Three-Dimensional Echocardiographic En Face Views of Ventricular Septal Defects: Feasibility, Accuracy, Imaging Protocols and Reference Image Collection




Background


Ventricular septal defect (VSD) is the most common congenital cardiac anomaly. Accurate assessment is critical for planning treatment. Recent advances in three-dimensional (3D) echocardiography have improved image quality and ease of use.


Methods


The feasibility and accuracy of three specific 3D echocardiographic protocols to demonstrate en face views of VSDs were analyzed in a retrospective review of 100 consecutive patients. Sixty-four patients underwent transthoracic echocardiography and 36 transesophageal echocardiography. Types of VSDs included 34 muscular, 32 perimembranous, 18 malaligned, 11 inlet, four outlet, and one acquired. Ages ranged from 1 day to 77 years, and body weights from 3 to 92 kg. Three-dimensional echocardiographic full-volume mode with standard XYZ and adjustable plane cropping, 3D full-volume mode with iCrop, and narrow-sector live 3D protocols were compared for feasibility and accuracy to obtain a diagnostic-quality en face view of a VSD.


Results


The success rate for obtaining a high-quality en face image for the three protocols was 100% for full-volume mode with iCrop, 97% for full-volume standard mode, and 94% for narrow-sector live 3D mode. The ability of both full-volume mode with iCrop and full-volume standard mode to demonstrate a VSD was slightly better than that of narrow-sector live 3D mode ( P < .001 for both vs narrow-sector live 3D mode). In all patients, the type, size, and location of the VSD were demonstrated accurately by two or more of the protocols.


Conclusions


Three-dimensional echocardiography of VSDs is feasible and accurate in most patients using defined protocols. The protocols are described and illustrated in detail, and a reference 3D image collection is presented.


Isolated ventricular septal defect (VSD) is the most common congenital cardiac anomaly, with an incidence of approximately 3.5 per 1,000 live births. VSDs are also commonly associated with other complex cardiac anomalies. There are six major types of VSD: perimembranous, muscular, inlet, malaligned, outlet, and acquired. Although many small perimembranous and muscular VSDs can be asymptomatic, larger defects and those associated with other cardiac anomalies often require surgical intervention. Accurate assessment of these defects and the surrounding anatomy is critical for appropriate treatment planning.


Two–dimensional (2D) transthoracic echocardiography (TTE) and transesophageal echocardiography (TEE) are well established as reliable methods for initial diagnosis and intraoperative assessment of VSDs. Initial reports of three-dimensional (3D) TTE and TEE have demonstrated their usefulness in evaluating the complex structure of VSDs. These reports have shown novel views of VSDs and the surrounding structures that are not possible using 2D echocardiography.


Specific image acquisition and cropping protocols for en face views of VSDs have not yet been reported, compared, or widely applied in the clinical setting. The development of such protocols, analogous to those developed for 2D echocardiography, may facilitate the more widespread application of 3D echocardiography (3DE) for VSDs. We used 3D TTE and 3D TEE to evaluate 100 consecutive patients with VSDs in order to (1) determine feasibility and accuracy, (2) describe 3D echocardiographic protocols that reliably demonstrate en face views of VSDs, and (3) provide a collection of 3D echocardiographic reference images of the various types of VSDs.


Methods


We studied 100 consecutive patients over a 6-month period who presented to our center with VSDs as isolated lesions or as part of more complex diagnoses. All echocardiographic studies were ordered for clinical indications, and complete 2D echocardiographic examinations were performed. Any patient with a VSD who underwent 2D TTE during the study period also underwent 3D TTE. Any patient with a VSD and body weight > 20 kg who underwent 2D TEE during the study period also underwent 3D TEE. All transesophageal echocardiographic studies were performed during either catheterization or surgical procedures. We considered 20-kg body weight to be a safe lower limit for the use of the 3D-capable transesophageal transducer. Three-dimensional TEE was not available for patients weighing <20 kg.


