Aortic Stenosis

2 Aortic Stenosis image


Given the prevalence of valvar aortic stenosis (AS), its huge clinical burden, and the impressive salvage rate with surgical valve replacement, and now with percutaneous valve replacement, identification and description of AS is a prime application of echocardiography. It is also one of the most elegant and relevant applications of Doppler physics in the evaluation of cardiac disease.


Although the basic principles of gradient and area determination are simple, the disease of AS and its innumerable permutations are not, and neither are the subtleties of testing that are responsible for many instances of discordance between different modalities. Rigorous attention to scanning details is paramount, as is proficiency with both noninvasive and invasive aortic valve assessment of hemodynamics, and the ability to navigate discordance with catheterization-derived estimates of aortic stenosis.




Scanning Issues




Scanning Notes




LVOT measurement


image To minimize error, a zoom view must be used.


image If the image quality confers ambiguity, repeat the measurement on several different zoom views.


image LVOT is the single most important measurement: a 2-mm error confers a 20% aortic valve area (AVA) calculation error.


image If the LVOT diameter is in doubt, consider:





V1 measurement


image Ensure that V1 Doppler sampling is correctly aligned—in some cases a more lateral apical sampling site is required to achieve acceptable alignment (within 20 degrees) if the LVOT long axis deviates markedly from that of the LV.


image Do not record the V1 from within the LV cavity, or the recorded velocity and AVA tend to be too low.


image Record the subvalvar V1 before the flow acceleration (easily depicted by the proximal isovelocity surface area [PISA]). Establishing the location of the PISA and sampling V1 to avoid the PISA is better technique than is arbitrarily placing the sample volume “1 cm beneath the aortic valve,” which may actually be within flow acceleration from the valvar stenosis or from concurrent subvalvar obstruction.


image Although ideally the spectral envelope would be measured (planimetered) at the modal frequency (velocity), this is seldom discernable; therefore, planimetry is, for clinical purposes, performed on the outermost aspect of the spectral profile.


image AVA calculations traditionally are made from integrals; peak velocities are a surrogate.


image If sinus rhythm: measure three spectral profiles.


image If in atrial fibrillation: measure five spectral profiles.


V2 measurement


image Ensure that the apical five-chamber and apical three-chamber V2 Doppler sampling views are correctly aligned (within 20 degrees). In some cases, a more lateral sampling site is needed.


image If a concurrent subvalvar stenosis (LVOT velocity [V1] >1.5 m/sec) is likely, record the pre-subvalvar flow velocity, the pre-valvar velocity (V1), and also the V2. The modified Bernouilli equation should not be used if V1 is ≥1.5 m/sec.


image AVA calculations traditionally are made from integrals; peak velocities are a surrogate, although a reasonably accurate one.


image If sinus rhythm: measure three spectral profiles.


image If in atrial fibrillation: measure five spectral profiles.


image If in the idea world: measure ten spectral profiles.


image Annotate the site of sampling for future reference comparison.


Confounders


image Verify whether there is/is not concurrent intracavitary flow acceleration (subaortic valve velocity ≥ 1.5 m/sec), as discussed in the following section.


image A narrow aortic root (STJ ≤ 30 mm): a narrow aorta facilitates the “pressure recovery” phenomenon that detects a higher gradient than the recovered gradient measured by catheterization.




Pathophysiology and Findings of Aortic Stenosis


The fundamental pathophysiology of AS is the excess pressure imposed on the LV, known as pressure overload. Secondarily, or indirectly, a lesser pressure load is imposed on the LA by diastolic failure of the LV. The LV response to pressure overload—concentric hypertrophy—is adaptive; it normalizes the wall stress on myocytes, but it is physiologically expensive. Essentially, the wall thickness increases in proportion to the pressure increase. Laplace’s law states that wall stress (i.e., myocyte stress) is proportional to intracavitary pressure and radius, and inversely proportional to wall thickness. Therefore, if the proportional increase in LV wall thickness is the same as the proportional increase in left ventricular pressure (due to imposition of the transvalvar gradient), then wall stress is normalized, and myocyte preservation is facilitated.


The pattern of hypertrophy in AS is typical of pressure overload, therefore, and should be present in most cases of severe AS. In concentric hypertrophy, myocardial sarcomeres replicate in parallel; therefore the walls thicken and the cavity dimensions do not increase (in fact, they often decrease). Therefore, the pattern is thick walls and no increase in cavitary dimensions, unless complications arise, or the disease is in a terminal state.


In severe AS, where the systolic pressure within the LV doubles, LV mass essentially doubles (increases to 178 g/m2 vs. normal of 86 g/m2),1 due to near doubling of the wall thickness to normalize wall stress that otherwise would be nearly doubled (Fig. 2-1).



