Circumferential and Longitudinal Ventricular Strain in the Normal Human Fetus




Background


Echocardiography with speckle tracking is a novel, angle-independent technique for assessing global and regional cardiac function. Normal data on longitudinal strain have been reported for human fetuses. The aim of this study was to define circumferential left ventricular (LV) strain in a normal fetal population.


Methods


Singleton fetuses between 17 and 42 weeks in gestational age with no adverse maternal health issues or fetal abnormalities were studied. Regional and averaged cardiac strain were measured using syngo Velocity Vector Imaging software.


Results


Data from 81 fetuses were analyzed (mean gestational age, 29.2 ± 5.7 weeks). Overall, average midventricular circumferential strain was 18.7 ± 3.3%, LV longitudinal strain was 15.2 ± 2.7%, and right ventricular longitudinal strain was 16.0 ± 3.3%, with no correlation with gestational age.


Conclusion


This is the first study to report normal fetal LV circumferential strain. These data may be useful as a reference for assessing fetal cardiac function. The retrospective study design and relatively low frame rates used in this study were important limitations.


Fetal cardiac function can be affected by a variety of conditions and prenatal interventions, including cardiac disease, twin-twin transfusion syndrome, thoracic anomalies, and procedures to treat them. The accurate assessment of fetal cardiac function may help improve the understanding of disease processes and prognosis, as well as the evaluation of prenatal interventions for conditions such as twin-twin transfusion syndrome or fetal cardiac diseases.


Echocardiography with speckle tracking (STE) is a novel, angle-independent technique for assessing global and regional cardiac function. STE makes it possible to assess not only the global condition of the fetal heart chambers but also segmental function, and the angle independency broadens its potential application in the face of varying fetal positioning. Although STE allows the assessment of longitudinal, circumferential, and radial strain, only data on longitudinal function have been reported for fetuses. However, some conditions, such as fetal aortic stenosis (AS) or consequent endocardial fibroelastosis, may have different effects on circumferential and longitudinal left ventricular (LV) contraction, and it may be important to assess cardiac function in different axes.


The aim of this study was to define circumferential strain values for the normal fetal population. In addition, to confirm the reliability of this study, we compared our findings with those of previously published studies. We also compared these values with strain measurement obtained from fetuses with AS to assess the usefulness of circumferential strain among fetuses with possible cardiac dysfunction.


Methods


Study Population


We retrospectively evaluated cardiac ultrasound studies in fetuses referred to Children’s Hospital Boston for fetal echocardiography for a family history of congenital heart disease or suggested fetal arrhythmia from 2004 to 2010. From these, we identified singleton fetuses between 17 and 42 weeks in gestational age (GA), with no maternal disease, fetal structural disease, or arrhythmia. After review of the echocardiographic studies, those with images felt to be sufficiently clear for STE were sequentially selected and analyzed. About 20 studies were selected for each GA subgroup (≤24, 25–29+6/7, 30–34+6/7, ≥35), with the intent of selecting a balanced distribution of fetal GA.


We identified fetuses with AS and moderate or less LV dysfunction. The AS gradient was estimated using the modified Bernoulli equation and LV volumes and LV ejection fraction using the single-plane area-length method. We selected five fetuses with AS to represent different physiologic states. Patients 1 and 2 had moderate LV dysfunction and substantial AS gradients, patients 3 and 4 had normal or mildly depressed LV function and substantial AS gradients, and patient 5 had normal LV function and a mild AS gradient. STE was performed retrospectively in these fetuses and was not used to determine whether to perform fetal aortic valvuloplasty (fAVP) or to determine the timing of fAVP.


The study was performed in conjunction with a protocol that was approved by the Children’s Hospital Committee for Clinical Investigations.


Image Acquisition


Images were acquired using Siemens Sequoia (Siemens Medical Solutions USA, Inc., Mountain View, CA) and Phillips Sonos 5500 and iE33 (Philips Medical Systems, Bothell, WA) machines. Phased-array sector or linear probes ranging from 4.0 to 7.0 Hz were used. The short-axis view was obtained by scanning perpendicular to the long axis of the heart. For circumferential strain, short-axis images containing some part of both papillary muscles without any mitral valve tissue through the whole cardiac cycle were selected as mid–papillary muscle–level images; for longitudinal strain, four-chamber views were selected. All the images were stored in Digital Imaging and Communications in Medicine format and analyzed offline. Fetal echocardiographic images were automatically stored at a frame rate of 30 Hz.