In addition to the standard clinical 2D echocardiographic protocol, 3DE was performed using a Phillips iE33 system and the appropriate 3D transthoracic (X5-1 and X7-2) or transesophageal (X7-2t) transducer (Philips Medical Systems, Andover, MA). Right- and left-side en face views of VSDs were obtained using three specific 3D echocardiographic protocols applied in an apical transthoracic or a midesophageal four-chamber view. Images were acquired using the iE33 system and acquisition processing and cropping software. Sixty-four patients underwent TTE and 36 TEE. None were excluded. Three-dimensional echocardiographic images were reviewed retrospectively with the reader blinded to the 2D echocardiographic, intraoperative inspection, cardiac catheterization hemodynamic and angiographic, cardiac magnetic resonance imaging (MRI) or magnetic resonance angiographic, and cardiac computed tomographic angiographic results of patients in whom those studies had been performed.


Description of 3D Echocardiographic Modalities


For anatomic assessment of VSDs, we used three different 3D echocardiographic methods: (1) large-sector full-volume 3DE plus standard cropping mode (FVstd), (2) large-sector full-volume 3DE plus iCrop mode (FViCrop), and (3) live narrow-sector 3D mode (L3D) with no additional cropping. All three methods are available for both TTE and TEE. Detailed descriptions of these techniques are as follows.


Three-dimensional full-volume mode consists of a large-sector 3D data set acquired using one to six heartbeats. A one-beat acquisition is an actual live 3D image, whereas the 3D data set in multiple-beat acquisitions is rapidly and automatically constructed in almost real time over two to six heartbeats. Individual live 3D segments are rapidly and automatically stitched together sequentially into a single large pyramid-shaped 3D sample volume. Acquisition of the individual sectors is triggered by electrocardiography. It is very rapid reconstruction, not actual real-time imaging. This allows the acquisition of a large 3D sample volume that includes most cardiac structures in a single volume. It is stored as a full-volume sample and is immediately available on the echocardiographic machine for extensive postacquisition cropping. Unfortunately, the electrocardiographically triggered rapid sequential reconstruction is prone to motion artifacts whereby the individual samples are stitched together because of respiratory motion or rapid heart rate. This problem can be mitigated using specific methods. After an initial learning curve, a 3D data set can be consistently acquired in <1 min. For analysis of a VSD, a four-beat full-volume acquisition usually provides the best balance between volume rate, resolution, and stitch artifact. The 3D full-volume data set is then cropped using standard cropping or iCrop to demonstrate the important information.


Standard cropping is a method in which the 3D echocardiographic data set is positioned within a cube. Any or all six sides of the cube (X, Y or Z planes) can be positioned to remove extraneous 3D echocardiographic data to reveal the important 3D echocardiographic information. These crop planes have their orientations fixed in the X, Y, and Z planes. If required, an additional cropping tool called adjustable plane cropping can be added. This tool provides for the placement of one or more oblique crop planes. Together, XYZ and adjustable plane cropping constitute what we refer to as standard cropping.


The iCrop mode, a more recently developed method of cropping, allows the placement of an adjustable crop box within the 3D data set. The data within the crop box are included, and the data outside the box are cropped away. The operator adjusts the size, position, and rotation of the box and selects from which of the six sides of the box to view the 3D data set. Although this method is rapid and easy to apply, the sonographer must be cognizant of the fact that the resolution is dependent on the viewing direction because of the inherent different resolutions with axial > lateral > elevation plane spatial resolution.


Narrow-sector live 3D mode, like full-volume 3DE, is an acquisition mode. When using the default one-beat acquisition setting, it provides live actual real-time imaging, with no reconstruction required. A narrow wedge-shaped 3D sector of adjustable size, position, and view aspect is manipulated to demonstrate the important information. The sector size and position are adjusted using lateral and elevation plane controls for width and position. The view aspect is controlled using the 3D rotation tool. When done properly, little or no cropping is required for L3D images. L3D usually provides the highest volume rate and overall resolution of the 3D acquisition modalities. However, it has a relatively small sector volume and must be adjusted by the sonographer to include the structure of interest entirely within the 3D sample volume.


