Understanding Imaging Artifacts




Abstract


Imaging artifacts encompass patterns in the image that seem to suggest the presence of structures that are in reality not present. They may in fact relate to both the appearance of nonexistent structures, as well as the concealing of existing structures. Artifacts are mostly caused by physical interactions between the imaged tissue and ultrasound itself that are more complex than assumed by the ultrasound system; however, they can also result from malfunctioning ultrasound equipment or their inadequate settings, as well as from interference caused by other electronic equipment. Imaging artifacts can occasionally encumber an echocardiographic examination (particularly for novices), and the knowledge of these occurrences should thus be used to minimize their effect.




Keywords

attenuation artifacts, refraction artifacts, reverberation artifacts, side-lobe artifacts, ultrasound artifacts

 




Introduction


Imaging artifacts encompass patterns in the image that seem to suggest the presence of structures that are in reality not present. They may in fact relate to both the appearance of nonexistent structures, as well as the concealing of existing structures. Artifacts are mostly caused by physical interactions between the imaged tissue and ultrasound itself that are more complex than assumed by the ultrasound system; however, they can also result from malfunctioning ultrasound equipment or their inadequate settings, as well as from interference caused by other electronic equipment. Imaging artifacts can occasionally encumber an echocardiographic examination (particularly for novices), and the knowledge of these occurrences should thus be used to minimize their effect. In this chapter, the most common artifacts have been categorized according to the authors’ discretion, although they recognize that alternative ways of classifying these artifacts are possible, as no standard nomenclature exists and reference to certain artifacts in the literature is limited ( Table 7.1 ).





B-Mode Artifacts


Shadowing (Attenuation) and Dropout Artifacts


Shadowing (i.e., attenuation) artifacts obscure certain (segments of) underlying structures. When imaging structures with mechanical properties (i.e., mass density and/or compressibility) substantially differ from soft tissue (e.g., metals such as used in prosthetic valves or air/contrast bubbles), very strong reflections will occur, resulting in little or no transfer of ultrasound energy to more distal regions, as (almost) all energy will have been reflected. This will manifest as a strong reflection in the area of the reflector, followed by an “acoustic shadow” that represents a signal void ( Fig. 7.1 , yellow dashed arrows, ).




FIG. 7.1


Apical four-chamber view of a patient with a prosthetic mitral valve, which is the origin of multiple acoustic shadows—that is, signal voids (yellow dashed arrows) —due to proximal strong reflecting structures. Furthermore, the cusps of the mitral prosthesis give rise to step ladder artifacts (as explained further in the text; blue arrows ), seen throughout the central portion of the left atrium and beyond the roof of the left atrium. These artifacts can also be appreciated on the cine loop of this figure ( ).


This shadowing will affect not only the two-dimensional (2D) image but also the color Doppler signal, which could be, for example, highly relevant in the assessment of regurgitant jets in the setting of prosthetic heart valves. Heavily calcified tissue is a similarly strong reflector; resorting to alternative scanning windows can mostly circumvent such artifacts. For more information on negative shadows, refer to Box 7.1 and Fig. 7.2 .



BOX 7.1


Although shadowing is typically seen as a darker zone (i.e., signal void) distal to a highly reflecting (most common) or absorbing (less common) structure, the opposite can also occur in that a structure is attenuating less than what is assumed by the scanner’s automatic time gain compensation (refer to Chapter 1 ). As a result, the distal echo signals are overamplified by the scanner, resulting in a brighter zone distal to the low-attenuating structure, which is referred to as a “negative” shadow. A typical example is that of a cyst, as the liquid in the cyst is attenuating the ultrasound wave less than soft tissue, while the scanner automatically corrects for attenuation, assuming it is imaging soft tissue only. An example of such a negative shadow following a cyst in the liver is given in Fig. 7.2 . In cardiac imaging, negative shadows do not normally occur.


Negative Shadows



FIG. 7.2


Example of a shadow (yellow dashed arrow) and negative shadow following a cyst (yellow arrow) in the liver.


Similarly, superficial structures leading to notable attenuation of ultrasound may significantly impair its penetration. As a result, the ribs or lung tissue can diminish the ability to image underlying structures, giving rise to “dropout” artifacts that typically occur at some phases of the respiratory cycle. Such artifacts can be reduced or avoided by scanning at different intervals of the respiratory cycle (i.e., breath hold), and occasionally only by choosing another transducer position ( Fig. 7.3 , ). A dropout artifact is thus similar to a shadowing artifact that occurs very near the transducer, thereby causing part of the image to become invisible.




FIG. 7.3


Dropout artifact imaged on an M-mode trace of the parasternal long-axis.

