Validation and Reliability Testing of Blood Pressure Monitors


Year

BHS (only)

BHS + AAMI

AAMI (only)

BHS (all)

AAMI (all)

ESH-IP

Total

2000

2 (20 %)

4 (40 %)

0 (0 %)

6 (60 %)

4 (40 %)


10

2001

1 (8 %)

5 (42 %)

1 (8 %)

6 (50 %)

6 (50 %)


12

2002

4 (29 %)

3 (21 %)

1 (7 %)

7 (50 %)

4 (29 %)

3 (21 %)

14

2003

2 (13 %)

0 (0 %)

6 (38 %)

2 (13 %)

6 (38 %)

8 (50 %)

16

2004

1 (6 %)

5 (29 %)

1 (6 %)

7 (41 %)

6 (35 %)

4 (24 %)

17

2005

2 (12 %)

4 (24 %)

0 (0 %)

6 (35 %)

4 (24 %)

7 (41 %)

17

2006

4 (13 %)

5 (17 %)

1 (3 %)

9 (30 %)

6 (20 %)

15 (50 %)

30

2007

1 (3 %)

2 (6 %)

3 (9 %)

3 (9 %)

5 (16 %)

24 (75 %)

32

2008

3 (9 %)

2 (6 %)

1 (3 %)

6 (17 %)

4 (11 %)

25 (71 %)

35

2009a

4 (14 %)

3 (10 %)

0 (0 %)

8 (28 %)

3 (10 %)

18 (62 %)

29

2002–2009a

21 (11 %)

24 (13 %)

13 (7 %)

48 (25 %)

38 (20 %)

104 (55 %)

190


AAMI Association for the Advancement of Medical Instrumentation, BHS British Hypertension Society, ESH-IP European Society of Hypertension International Protocol

aUntil 15/06/2009



A145037_3_En_5_Fig1_HTML.gif


Fig. 5.1
Cumulative graph of validation studies performed according to the European Society of Hypertension International Protocol (ESH-IP) compared with the British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) protocols from 2002 (publication of ESH-IP) until June 2009. Reproduced from Stergiou et al. [25] with permission from the authors


In its introduction, the 2002 ESH-IP protocol included the sentence “It is anticipated that the relative ease of performance of the International Protocol will encourage manufacturers to submit blood pressure measuring devices for validation in order to obtain the minimum approval necessary for a device to be used in clinical medicine, and that, in time, most devices on the market will be assessed according to the protocol for basic accuracy”. The figures demonstrate that while the goal of having “most devices on the market” validated has still some way to go, the protocol did encourage manufacturers to submit devices for validation.

The analysis demonstrates the importance of practicality in the development of protocols. Protocols that are difficult and onerous to complete are less attractive both to centres and to manufacturers. Furthermore, they are rarely carried out correctly. Ideal evaluation is therefore best achieved, and perhaps can only be achieved, by a set of simple validations rather than by a single complex one.



Differences Between Protocols


At present, the ESH-IP validation protocol is the most widely used, followed by the AAMI/ANSI/ISO standard. There are some substantial differences between these validation procedures, as described below and shown in Table 5.2. A detailed comparison has also been made by Ng [28]. Nevertheless, they both have a common objective, namely the standardisation of validation procedures to establish minimum standards of accuracy and performance, and to facilitate comparison of one device with another. With the substantial data available, it is perhaps timely that a common protocol be developed that will become a world standard.


Table 5.2
Comparison of ESH-IP 2010 [24] and AAMI/ANSI/ISO 81060-2:2013 [14] requirements for noninvasive validation of blood pressure devices





































































































































































































































































Protocol provision

ESH-IP 2010

AAMI/ANSI/ISO 81060-2:2013

Sample demographics

Sample size

33

85+

Age

≥25 years

Adult and adolescent

>12 years

Adult, adolescent, and children

35 subjects 3–12 years and

50 + subjects >25 years

Sex

≥10 Male and ≥10 Female

≥30 % Male and ≥30 % Female

Arm circumference

1 cuff

No requirements

≥40 % upper ½, ≥40 % lower ½,

≥20 % upper ¼, ≥20 % lower ¼

n cuffs, n > 1

≥1/2n subjects/cuff

Methodology

Sequential same arm

7 alternating measurements (4 observer, 3 test device)

