The Value of Three-Dimensional Echocardiography Derived Mitral Valve Parametric Maps and the Role of Experience in the Diagnosis of Pathology




Background


Accurate segmental mitral valve (MV) analysis is essential for surgical planning. Although real-time three-dimensional (3D) transesophageal echocardiography has improved the ability to visualize the MV, accurate localization of MV pathology from 3D transesophageal echocardiographic images still remains experience dependent. Three-dimensional parametric maps of the MV obtained from these images further simplify the visualization of MV anatomy. The aims of this study were to examine whether 3D parametric maps of the MV could improve the diagnostic accuracy in localizing pathology and to determine their usefulness for readers with different levels of training.


Methods


Five novice (American Society of Echocardiography [ASE] level 2), three intermediate-level (ASE level 3; <500 MV cases), and two expert (ASE level 3; >500 MV cases) readers interpreted MV segmental anatomy in 50 patients (30 with degenerative MV disease, 20 with normal MVs). All readers reviewed two-dimensional and 3D transesophageal echocardiographic and 3D parametric maps at sequential weekly sessions. The results were compared with surgical findings.


Results


Expert readers were the most accurate irrespective of image type. Novice readers were the least accurate and commonly misinterpreted P2 and P3 scallops. Their accuracy was highest when interpreting 3D parametric maps (from 87% with two-dimensional transesophageal echocardiography to 92%). Intermediate readers’ accuracy fell between the other two groups irrespective of image type and showed no change with the use of parametric maps.


Conclusions


This is the first study to show that the interpretation of 3D parametric maps improves the accuracy of localization of MV pathology by novice readers. Therefore, parametric maps should be used routinely by less experienced readers during the assessment of degenerative MV disease.


Accurate preoperative assessment of the mitral valve is critical in the surgical management of patients with severe mitral regurgitation. This information determines whether patients should undergo valve repair or replacement, which has implications in terms of timing of surgery and long-term morbidity and mortality. Although two-dimensional (2D) transthoracic echocardiography and transesophageal echocardiography (TEE) are part of the standard assessment of mitral valve anatomy, three-dimensional (3D) transthoracic echocardiography and TEE have been shown to facilitate the understanding of more complex abnormalities of the mitral valve apparatus and individual scallop identification. However, even with 3D echocardiography, accuracy in diagnosing mitral valve pathology is still not perfect. Specifically, mitral valve pathology involving the posteromedial and less commonly the anterolateral commissures is often missed or erroneously interpreted.


Although 3D echocardiography provides “realistic” images of the mitral valve, training is required to differentiate normal anatomy from dropout artifacts. The majority of studies demonstrating improvement in diagnosing mitral valve pathology using 3D echocardiography have included readers with significant 2D (and presumably 3D) echocardiographic experience. It is plausible that readers without sufficient experience may not adequately harness the information available from 3D echocardiography.


Three-dimensional parametric maps transform the 3D images of the mitral valve into color-encoded topographic displays of mitral valve anatomy, in which the color gradations indicate the distance of the leaflet from the mitral annular plane toward the left atrium. Thus, in mitral valve prolapse, the location of maximal prolapse can be easily appreciated by the presence of orange color on the affected leaflet. We hypothesized that these 3D parametric maps could improve the interpretation of mitral valve pathology even in inexperienced readers. Accordingly, the aims of this study were (1) to examine whether 3D parametric maps could improve diagnostic accuracy in localizing mitral valve pathology and (2) to determine to what extent reader’s level of training affects the utility of 3D parametric maps on the sensitivity, specificity, and accuracy of the TEE-based interpretation of mitral valve pathology.


Methods


Patient Population


We retrospectively identified 50 patients (mean age, 59 ± 15 years; 37 men) who had undergone clinically indicated TEE and had standard 2D and 3D transesophageal echocardiographic studies, of whom 20 had normal mitral valves and 30 had severe mitral regurgitation requiring surgical intervention. The localization of mitral valve pathology in the patients with mitral regurgitation was confirmed by surgical exploration. All 50 patients were in normal sinus rhythm. The institutional review board approved this study.


Image Acquisition and Formats


Two-dimensional TEE was performed according to a standard protocol using the iE33 ultrasound system (Philips Medical Systems, Andover, MA). The following midesophageal, zoomed views of the mitral valve were obtained: four-chamber view (0°), bicommissural view (45°), two-chamber view (90°), and long-axis view (125°). These 2D images were exported for interpretation as digital cine loops.


Real-time 3D echocardiography (RT3DE) of the mitral valve was performed using a fully sampled matrix transesophageal echocardiographic transducer (X7-2t). Initially, gain settings were optimized using the narrow-angle acquisition mode, which allows imaging of a pyramidal volume of approximately 30° × 60° without the need for electrocardiographic gating. Zoomed real-time 3D echocardiographic images of the entire mitral valve were then acquired in a single cardiac cycle, resulting in frame rates between 5 and 35 Hz (mean, 9 ± 5 Hz). Acquisition of 3D data sets was repeated at least three times to ensure optimal image quality.