Both of the 3D acquisition modes (full-volume 3D and L3D) and the cropping modes (standard and iCrop) are selected and controlled with typical echocardiographic machine touchpad, dial, and trackball controls. Multiple echocardiographic parameter adjustments can be made at the time of acquisition and during postacquisition processing, including gain, contrast, smoothing, image algorithms, color adjustments, color depth shading adjustments, and rotation throughout 360° in all planes. Image manipulation, cropping, and quantification can be performed on the iE33 ultrasound platform or on the digital storage and review station using dedicated software (Philips Medical Systems).


Three-Dimensional Echocardiographic VSD Protocols


Transthoracic 3D images were obtained from the standard apical four-chamber view. Transesophageal 3D images were obtained from the four-chamber midesophageal view.


In protocol 1, FVstd ( Figure 1 ), full-volume 3D data sets were acquired using a four-beat setting. The data sets were then processed using the iE33 echocardiographic machine at the time of acquisition or on a desktop workstation using Phillips Q-LAB image-processing software at a later time. The front of the volume was cropped away using the green (z axis) plane adjustment, revealing a four-chamber view. Using the red (y axis) plane adjustment, the right and left sides of the volume were cropped to isolate the interventricular septum. The volume was then rotated to visualize the right or left side of the interventricular septum. Finally, the anterior aspect of the sample was filled in using the green plane adjustment to fill in a complete en face view of the interventricular septum.




Figure 1


Three-dimensional full-volume with standard cropping protocol (FVstd): (1) Acquire a 3D full-volume sample using the four-beat acquisition mode from the apical or midesophageal four-chamber view. Open the image using Q-LAB software. (2) Crop the volume from the front using the green plane adjustment until the center of the VSD is seen. (3) Crop both sides of the volume using the red plane adjustment to isolate the interventricular septum. (4) Rotate the volume to view either the right or left face of the interventricular septum. (5) Fill in the front of the volume using the green plane adjustment. Note: In the case of an oblique septum, additional cropping planes using the adjustable crop plane tool can be activated and oriented to crop in any 3D plane from both sides of the ventricular septum. LV , Left ventricle; RA , right atrium; RV , right ventricle.


In protocol 2, FViCrop ( Figure 2 ), acquisition of iCrop 3D images involved first setting up a full-volume sample using a four-beat acquisition setting in the midesophageal or apical four-chamber view. Activating the iCrop function results in the appearance of a three-panel display with two 2D orthogonal (frontal and lateral) planes and the 3D volume above them in the center. The adjustable crop box appears superimposed on the two 2D images. The sonographer then adjusts the size, position, and rotation of the crop box in both the right and left 2D panels to include the VSD and important surrounding anatomy. Next, the side of the crop box that demonstrates the right or left en face view of the interventricular septum is selected. The 2D images are deactivated to display only the 3D image, which is then acquired and stored in Digital Imaging and Communications in Medicine format.




Figure 2


Three-dimensional full-volume with iCrop protocol (FViCrop): (1) Orient the transducer in an apical or midesophageal four-chamber view and activate 3D full-volume with a four-beat acquisition. (2) Activate iCrop. The 2D reference images and 3D echocardiographic volume will appear. (3) Adjust the size, position, and rotation of the iCrop box in both right and left 2D reference panels to include the VSD and important surrounding anatomy. Toggle the view selector to visualize the right or left face of the interventricular septum. (4) Deactivate the reference images to display only the 3D image and acquire the sample. LA , Left atrium; LV , left ventricle; RA , right atrium; RV , right ventricle.


In protocol 3, L3D ( Figure 3 ), volumes were acquired by first orienting the transducer to obtain a 2D four-chamber view. Narrow-sector live 3D mode was then activated. The lateral sector width was then decreased to isolate the interventricular septum. The volume was rotated left or right to view one face of the interventricular septum. Finally, the elevation width was increased until the entire ventricular septum and VSD were visualized.