Given the increase in both myocardial mass and generated systolic pressure, myocardial O2 demand increases (mVO2 ∝ LV mass, developed systolic pressure, contractility, and heart rate). In AS, coronary flow (supply) is a problematic issue that compounds the potential problem of increased demand; about half of adult patients with AS have concurrent coronary artery disease (CAD).


Severe AS may remain seemingly static and compensated for years, or may progress. The rate of progression depends on many factors, few of which are well understood. Progression rates between –0.01 cm2/year and –0.1 cm2/year have been described. Calcification of the aortic valve is somewhat predictive of a faster rate of progression, but it is not known how to quantify calcification toward this purpose. Electron beam CT studies have not shown good correlation of calcification quantification by Agatston score with aortic valve area, especially for moderate and severe AS.2


Both increasing left ventricular systolic pressure and LV hypertrophy impair LV compliance; left atrial hypertrophy and augmented atrial systolic function occur to maintain left ventricular filling. In severe AS, the left atrium typically is mildly enlarged.


In summary, the pathophysiology of severe AS generates expected findings: increased wall thickness, no increase in left ventricular cavitary size, and left atrial dilation. An increase in cavitary size requires additional explanation, such as concurrent volume overload from aortic insufficiency or mitral regurgitation, impaired systolic function from CAD or cardiomyopathy, or terminal decompensated AS.




Reporting Issues


Aortic valve gradient is a per-beat (volume) function. The gradient of aortic stenosis occurs through the systolic ejection period. The mean gradient is the most suitable expression of the average pressure load imposed on the LV through this period. The gradient reflects both the severity of the impedance to ejection (imparted by the aortic valve) and the function of the LV. The expression of LV function that is most germane to aortic stenosis is the stroke volume, not the EF or CO.



Gradient Issues


When reporting, the combination of mean gradient and stroke volume must be emphasized. For example: “The mean gradient is 65 mm Hg with a normal stroke volume of 75 mL.” Normal stroke volume index is 45 ± 13 mL/m2. It is not necessary to describe the peak instantaneous gradient; this just generates unjustifiable confusion with the (different) catheterization-derived peak-to-peak gradient, and is not the criterion for severity. Gradients, valve area, and LV function must be compared to previous determinations. If there is a difference in the gradient, ensure that the recording was obtained from a comparable sampling site.


It is important to enter annotations (e.g., apical, right parasternal, suprasternal) on all spectral Doppler recordings, to make appropriate comparisons possible. Gradients may be enhanced by numerous factors, only a few of which may be apparent when reading an echocardiogram. Factors that increase the aortic valve gradient include the following:



image Bradycardia in the context of normal LV function or contractile recruitment


image Causes of larger stroke volume ± lower peripheral resistance









image Undersampling of the V2 spectral profile is a common clinical problem, as the spectral profile boundary may be vague.




image Fewer than 5% of cases of AS will have inadequate spectral profiles; the percentage is very much determined by the time and effort made to acquire the spectral recording.


image Adequate spectral profiles are complete and plausible parabolic profiles.


image If the peak velocity is to be marked with cross-hairs, the cross-hairs should not be placed on the peak of a spectral profile, because that may falsely confer an impression of the true peak to the reviewer. The cross-hairs should be placed off to the side of the presumed peak to allow the reviewing or reporting physician to establish independently that the peak was true, as a useful internal check.


image If the profile initially does not appear complete, persistence in attempting better ones is required, as there is a (albeit diminished) return. Techniques that should be used include the following:






image The published correlation of Doppler versus catheterization mean gradients averages 0.90,3,4 but importantly, 1 standard error of estimate (SEE) of Doppler gradient versus catheterization gradient is actually 10 mm Hg [SEE range, 6–12 mm Hg].5


image Rahimtoola5 emphasized consideration of the SEE by Doppler when describing AS severity (by gradient):





image Recall that the pressure recovery phenomenon at aortic root level is seen with small aortic roots (dimensions <30 mm) and restrictive planar divergent orifices. Pressure recovery may add 15% to 30% to the gradient.


image LVSWL, originally a catheterization-based parameter, can be approximated by echocardiography, and has been validated as a way to discriminate clinical end-points: an LVSWL ≤25 is the best predictor.6 The concept expresses kinetic energy loss across the aortic valve in reference to the aortic pressure.