The 3D images were reviewed offline on an Xcelera workstation (Philips Medical Systems) by a trained echocardiographer. To improve visualization, pyramidal data sets were cropped along designated x-y-z axes or using a manually positioned cropping plane of choice. The images were then oriented from the left atrial perspective with the aortic valve at the 12-o’clock position and exported for interpretation as digital cine loops.


Parametric maps of the mitral valve were generated using commercial software (MVQ, QLAB version 8.0; Philips Medical Systems). Initially, the end-systolic frame was defined as the second-to-last frame before the initiation of mitral valve opening. Then, a long-axis view of the mitral apparatus was used to determine anterior, posterior, anterolateral, and posteromedial annular coordinates ( Figure 1 ). The annulus was manually outlined by defining annular points in multiple planes rotated around the axis perpendicular to the mitral annular plane. The annulus was then segmented to identify leaflet geometry and coaptation points by manually tracing the leaflets in multiple parallel long-axis planes spanning the annulus from commissure to commissure. The reconstructed mitral valve was subsequently displayed as a color-encoded 3D surface-rendered image representing a topographic map of the mitral leaflets ( Figure 1 D). This parametric map of the mitral valve was oriented as viewed from the left atrium with the aortic annulus at the 12-o’clock position and exported as a still image.




Figure 1


Mitral valve quantification. From the 3D transesophageal images, three orthogonal mitral annular images were displayed at end-systole (A–C) . Four annular key MA reference points (anterolateral and posteromedial hinge points of the leaflet insertion, A ; anterior and posterior points, B ) as well as the aortic annulus were used for initialization. Subsequently, with the three orthogonal planes locked in, 14 markers around the annulus at leaflet insertion were initialized in seven rotational planes (C) . Finally, a parametric model of the mitral valve was produced (D) . A , Anterior; AL , anterolateral; Ao , aortic valve; P , posterior; PM , posteromedial.


Interpretation Definitions


We used the mitral valve nomenclature proposed by Carpentier, which divides the posterior leaflet into three scallops: lateral (P1), middle (P2), and medial (P3). The anterior leaflet was divided into three segments: lateral third (A1), middle third (A2), and medial third (A3).


On 2D TEE, a scallop was considered prolapsed if any portion moved above the mitral annular plane during systole. Flail scallop was defined as a free leaflet edge or chordae tendineae detected in the left atrium during systole. Leaflet perforation was defined as the presence of normal coaptation of the leaflets with regurgitant jet originating from the leaflet body, in a location separate from the coaptation line.


For RT3DE, a scallop was considered prolapsed if any of the following features were noted: billowing or hammock shape, coaptation gap with correspondent or adjacent scallops, straight lines from the annulus to the free border delineating the border between a prolapsing and a nonprolapsing scallop, and image of negative relief, depression, or fingerprint corresponding to one or several scallops. Flail scallop on RT3DE was diagnosed as a flail shape or line between the annulus and the free border of the scallops that was directed toward the left atrium. Perforation on RT3DE was diagnosed as a discontinuity in the leaflet surface.


The definition of prolapse from the parametric maps was the presence of a coaptation gap between corresponding segments or billowing represented by color gradations on the leaflet segments ( Figure 2 ). Because of limitations of the parametric maps, flail segments could not be identified, and consequently segmental prolapse was considered a correct answer.




Figure 2


Three-dimensional parametric maps of the mitral valve demonstrating (A) normal anatomy, (B) P3 flail scallop, (C) P2 and P3 prolapse, and (D) Barlow’s disease.


Study Interpreters


Five cardiologists and five cardiology fellowship trainees blinded to the surgical findings were asked to interpret mitral valve segmental anatomy. Expertise was determined according to the American Society of Echocardiography (ASE) standards of training, resulting in five novice (ASE level 2), three intermediate-level (ASE level 3; <500 mitral valve cases), and two expert (ASE level 3; >500 mitral valve cases) readers. Readers initially interpreted the 2D transesophageal echocardiographic images, then after 1 week the 3D transesophageal echocardiographic images, and then after another week the 3D parametric maps. Each mitral valve scallop or segment was classified using the above definitions.


Surgical Inspection and Validation


Intraoperatively, the surgeons described the anatomy of the mitral valve using the same Carpentier classification as the readers. The surgeons had access to the clinical echocardiography reports but were unaware of the interpretations made for this study.


Statistical Analysis


The sensitivity, specificity, and accuracy of the evaluation of the involved scallops and segments were calculated for each image type and reader level using the surgical findings as the reference. Reproducibility was calculated as the absolute rate of agreement between two readers for each scallop in each modality.




