The Comprehensive Transthoracic Echocardiographic Examination



The Comprehensive Transthoracic Echocardiographic Examination


Pei-Hsiu Huang



Getting Started

A quality echocardiogram study starts with the setup (Fig. 2-1).











Figure 2-1. Echocardiogram setup.

Additional Imaging Considerations














Acoustic Windows
The optimal acoustic window allows acquisition of bright and clear images. Poor image quality will affect the accuracy of 2D measurements and Doppler quality.


  • Acquire the “big picture” first
  • Move the transducer along and between the intercostal spaces
  • Keep the transducer movements small
Respiration
Heart position changes with respiration.


  • Acquire most parasternal and apical views at held end-expiration
  • Acquire the apical two-chamber and subcostal views at held end-inspiration
Transducer Pressure
Firm pressure is necessary for good transducer contact. However, applying too much pressure, especially on pain sensitive bony surfaces, will cause the patient pain!


  • When imaging through areas of increased subcutaneous tissue (e.g., adipose tissue or under breasts), applying increased pressure often improves the image quality
  • Slightly reduce pressure, once a good quality image is found
  • Release the transducer pressure when switching intercostal spaces
Ultrasound Gel
A medium used to conduct sound waves between the patient and the ultrasound transducer


  • Use plenty of gel!
  • Reapply when the gel on the patient spreads to a thin layer



Adjusting the Image


2D Gain

The intracardiac blood pool should be as dark as possible without losing definition of the cardiac structures (Fig. 2-2).






Figure 2-2A. Optimum gain.






Figure 2-2B. Over-gained.



Image Depth

The depth should be set to extend approximately 1 to 2 cm beyond the cardiac boundary most distant from the transducer to ensure that none of the structures are cut off (Fig. 2-3).






Figure 2-3A. Optimum depth.






Figure 2-3B. Too deep.

Note: Some laboratories use standard default depths to facilitate comparison of serial examinations. In this case, acquire images at the default depth as well as at the appropriate depth.



Color Doppler






Figure 2-4. Color Doppler sampling region with Nyquist limit (inset).

Adjust the color Doppler sampling region to include only the structures of interest in order to avoid decreases in temporal resolution and color quality. Keep the default Nyquist limit at 50–60 cm/s (Fig. 2-4).


Color Gain

This can be calibrated by moving the color box into the extracardiac space and increasing Doppler gain until there is visible noise. Slowly decrease gain until the noise first disappears.


Frequency

Start imaging with the transducer set at 1.7/3.4 MHz (transmitted/received; second harmonic imaging) (Fig. 2-5). Select a higher (for near-field imaging) or lower (for deeper penetration) frequency to optimize the image quality.






Figure 2-5. Transducer setting.


Focus






Figure 2-6. The focus of the ultrasound beam is shown by the arrow (inset).

Try adjusting the ultrasound focus if an unclear image or artifact is encountered (Fig. 2-6).



Pulsed-wave Doppler Sample Volume Sizes


An inappropriate sample size may contaminate the Doppler acquisition and nullify spectral Doppler’s greatest advantage of range specificity.


Spectral Gain

The background should be dark and the signal bright to ensure it is not under-gained. Over-gained images may result in over-estimation of blood flow velocities. Measurements should be taken of the modal velocities (bright envelope) and not the spectral broadening (“feathery spray”) especially seen in over-gained or poor quality Doppler (Fig. 2-7).






Figure 2-7. (A) Pulsed-wave Doppler with appropriate gain and (B) over-gained.



Sweep Speed

Generally set at 50 mm/s for M-mode and spectral Doppler with normal heart rates. Set the sweep speed at 100 mm/s to get measurements at a high temporal resolution (Fig. 2-8).






Figure 2-8. PLAX view with M-mode through the mitral valve leaflets and PSAX view with spectral Doppler of the RVOT with sweep speed at 50 mm/s (A, B) and 100 mm/s (C, D).



The Views


Parasternal Long Axis (PLAX)





image







Figure 2-9. Parasternal long-axis (PLAX) view. N, transducer notch.



Oct 20, 2016 | Posted by in CARDIOLOGY | Comments Off on The Comprehensive Transthoracic Echocardiographic Examination

Full access? Get Clinical Tree

Get Clinical Tree app for offline access