7 alternating measurements (4 observer, 3 test device)

Control measurement is average of 4 values (preceding and succeeding measurements of each observer)

Control measurement is the nearer on the preceding and succeeding mean observer measurements to test measurement

Exclude measurement if observer measurements differ by >12 mmHg SBP or >8 mmHg DBP from other 2 measurements. (Max. 10 % of subjects; exclude subject if >10 %)

Simultaneous same arm

Not permitted

Test device: Inflates ≥SBP + 20 mmHg, deflates ≤DBP—20 mmHg, deflation rate 2 mmHg/s to 3 mmHg/s

3 Pairs of measurements

Control measurement is mean observer measurement

Exclude subject if observer measurements differ by >12 mmHg SBP or >8 mmHg DBP

Simultaneous opposite arm

Not permitted

3 Pairs of measurements with test device on left arm +3 pairs with test device on right arm

Control measurement is mean observer measurement with lateral arm difference adjustment

Exclude subject if observer measurements on same arm differ by >12 mmHg SBP or >8 mmHg DBP of if lateral difference >15 mmHg SBP or >10 mmHg DBP

Recruitment blood pressure

SBP

90–100 mmHga

10–12

≥1

Initial pressures are recorded on each subject but they are not used in any calculations. Blood pressure distribution requirements are based on the reference control measurements only

101–129 mmHg
 

130–160 mmHg

10–12

161–169 mmHg

10–12
 

170–180 mmHga

≥1

DBP

40–50 mmHga

10–12

≥1

51–79 mmHg
 

80–100 mmHg

10–12

101–119 mmHg

10–12
 

120–130 mmHga

≥1

Control BP distribution

SBP

≤100 mmHg

22/99–44/99

Max. difference ≤19/99

≥5 % (13/255)

101–129 mmHg

≤75 % (191/255)

130–139 mmHg

22/99–44/99

140–159 mmHg

≥20 % (51/255)
 

160 mmHg

≥5 % (13/255)

>160 mmHg

22/99–44/99

DBP

≤60 mmHg

22/99–44/99

Max. difference ≤19/99

≥5 %

61–79 mmHg

≤75 %

80–84 mmHg

22/99–44/99

85–99 mmHg

≥20 %
 

100 mmHg

≥5 %

>100 mmHg

22/99–44/99

Passing requirements (SBP and DBP)

ESH-IP

Part 1 based on individual measurement errors

No. of errors

≤5 mmHg

≤10 mmHg

≤15 mmHg
 

≥2 of

≥73/99

≥87/99

≥96/99

All of

≥65/99

≥81/99

≥95/99

Part 2 based on measurement errors per subject

No. of subjects

≥2/3 errors

0/3 errors

≤5 mmHg

≤5 mmHg

Both

≥24/33

≤3/33

81060-2:2013

Criterion 1 based on individual measurement errors

For comparison with ESH-IP, expected error rates within bands are shown when the mean error is 5 mmHg ± 8 mmHg

≤5 mmHg

≤10 mmHg

≤15 mmHg

Mean error (255 measurements)

Max. SD

39.3 %

70.1 %

88.5 %

5 mmHg

8 mmHg

Criterion 2 based on mean subject errors such that, using the normal distribution, the probability of a mean subject error being within 10 mmHg is ≥0.85. Selected examples shown

Probability of mean subject error within bands, based on normal distribution

Sample mean error and SD combinations that satisfy criteria

≤5 mmHg

≤10 mmHg

≤15 mmHg

Mean error (85 subject mean BPs)

Max. SD

0.528

0.850

0.969

0 mmHg

6.95 mmHg

0.529

0.850

0.969

1 mmHg

6.87 mmHg

0.528

0.850

0.969

2 mmHg

6.65 mmHg

0.525

0.850

0.971

3 mmHg

6.25 mmHg

0.515

0.850

0.974

4 mmHg

5.64 mmHg

0.482

0.851

0.982

5 mmHg

4.79 mmHg

Special circumstances

Exercise

No requirements

Supplementary study using 35 subjects


aMaximum of four instances in total with SBP < 90 mmHg, SBP > 180 mmHg, DBP < 40 mmHg or DBP > 130 mmHg


Passing Requirements


The AAMI/ISO standard requires two criteria to be fulfilled. In Criterion 1, the mean error of at least 255 measurements must be at most 5 mmHg with a standard deviation of at most 8 mmHg. In Criterion 2, the mean error of the mean measurements from at least 85 subjects must be at most 5 mmHg and the standard deviation at most a value to ensure that the mean error on at least 85 % of subjects is within 10 mmHg. (By extension, this definition also means that the expected number of subjects with average errors within the more commonly accepted error of 5 mmHg is around 50 %.) A table, showing the acceptable standard deviations for ranges of mean errors, is provided. For ABPM devices, similar criteria must also be applied to a further raised-heartrates test.