Results


We were able to obtain interpretable parametric maps in all 50 patients, the creation of which required on average 10 min. Figures 3 through 5 show examples of images including parametric maps obtained in three patients with differing anatomy. Table 1 shows the anatomic (surgical inspection) characteristics of the 30 surgical patients. Half of the patients ( n = 15) had complex pathologies involving more than one mitral valve scallop or segment or bileaflet pathology.




Figure 3


Two-dimensional transesophageal echocardiographic zoomed image of a normal mitral valve in the midesophageal, four-chamber (A) , and two-chamber (B) views. Three-dimensional transesophageal echocardiographic image in zoomed mode of the mitral valve as viewed from the left atrium (C) . Three-dimensional parametric map of the mitral valve obtained from 3D TEE as view from the left atrium (D) .



Figure 4


Example of images of the mitral valve obtained in a patient with isolated P2 prolapsed scallop, shown in the same format as in Figure 2 .



Figure 5


Example of images of the mitral valve obtained in a patient with an isolated flail A2 segment, shown in the same format as in Figure 2 .


Table 1

Anatomic characteristics of the mitral valve in the study patients on the basis of surgical inspection








































Characteristic n
Isolated A1 0
Isolated A2 2
Isolated A3 0
Isolated P1 1
Isolated P2 8
Isolated P3 3
Posterior leaflet (>1 scallop) 7
Anterior leaflet (>1 scallop) 1
Anterior and posterior leaflets 8
Associated lesion
Chordal rupture 16


Sensitivity, specificity, and accuracy of the three methodologies for the overall and leaflet lesions are reported in Table 2 . As expected, we found that novice readers had the lowest sensitivity, specificity, and accuracy, while expert readers had the highest values, regardless of image type used to interpret the mitral valve anatomy. Both novice and intermediate readers had greater difficulty interpreting 3D images compared with 2D images. However, in novice readers, 3D parametric maps improved the sensitivity, specificity, and accuracy, surpassing the 2D image values, with particular improvements in interpreting P2 and P3 scallop involvement ( Table 3 ). In contrast, the accuracy of intermediate readers did not improve with the use of 3D parametric maps to values greater than those achieved using 2D images, except in the interpretation of P2 and P3 scallop morphology. For these scallops, accuracy and sensitivity improved but at the expense of specificity. On the other hand, expert readers were very proficient at accurately localizing mitral valve lesions regardless of the type of images used.



Table 2

Detection of mitral leaflet pathology with the three echocardiographic techniques




























































































































Sensitivity (%) Specificity (%) Accuracy (%)
Novice Intermediate Expert Novice Intermediate Expert Novice Intermediate Expert
AML + PML
2D TEE 66 82 99 94 99 100 87 95 100
3D TEE 57 72 100 94 97 100 84 91 100
Parametric maps 76 94 97 97 92 98 92 92 99
AML
2D TEE 65 78 100 93 99 100 88 95 100
3D TEE 51 60 100 95 99 100 87 92 100
Parametric maps 67 68 96 97 98 98 92 93 98
PML
2D TEE 66 85 99 94 99 100 85 94 100
3D TEE 60 78 100 92 95 100 82 90 100
Parametric maps 81 96 100 96 91 100 92 92 100

Sensitivity, specificity, and accuracy are presented by reader experience and imaging methodology, with overall leaflet values followed by anterior mitral leaflet (AML) and posterior mitral leaflet (PML) values.


Table 3

Detection of mitral scallop pathology with the three echocardiographic techniques





































































































































































































































Sensitivity (%) Specificity (%) Accuracy (%)
Novice Intermediate Expert Novice Intermediate Expert Novice Intermediate Expert
A1
2D TEE 60 71 100 99 99 98 93 95 98
3D TEE 37 57 100 95 98 100 87 93 100
Parametric maps 66 100 100 96 95 100 92 96 100
A2
2D TEE 71 85 96 86 98 99 83 96 98
3D TEE 64 61 100 91 98 100 85 95 100
Parametric maps 73 85 96 97 96 100 92 90 99
A3
2D TEE 60 75 100 94 99 99 88 95 99
3D TEE 45 63 100 97 99 100 89 93 100
Parametric maps 60 88 94 97 89 100 91 89 99
P1
2D TEE 59 77 100 96 98 100 86 93 100
3D TEE 54 74 100 95 95 100 84 89 100
Parametric maps 77 87 100 97 94 100 92 92 100
P2
2D TEE 72 87 98 90 100 100 82 87 99
3D TEE 67 81 100 88 93 100 78 81 100
Parametric maps 86 100 100 93 90 100 90 100 100
P3
2D TEE 63 81 100 96 98 100 88 81 100
3D TEE 52 78 100 93 97 100 83 78 100
Parametric maps 77 97 100 98 88 100 93 97 100