Figure 3


Live 3D protocol (L3D): (1) Orient the transducer in an apical or midesophageal four-chamber view with the ventricular aligned vertically. Next, activate live 3D. (2) Decrease the lateral sector width to isolate the interventricular septum. (3) Rotate the volume to visualize the right or left face of the interventricular septum. (4) Increase the elevation width to include the entire ventricular septum and VSD. LV , Left ventricle; RV , right ventricle.


Assessment of Feasibility


Assessment of the quality of the 3D images of the VSD was done by scoring each volume on a scale from 0 to 3: 0 = no visualization of the VSD; 1 = the VSD can be identified, but the entire defect is not seen, and visualization of the surrounding anatomy is inadequate; 2 = all borders of the VSD are visualized, but the surrounding anatomy is not adequately shown; and 3 = the entire VSD and all the important surrounding anatomy are adequately visualized ( Figure 4 ). All three protocols were applied in each patient. The highest rated loop was used for quality scoring for each of the three respective acquisition protocols. We defined feasibility for each of the three respective 3D echocardiographic modalities as the percentage of cases with en face views of the VSD having image quality grades of 2 or 3.




Figure 4


Image scoring: 0 = no visualization of the VSD; 1 = the VSD can be identified, but the entire defect is not seen, and visualization of the surrounding anatomy is inadequate; 2 = all borders of the VSD are visualized, but the surrounding anatomy is not adequately shown; and 3 = the entire VSD and all the important surrounding anatomy are adequately visualized.


Assessment of Accuracy


The 3D echocardiographic findings were compared with each patient’s definitive clinical diagnosis on the basis of intraoperative inspection, cardiac catheterization with angiography, cardiac MRI or magnetic resonance angiography, cardiac computed tomographic angiography, and standard 2D and color Doppler echocardiography. Accuracy was defined as the percentage of cases in which the correct type, subjective assessment of size, and the location of the VSD were all correctly diagnosed using each of the 3D echocardiographic modality en face views.


Statistical Analysis


Continuous data are reported as mean ± SD. Categorical variables are expressed as proportions. A scoring system was used to evaluate the feasibility of each of three modalities with regard to obtaining a diagnostic-quality 3D echocardiographic en face view. The Pearson χ 2 test with the Fisher exact test was used to determine the statistical difference between any two of the three modalities. A 2 × 2 classification table was used to calculate accuracy. Statistical significance was declared when computed P values from two-sided tests were <.05. Analyses were performed using IBM SPSS version 20 for Windows (IBM, Inc, Somers, NY).


Ethics


This study was in compliance with all institutional guidelines regarding research conduct, including patient confidentiality and institutional research board approval. All patients or parents provided written informed consent for 3D transesophageal echocardiographic studies.




Results


Clinical Features


Sixty-four patients underwent TTE, 31 female and 33 male, ranging in age from 1 day to 77 years (mean, 1.8 ± 9.4 years) and body weight from 3 to 72 kg (mean, 11.3 ± 19.7 kg). Thirty-six patients underwent TEE, 19 female and 17 male, ranging in age from 4.8 to 77 years (mean, 20.9 ± 13.9 years) and body weight from 20 to 92 kg (mean, 42 ± 18.4 kg). The frequency of each of the VSD types was as follows: 34 muscular, 32 perimembranous, 18 malaligned, 11 inlet, four outlet, and one acquired. There were no complications related to either imaging modality.


Feasibility


Analysis of the image quality of each of the three methods for each VSD was performed in all 100 patients and scored from 0 to 3 on the aforementioned scale. The success rate for obtaining a good diagnostic-quality en face image, defined as an image quality score ≥ 2, was high for each of the three protocols: 100% for FViCrop, 97% for FVstd, and 94% for L3D. The higher success rates for both FViCrop and FVstd versus L3D were both statistically significant ( P < .001 for both comparisons). The higher success rate for FViCrop versus FVstd was not statistically significant ( P = .076). In no patient did we encounter a score of 0 for any of the three methods. The FViCrop protocol produced the highest overall score (mean, 2.91 ± 0.29), followed by FVstd (mean, 2.87 ± 0.44), and L3D (mean, 2.46 ± 0.67). See Table 1 for a summary of these results.