Offline Analysis


Regional cardiac function was measured using syngo Velocity Vector Imaging (VVI) software (Siemens Medical Solutions, USA, Inc.). Among images stored as Digital Imaging and Communications in Medicine clips in routine screening, well-visualized four-chamber and short-axis views were selected. Images that were acquired by Phillips Sonos 5500 or iE33 machines were reformatted to be compatible with syngo VVI software.


To define a fetal cardiac cycle, using superimposed M-mode tracings, end-diastole was defined as the point just before closure of the atrioventricular valve. For short axis, the longest LV dimension was used to define end-diastole. After manually tracing the subendocardial surface of the ventricle on the frame, VVI software calculated strain automatically using the feature-tracking algorithm ( Figure 1 ). Feature-tracking accuracy was visually confirmed, and the tracing was corrected until consistent endocardial tracking was obtained. Only one cardiac cycle was used for analysis instead of averaging over multiple cardiac cycles, to avoid artifacts due to fetal movement or change of cardiac cycle. Analysis was performed by one examiner, who was blinded to outcomes and other echocardiographic parameters.




Figure 1


A still frame of the short axis of the left ventricle with velocity vectors. The arrow represents magnitude of the tissue velocity and the direction of movement of the myocardium.


Because of the variable fetal cardiac location, some rearrangement was needed to match the actual segment of the fetal heart and the segment displayed on the screen. The segment consisting primarily of the septum and closest to the tricuspid valve ( Figure 2 ) was defined as septal, and other segments were defined in order.




Figure 2


These short-axis echocardiographic images demonstrate the method of segmenting the left ventricle (LV) into six segments. The syngo software calculates the center of the gravity of the tracing circle automatically. Segment 1 is defined by the software as the 60° region centered on a vertical line drawn upward from the center of the gravity (30° to each side of the line). Moving clockwise around the wall of the LV, successive 60° segments are defined as segments 2 to 6. In the case depicted in the figure, the segment consisting primarily of the septum and closest to the tricuspid valve was segment 1. The liver was on the right, the stomach (S) was below the heart, and the head was to the left of the image. Therefore the six segments were defined as follows: anterior, 5; lateral, 4; posterior, 3; inferior, 2; septal, 1; and anteroseptal, 6. P1 , Posteromedial papillary muscle; P2 , anterolateral papillary muscle.


Longitudinal and circumferential negative peak strains were used for strain values. Longitudinal strains for all segments were averaged to express global strain.


Data Analysis


To assess intraobserver and interobserver variability, a randomly selected subset of 15 data sets, analyzed by two examiners who were blinded to each other’s measurements, was used. Bland-Altman analysis bias was calculated and presented with the standard deviation. To assess the success rate of analysis, χ 2 tests were conducted, and significance was set at P ≤ .05. To assess relationships between GA and speckle-tracking echocardiographic data, linear regression was performed. Data are presented as mean ± SD.




Results


General Characteristics


Indications for fetal echocardiograms in the normal cohort included the following: sibling of child with congenital heart disease, possible fetal arrhythmias detected by primary obstetrician, and maternal congenital heart disease. From 150 sequential fetal studies, mothers with congenital heart disease (except for one with a small atrial septal defect), significant fetal arrhythmias, and fetuses with any anomalies were excluded. Because a majority of fetuses were evaluated in the middle of the second trimester, we selected 81 of those 150 fetuses with the intent of selecting a balanced distribution of fetal GA that met demographic and diagnostic inclusion criteria for analysis. At the time of the examination, there were no identified obstetric abnormalities. GA at the time of echocardiogram ranged from 19 to 42 weeks (mean, 29.2 ± 5.7 weeks), and the average heart rate was 140 ± 9 beats/min. Among 81 fetuses, strain analysis was performed successfully in 62 (77%) for LV longitudinal function, 64 (79%) for RV longitudinal function, and 49 (64%) for LV circumferential function ( Table 1 ). The frequency of successful analysis was lower for LV circumferential analysis, especially in late-gestation fetuses.