Continuity Equation–Derived Aortic Valve Area Issues


The idea of aortic valve area as a conceptual expression of AS is attractive, because it is intended to be a flow-independent description of AS severity, and it is widely accepted that flow does vary across stenotic aortic valves, reducing the predictiveness of gradient alone. Because flow across the aortic valve per cardiac cycle may vary, the gradient should be offered in the context of the stroke volume (index). Unfortunately, precise determination of the stenotic aortic valve area, by echo and also by catheterization, is neither as simple nor accurate as is believed. The calculation of AVA by echo and by catheterization actually is not “flow-independent” by either technique, because so many parameters are involved in the equations that the total introduced error becomes significant and sometimes problematic. Furthermore, plastic deformation of the aortic valve (i.e., greater opening under higher gradients and less opening under lower gradients) is known to occur.






Other Issues


Although the focus of an echocardiographic study is on detailed examination and description of the valve and related hemodynamics, it also is necessary to assess for associations, complications, and concurrent diseases.


If the aortic valve is bicuspid, describe possible associations. Recall that a high-velocity AS jet may carry around the aortic arch—in other words, higher-velocity flow in the proximal descending aorta may not be coarctation generated, and represents only transmission of the AS jet. The diagnosis of coarctation of the aorta by echocardiograophy requires demonstration of a focal step-up of flow velocity.


Describe the dimensions and appearance of the root and ascending aorta. These are potentially important surgical details that may modify the approach to surgery. Furthermore, if a case is obviously severe, and the aortic annulus is small, the size of the annulus should be mentioned. An unusually small root will allow only a small prosthesis, which engenders a gradient little better than the AS it was supposed to relieve—an outcome known as “patient–prosthesis mis-match.” Although such a mismatch probably is not as significant a clinical problem as has been purported, it still has validity in some cases.


The presence of LV hypertrophy is expected. Its absence is conspicuously inconsistent with severe AS. The presence of wall motion abnormalities is important.



Concurrent Subvalvar Stenosis


Subvalvar AS may masquerade as valvar AS, or may be an unsuspected concurrent lesion. Concurrent subvalvar AS confounds accurate estimation of the aortic valve hemodynamics, and is not treated by AVR alone.




Low-Gradient Severe Aortic Stenosis




image The definition of “low-gradient severe AS” is unresolved and variable, but may be approached as follows:


Severe AS with an AVA ≤ 0.7 cm2 with a mean gradient of ≤30 mm Hg due to reduced stroke volume.9 Several papers have used an AVA of ≤1 cm2 and a mean gradient of ≤40 mm Hg.10

image The challenge is to distinguish patients with end-stage AS with failing LV systolic function (responsible for the low gradient) from patients with moderate AS with poor LV systolic function, in whom a factor other than AS is responsible for the low gradient.


image Low-gradient severe AS can be discounted by a convincingly normal stroke volume: if the gradient is low and the stroke volume is normal, then low-gradient severe AS is unlikely to be present.


image Citing low EF% or “grade” to describe low-output AS is not adequate. The stroke volume should be shown to be low to establish that the per-beat output is low, and potentially responsible for the low gradient. There may be normal stroke volume with a low EF% LV (if dilated) and low stroke volume with an LV with normal EF% (if it is under-loaded or if there is severe MR). Therefore, LV grade is not synonymous with output.


image EF% generally increases after AVR10,11 unless a perioperative infarction occurs, or the LV is intractably stiff. It appears that selected cases of low-gradient severe AS still benefit from AVR,12 although the data in this field are preliminary.13


image In the presence of low output, both the Gorlin catheterization formula11,14 and continuity methods are less accurate in the estimation of AVA.


image When catheterizing suspected low-gradient AS, use of either two catheters or double-lumen catheters should be considered to eliminate phase shift artifacts, which can introduce a difference of >10 mm Hg in the gradient recording from a femoral side-arm.


image Low-gradient AS should be reported as a possibility or probability, depending on how strongly it is believed to be present.


image Valve resistance has been proposed as a means of establishing the severity of AS (>300 dynes/sec/cm−5 = severe disease) that is independent of flow output, but it has not been proved to be better than valve area calculation,9 nor has it been popular. It appears less useful than LVSWL.6


image Determination of contractile reserve in low-gradient severe AS cases







Rahimtoola15 would caution that changes in stroke volume are labeled as an index of contractility and as the hemodynamic responses to dobutamine are complex.

image In cases of suspected low-gradient severe AS, particular scrutiny should be afforded to the size of the aortic annulus/root. If the root is unusually small (e.g., 19–20 mm), then the size of prosthesis that would be implanted would confer a moderate gradient, which might be little different from the preoperative gradient.


Jun 12, 2016 | Posted by in CARDIOLOGY | Comments Off on Aortic Stenosis

Full access? Get Clinical Tree

Get Clinical Tree app for offline access