pressure alarms Eva Nourbakhsh MD

Peak pressure with normal plateau pressure


 


  • Monitor will show a peak pressure >40 cm H2O and a plateau pressure <30 cm H2O
  • Possible causes: image secretions, bronchospasm, kinked tube
  • Solution: suction, give bronchodilators, check tube

image Peak pressure with image plateau pressure


 


  • Monitor will show a peak pressure >40 cm H2O and a plateau pressure >30 cm H2O
  • Possible causes: pneumothorax, ARDS, atelectasis, pulmonary edema, ETT in main bronchus
  • Solution: treat the cause. Also try to image pressures by image VT by 1–2 ml/kg IBW

image Plateau pressure


 