Table 1

Success measuring strain according to GA, ventricle, and strain dimension


































GA (wk) Longitudinal RV strain Longitudinal LV strain Circumferential LV strain
≤24 16/21 (76%) 17/21 (81%) 16/20 (80%)
25–29+6/7 17/22 (77%) 16/22 (72%) 16/22 (72%)
30–34+6/7 17/20 (85%) 17/20 (85%) 12/19 (63%) †,‡
≥35 12/18 (67%) 12/18 (67%) 5/16 (31%) ∗,†,‡
Total 62/81 (77%) 64/81 (79%) 49/77 (64%) †,‡

RV , Right ventricular.

P < .05 versus previous GA group in the same strain analysis subgroup.


P < .05 versus RV longitudinal strain in the same GA subgroup.


P < .05 versus LV longitudinal strain in the same GA subgroup.



Longitudinal and Circumferential Strain


Global RV and LV longitudinal strains were 16.0 ± 3.3% and 15.2 ± 2.7%, respectively, and average mid-LV circumferential strain was 18.7 ± 3.3%. There was no correlation between GA and RV longitudinal strain, LV longitudinal strain, or LV circumferential strain ( Figure 3 ). Values of segmental longitudinal and circumferential strain are shown in Table 2 . Figure 4 is a representative strain curve for circumferential strain.




Figure 3


Peak negative global strain plotted against GA in right ventricular (RV)longitudinal (A) , LV longitudinal (B) , and LV circumferential strain (C) . Blue symbols indicate RV longitudinal strain, red symbols indicate LV longitudinal strain, and black symbols indicate LV circumferential strain.


Table 2

Average and segmental longitudinal and circumferential strain




































Free wall Septum
Basal Mid Apical Basal Mid Apical Global strain
RV longitudinal strain (%) 17.6 ± 5.6 17.7 ± 4.4 16.8 ± 5.0 15.4 ± 4.4 14.4 ± 4.6 14.3 ± 5.0 16.0 ± 3.3
LV longitudinal strain (%) 16.3 ± 4.5 15.9 ± 3.8 15.4 ± 4.8 14.2 ± 4.6 14.0 ± 4.0 15.2 ± 2.7 15.2 ± 2.7






















Anterior Lateral Posterior Inferior Septal Anteroseptal Average strain
Mid-LV circumferential strain (%) 19.9 ± 5.5 19.1 ± 5.7 17.5 ± 5.4 18.7 ± 5.8 19.1 ± 5.4 18.0 ± 5.4 18.7 ± 3.3

RV , Right ventricular.



Figure 4


Circumferential strain curves. Each curve represents each segmental strain and its average.


Fetuses with AS


Five fetuses with AS and no other cardiac malformation, including LV hypoplasia, were selected for analysis. Patients 1 and 2 met criteria for fetal intervention and underwent fAVP for AS soon after the echocardiographic evaluation used in this study. Patients 3 and 4 did not meet criteria for intervention and underwent fAVP 2 and 3 weeks later, respectively. Patient 5 (still in utero) was observed without intervention for 3 months.


One fetus (patient 1) that had both longitudinal and circumferential strains > 2 SDs below average underwent immediate fAVP. The two fetuses (patients 2 and 3) that had borderline low circumferential strain underwent intervention the next day ( n = 1) or after 2 weeks of observation ( n = 1). In the two fetuses with AS that had average or above-average circumferential strain (patients 4 and 5), only the fetus that had borderline high circumferential stain underwent fAVP ( Table 3 ).



Table 3

Profiles of fetal AS




























































Patient GA (wk) LV strain (%), mean ± SD LV dysfunction Peak ASPG (mm Hg) LVEF (%) Outcome
Global longitudinal Average mid-LV circumferential
1 22 8.8 ± −2.3 8.4 ± −3.1 Moderate 35 46 Immediate fAVP
2 22 10.8 ± −1.6 12.8 ± −1.8 Moderate 62 40 Immediate fAVP
3 23 5.3 ± −3.6 12.6 ± −1.8 Mild 63 55 2 wk to fAVP
4 23 11.5 ± −1.3 23.0 ± 1.3 None to mild 28 NA 3 wk to fAVP
5 23 13.2 ± −0.7 18.5 ± −0.1 None to mild 10 68 3-mo observation (in utero)