In this case, during inspiration (as appreciated from the respirometer trace— white arrow ) the tissue close to the transducer (e.g., ribs or lung tissue) causes attenuation of the echo signal, thus obscuring the underlying image. The cine loop of the corresponding B-mode image can be found under , where it can be appreciated that the visibility of the underlying structures clearly depends on the phase of the respiratory cycle. is taken during normal respiration, while shows a corresponding cine loop taken during breath hold at end-expiration, demonstrating the relevance of scanning at various intervals of the respiratory cycle for the optimization of image acquisition.


Reverberation Artifacts


During image generation, the signal of the reflected wave arriving at the transducer is transformed to electrical energy (i.e., the radiofrequency signal) to be used in the image reconstruction process (for more details, see Chapter 1 ). However, a certain portion of the reflected wave is not converted to electrical energy and is merely reflected on the transducer surface to start repropagating through the tissue as if it were another ultrasound transmission. This secondary “transmission” is yet again reflected by the tissue and ultimately detected by the transducer ( Fig. 7.4 , upper panel ). Since the ultrasound system assumes all echo signals are resulting from the original transmission, these secondary reflections that arrive late (as they have to travel back and forth twice) will thus be depicted at a greater depth, mimicking a reflecting structure at a greater distance (see Fig. 7.4 , lower panel ).




FIG. 7.4


The generation of a reverberation artifact.

A transmitted pulse (green) reflects form a structure and results in an echo signal (green) . The reflected pulse will partially reflect at the surface of the transducer (red) and be the source of secondary echo signals (red; upper panel) . The additional time required for the “back-and-forth” travel of the signal causes the ultrasound system to depict the artifact distal to the true image, positioned at a distance that is an integer multiple of the true distance (d) between the transducer and the reflective structure (lower panel) .

Modified from D’hooge J, Mertens LL. Ultrasound physics. In: Lai WW, Mertens LL, Cohen MS, Geva T, eds. Echocardiography in pediatric and congenital heart disease: from fetus to adult, Second Edition. Chichester: John Wiley and Sons; 2016:8.


In fact, the secondary reflections will also be partially reflected at the transducer surface, resulting in an apparent third transmit; this process can be repeated multiple times. Although these second- and/or higher-order reflections always occur, they are normally negligible in amplitude with respect to the primary echo amplitudes and can therefore simply be ignored. However, when strong reflections of the primary wave occur, the reverberated signal (i.e., the signal caused by two or more round-trips) can be substantial and give rise to image artifacts (i.e., reverberation artifacts) . This type of artifact can typically be recognized as the repetition of a strongly reflecting structure at integer multiple distances from the transducer. When the reflecting structure causing the reverberation is moving, so will the induced reverberation (with an amplified motion amplitude; ). If, on the other hand, this structure is not moving (e.g., a rib), the induced artifact will be stationary as well (i.e., a stationary reverberation artifact ; ).


Similarly, reverberations can occur within structures that are being visualized. Indeed, on their way back to the transducer, reflections from the primary wave can meet strong reflectors that cause these reflections to start propagating in the direction away from the transducer again ( Fig. 7.5 ). As such, these waves are propagating with some time lag with respect to the primary wave and create secondary reflections. Obviously, similar as for the transducer-related reverberations described previously, higher-order reflections can arise by multiple round-trips between both boundaries of the object. Structures with well-aligned, flat, strong reflecting boundaries (due to a strong mismatch in mechanical properties), such as catheters or the metallic leaflets of a prosthetic valve, are most likely to give rise to this acoustic “ping-pong.” When the reverberating structure is substantially larger than the length of the ultrasonic pulse, individual “ping-pong” reflections can be detected ( Fig. 7.6 , blue arrows; Fig. 7.7 , yellow arrows; and corresponding ), in which case the name “ step ladder ” artifact is used. On the other hand, if this structure is small compared with the pulse length, all “ping-pong” reflections start to interfere, giving a more hazy appearance in the image referred to as a “ comet tail ” artifact ( Fig. 7.8 ). Depending on the attenuation the reverberating wave experiences, the “step ladder” or “comet tail” fades out faster (i.e., higher attenuation; see Fig. 7.6 ) or slower (i.e., lower attenuation; see Fig. 7.1 , blue arrow; see and Fig. 7.8 ).




FIG. 7.5


The generation of an internal reverberation artifact.

A transmitted pulse (green) partially reflects form two parallel structures resulting in the expected echo signals at a time corresponding to a distance “d” and “d+w” (green) . However, the reflection from the distal boundary reflects at the proximal one (red) resulting in a late echo signal (red) . The additional time required for the “back-and-forth” travel within the layer causes the ultrasound system to depict the artifact distal to the true boundaries, positioned at a distance equal to twice the thickness (w) of the layer. Most often, this process is repeated resulting in a series of additional (artefactual) echo signals. d, distance; w, thickness of the layer; c, ultrasound propagation speed.

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Sep 15, 2018 | Posted by in CARDIOLOGY | Comments Off on Understanding Imaging Artifacts

Full access? Get Clinical Tree

Get Clinical Tree app for offline access