TABLE 1.1 Transducer frequencies are measured in MHz (1,000,000 cycles/s). Doppler signal frequencies are measured in KHz (1000 cycles/s). Image resolution is greatest (about 1 mm) with a shorter wavelength (higher frequency). Depth of tissue penetration is greatest with a longer wavelength (lower frequency). A log scale is used for dB. On the dB scale, 80 dB represents a 10,000-fold and 40 dB indicates a 100-fold increase in amplitude. TABLE 1.2 TABLE 1.3 Transducer characteristics and configuration Most cardiac transducers use phased array of piezoelectric crystals Transthoracic (adult and pediatric) Non-imaging CW Doppler 3 D echocardiography TEE Intracardiac Each transducer type is optimized for a specific clinical application. More than one transducer may be needed for a full examination. A higher frequency transducer provides improved resolution but less penetration. Doppler signals are optimal at a lower transducer frequency than used for imaging. A larger aperture allows a more focused beam. A smaller aperture allows improved transducer angulation on TTE imaging.
Principles of Echocardiographic Image Acquisition and Doppler Analysis
Basic Principles
Key Points
Ultrasound Waves
Key Points
Transducers
Definition
Examples
Clinical Implications
Frequency (f)
The number of cycles per second in an ultrasound wave: f = cycles/s = Hz
Different transducer frequencies are used for specific clinical applications, because the transmitted frequency affects ultrasound tissue penetration, image resolution, and the Doppler signal.
Velocity of propagation (c)
The speed that ultrasound travels through tissue
The average velocity of ultrasound in soft tissue about 1540 m/s.
The velocity of propagation is similar in different soft tissues (blood, myocardium, liver, fat, etc.) but is much lower in lung and much higher in bone.
Wavelength (λ)
Wavelength is shorter with a higher frequency transducer and longer with a lower frequency transducer.
Amplitude (dB)
Height of the ultrasound wave or “loudness” measured in decibels (dB)
A very wide range of amplitudes can be displayed using a gray scale display for both imaging and spectral Doppler.
Definition
Examples
Clinical Implications
Acoustic impedance (Z)
A characteristic of each tissue defined by tissue density (ρ) and propagation of velocity (c) as: z = ρ × c
Lung has a low density and slow propagation velocity, whereas bone has a high density and fast propagation velocity. Soft tissues have smaller differences in tissue density and acoustic impedance.
Ultrasound is reflected from boundaries between tissues with differences in acoustic impedance (e.g., blood versus myocardium).
Reflection
Return of ultrasound signal to the transducer from a smooth tissue boundary
Reflection is used to generate 2D cardiac images.
Reflection is greatest when the ultrasound beam is perpendicular to the tissue interface.
Scattering
Radiation of ultrasound in multiple directions from a small structure, such as blood cells
The change in frequency of signals scattered from moving blood cells is the basis of Doppler ultrasound.
The amplitude of scattered signals is 100 to 1000 times less than reflected signals.
Refraction
Deflection of ultrasound waves from a straight path due to differences in acoustic impedance
Refraction is used in transducer design to focus the ultrasound beam.
Refraction in tissues results in double image artifacts.
Attenuation
Loss in signal strength due to absorption of ultrasound energy by tissues
Attenuation is frequency dependent with greater attenuation (less penetration) at higher frequencies.
A lower frequency transducer may be needed for apical views or in larger patients on transthoracic imaging.
Resolution
The smallest resolvable distance between two specular reflectors on an ultrasound image
Resolution has three dimensions—along the length of the beam (axial), lateral across the image (azimuthal), and in the elevational plane.
Axial resolution is most precise (as small as 1 mm), so imaging measurements are best made along the length of the ultrasound beam.
Key Points
Ultrasound Imaging
Principles
Key Points
Definition
Examples
Clinical Implications
Type
Transmission frequency
The central frequency emitted by the transducer
Transducer frequencies vary from 2.5 MHz for transthoracic echo to 20 MHz for intravascular imaging.
Power output
The amount of ultrasound energy emitted by the transducer
An increase in transmitted power increases the amplitude of the reflected ultrasound signals.
Excessive power output may result in bioeffects measured by the mechanical and thermal indexes.
Bandwidth
The range of frequencies in the ultrasound pulse
Bandwidth is determined by transducer design.
A wider bandwidth allows improved axial resolution for structures distant from the transducer.
Pulse (or burst) length
The length of the transmitted ultrasound signal
A higher frequency signal can be transmitted in a shorted pulse length compared with a lower frequency signal.
A shorter pulse length improves axial resolution.
Pulse repetition frequency (PRF)
The number of transmission-receive cycles per second
The PRF decreases as imaging (or Doppler) depth increases because of the time needed for the signal to travel from and to the transducer.
Pulse repetition frequency affects image resolution and frame rate (particularly with color Doppler)
Focal depth
Beam shape and focusing are used to optimize ultrasound resolution at a specific distance from the transducer
Structures close to the transducer are best visualized with a short focal depth, distant structures with a long focal depth.
The length and site of a transducer’s focal zone is primarily determined by transducer design, but adjustment during the exam may be possible.
Aperture
The surface of the transducer face where ultrasound is transmitted and received
A small non-imaging CW Doppler transducer allows optimal positioning and angulation of the ultrasound beam.
Imaging Artifacts
Key Points
Doppler
Key Points
Stay updated, free articles. Join our Telegram channel
Full access? Get Clinical Tree
Principles of Echocardiographic Image Acquisition and Doppler Analysis
1
The distance between ultrasound waves:
λ = c/f = 1.54/f (MHz)