The ESH-IP protocol provides two sets of cut-off points. The primary set is for at least 73/99 measurement errors to be within 5 mmHg, at least 87/99 to be within 10 mmHg and at least 96/99 to be within 15 mmHg. The secondary set is for at least 65/99 measurement errors to be within 5 mmHg, at least 81/99 to be within 10 mmHg and at least 95/99 to be within 15 mmHg. Devices must fulfil at least two of the primary set and all of the secondary set. In addition, in at least 24/33 subjects, the errors in at least two of the three measurements recorded must be no more than 5 mmHg and there can be no more than 3/33 subjects with errors over 5 mmHg in all three measurements.


Methodology


The AAMI/ANSI/ISO standard allows simultaneous same-arm comparisons if the test device inflates to at least 20 mmHg above SBP and deflates to at least 20 mmHg below DBP at a rate of between 2 and 3 mmHg/s. Otherwise, sequential same-arm comparisons or simultaneous opposite-arm comparisons must be used. The ESH-IP protocol stipulates sequential same-arm comparisons in all circumstances.


Recruitment Requirements


For Criterion 1, the AAMI/ISO standard requires 255 measurements to be recorded in at least 85 subjects, with at most three measurements per subject. For Criterion 2, 85 subjects with at least three measurements per subject are required. This number of measurements per subject is doubled if the simultaneous opposite-arm methodology is used, with each arm being used for half the control and test measurements. For ABPM devices, a further 35 subjects are required for a validation with raised heartrates following exercise.

The ESH-IP protocol requires three measurements to be recorded in each of 33 subjects. However, strict requirements generally necessitate several more subjects to be recruited in order for these to be satisfied.


Pressure-Range Requirements


The ESH-IP protocol requires between 10 and 12 subjects with SBP in the ranges <130, 130–160 and >160 mmHg; between 10 and 12 subjects with DBP in the ranges <80, 80–100 and >100 mmHg; a total of at most for subjects with pressures outside the “ideal” range of 90–180 mmHg for SBP and 40–130 mmHg for DBP and minimum ranges of 100–170 mmHg for SBP and 50–120 mmHg for DBP. In addition, the 99 control measurements must not be statistically different from the ideal 33 measurements in each range; that is each range must contain 22 and 44 measurements and the difference between the range with the highest count and that with the lowest count cannot exceed 19.

The AAMI/ISO standard requires at least 5 % (13 measurements, requiring at least five subjects) of reference systolic blood pressures have to be at least 160 mmHg, at least 20 % (51 measurements, requiring at least 17 subjects) at least 140 mmHg and at least 5 % at most 100 mmHg. Similarly, at least 5 % of reference diastolic blood pressures have to be at least 100 mmHg, at least 20 % at least 85 mmHg and at least 5 % at most 60 mmHg.


Age Distribution Requirements


The ESH-IP protocol requires all subjects to be at least 25 years of age. The AAMI/ISO standard defines two types of studies based on age. The first is for an adult/adolescent population defined simply as on persons over the age of 12. The alternative population is a combined adult/paediatric population where 35 subjects must be between the ages of 3 and 12; the remainder (minimum 50) must be over the age of 12.


Sex Distribution Requirements


Both the ESH-IP protocol and the AAMI/ISO standard are at ad idem requiring a minimum of 30 % (stated as 10/33 in the ESH-IP protocol) male and 30 % female subjects.
< div class='tao-gold-member'>

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Jul 8, 2016 | Posted by in CARDIOLOGY | Comments Off on Validation and Reliability Testing of Blood Pressure Monitors

Full access? Get Clinical Tree

Get Clinical Tree app for offline access