from this analysis.


Demographic and clinical data of the patients were obtained from review of the electronic medical records, including patient age, gender, New York Heart Association class of heart failure, and left ventricular ejection fraction. Co-morbid conditions including coronary artery disease, diabetes mellitus, and renal disease were also included in the data set. The use of antiplatelet and anticoagulation agents in the perioperative period was ascertained from medical orders and pharmacologic database. Preoperative laboratory values including hemoglobin levels, serum creatinine, and the international normalized ratio were recorded. Lead revision surgery was considered the index procedure. Previous device and lead data were obtained from the electronic medical record and included the type of device, the number, chamber, dwell time, manufacturer, and model number of each lead. Details of the index procedure and subsequent CIED-related procedures were abstracted from the operative notes.


Patients were followed to the primary end point of the first unanticipated CIED-related procedure, which was defined as any CIED-related procedure performed after the index procedure for any reason other than battery depletion or device upgrade, including procedures for device infection or malfunction or venous complications. Mortality data were obtained from the electronic medical records, including scanned death certificates and Social Security Death Index records. All-cause 30-day readmissions, all-cause hospitalizations, and cardiac-specific hospitalizations were also recorded through August 2014. Cardiac-specific hospitalization was defined as any hospitalization in which the admitting physician was a member of the Heart and Vascular Institute of UPMC. Major and minor complications were also collected during the follow-up period with definitions adopted from the 2009 Heart Rhythm Society expert consensus statement on lead extraction in patients with CIEDs. Operators were classified as experienced extractors or nonextractors. An experienced extractor was defined as one who has experience performing long-term lead extractions using powered laser sheaths and who is familiar with extraction tools including locking stylets. A total of 16 operators at our institution performed 488 procedures of CIED lead revisions.


Baseline characteristics were presented for the LE and the LC groups as mean ± standard deviation for continuous variables and as rates for dichotomous variables and were compared using the Student’s t and chi-square tests, respectively. A p value <0.05 was considered statistically significant. Kaplan–Meier curves were constructed for overall survival and Nelson–Aalen cumulative hazard curves were constructed for other clinical outcomes, comparing LE and LC groups using the log-rank test. Cox proportional-hazard models were constructed for each clinical outcome to adjust for any unbalanced covariates (p <0.10 at baseline comparison). All statistical analyses were performed on SPSS version 10.1 (Armonk, NY).




Results


Of the original 644 lead revisions performed from January 2006 to December 2012 at UPMC hospitals, 488 met the inclusion criteria of replacement of a sterile malfunctioning or recalled lead or upgrade of device resulting in superfluous leads. Table 1 presents the baseline characteristics of the 488 patients included in this analysis. Patients in the LE group were younger, had a marginally lower INR value, and were more likely to have an operator classified as “experienced extractor” (76% vs 26%, p <0.001). All other baseline characteristics were balanced between the 2 groups.



Table 1

Baseline characteristics
















































































































































Variable Extract Lead
(N= 296)
Abandon
Lead
(N= 192)
P-Value
Age (years) 60±17 67±13 <0.001
Women 35% 31% 0.42
Coronary artery disease 40% 50% 0.07
Diabetes Mellitus 22% 28% 0.13
End-stage renal disease 3% 1% 0.17
New Heart Association Class 2.4±0.7 1.3±0.7 0.26
New Heart Association Class 0.47
I 5% 10%
II 44% 46%
III 50% 43%
IV 1% 1%
Ejection fraction (%) 37±15 38±15 0.58
Serum Creatinine (mg/dL) 1.21±0.98 1.13±0.47 0.31
Hemoglobin (g/dL) 13.1±1.9 13.0±1.6 0.78
International normalized ratio 1.23±0.27 1.30±0.37 0.011
Aspirin 58% 65% 0.12
Clopidogrel 12% 16% 0.27
Warfarin 36% 43% 0.12
New oral anticoagulation agents 1.3% 2.0% 0.35
Type of device implanted 0.74
Pacemaker 21% 20%
Defibrillator 40% 37%
Cardiac resynchronization pacemaker 2% 2%
Cardiac resynchronization defibrillator 37% 41%
Experienced extractor operator 76% 26% <0.001
Use of Laser-powered sheaths 57%
Follow-up duration (years) 2.9±2.1 3.1±1.9 0.30


Six operators were classified as experienced extractors and 10 as nonextractors. The experienced extractors performed 275 of the 488 procedures (56%) in this study. Leads extracted by experienced extractors versus nonextractors had longer dwell times (4.2 ± 3.6 vs 0.9 ± 1.1 years, p <0.001). Leads with dwell times greater than 1 year were removed by an experienced extractors in 80% of cases versus only 12% of cases for nonextractors (p <0.001).


Of the 275 cases performed by experienced extractors, the decision to extract versus cap a lead was influenced by the patient’s age whereby patients who underwent extraction were younger (58 ± 17 years in LE group vs 66 ± 14 years in LC group, p = 0.003). This was not the case for the 213 cases performed by nonextractors (64 ± 16 years in LE group vs 67 ± 13 years in LC group, p = 0.13).


Outcomes during follow-up are detailed in Table 2 . Over a median follow-up of 3.0 (interquartile range = 3.2) years, the risk of unanticipated CIED-related procedures adjusted for patients’ age, operator experience in lead extraction, and presence of coronary artery disease was similar for LE versus LC (hazard ratio 1.04 [95% confidence interval 0.62 to 1.75], p = 0.88, Figure 1 ). Major and minor complications, all-cause and cardiac re-hospitalization rates, and mortality rates ( Table 2 and Figure 2 ) were also similar between the 2 study groups.



Table 2

Patient outcomes

















































Extract
Lead
(N= 296)
Abandon Lead
(N= 192)
P-Value
Major Complications 6% 3% 0.13
Minor Complications 3% 3% 0.63
30-day readmission rate 9% 5% 0.13
All-cause hospitalization rate 49% 50% 0.81
Cardiac-specific hospitalization rate 30% 30% 0.91
Unanticipated device-related procedure 18% 15% 0.34
Death 24% 27% 0.42
Length of stay (days) 1.8±2.4 1.6±1.9 0.17