Predicting Paravalvular Regurgitation Following Transcatheter Valve Replacement: Utility of a Novel Method for Three-Dimensional Echocardiographic Measurements of the Aortic Annulus




Background


Studies evaluating three-dimensional echocardiographic (3DE) annular sizing for balloon-expandable transcatheter aortic valve replacement (TAVR) are limited. In this study, a retrospective analysis of transesophageal echocardiographic images was performed to assess the feasibility of multiplanar measurements of annular dimensions by the novel off-label use of commercially available 3DE software and correlate annular sizing with severity of paravalvular regurgitation (PVR).


Methods


Intraprocedural transesophageal echocardiography was performed in 58 patients undergoing TAVR for severe, symptomatic aortic stenosis. Off-label use of commercially available software was used to measure transesophageal 3DE volumes. Pre-TAVR annular linear dimensions included two-dimensional echocardiographic sagittal diameter and 3DE measurements of minimal diameter, maximal diameter (MaxDiam), and the average or mean diameter. Three-dimensional echocardiographic average annular diameter derived from annular perimeter (AveAnnDiam P ) and average annular diameter derived from annular area (AveAnnDiam A ) were calculated. A cover index was calculated using each measurement. Short-axis PVR color jet areas were summed after deployment and at the end of study.


Results


Two-dimensional echocardiographic sagittal diameter was significantly smaller than 3DE MaxDiam ( P < .0001) and AveAnnDiam P ( P = .017), significantly larger than 3DE minimal diameter ( P < .0001), and not significantly different from 3DE mean diameter ( P = .36) and AveAnnDiam A ( P = .38). There was a linear relationship between all 3DE annular measurements and immediate post-TAVR PVR area ( P ≤ .0001), with the largest R 2 value (0.48) for 3DE MaxDiam, 3DE AveAnnDiam P , and 3DE AveAnnDiam A . The largest areas under the curve to detect greater than mild PVR were for 3DE AveAnnDiam P cover index (0.772) and 3DE AveAnnDiam A cover index (0.769). Intraclass correlation coefficients for interobserver and intraobserver variability were high for 3DE AveAnnDiam P and 3DE AveAnnDiam A and lower for 3DE MaxDiam.


Conclusions


Using a novel approach to 3DE annular measurements, 3DE AveAnnDiam P and 3DE AveAnnDiam A can be reliably measured and correlate best with post-TAVR PVR area. New sizing algorithms with 3DE measurements should be developed.


Although transcatheter aortic valve replacement (TAVR) has emerged as an alternative to surgical aortic valve replacement for patients with severe aortic stenosis who are at “high risk” or deemed inoperable, recent studies have shown an association between postprocedural paravalvular regurgitation (PVR) and increased late mortality. Some degree of PVR has been reported in 80% to 96% of patients undergoing TAVR. In most cases, the degree of regurgitation is mild; however, moderate and severe aortic regurgitation are found in 7% to 24% of patients. A small cover index or large annular size, annular eccentricity, the degree and distribution of leaflet calcifications, and malpositioning of the valve have all been identified as risk factors for PVR. A recent meta-analysis confirmed that implantation depth, valve undersizing, and Agatston calcium score were the important predictors of PVR after TAVR. Numerous studies have suggested that three-dimensional (3D) assessment of the aortic annulus using multislice computed tomography can accurately assess the noncircular anatomy of the aortic annulus and improve annular sizing of the transcatheter heart valves (THVs) by measuring the perimeter or area of the annulus. Three-dimensional transesophageal echocardiographic (TEE) imaging can also accurately measure the annular aortic annulus. Both techniques rely primarily on direct planimetry of the annulus in the transverse plane, which has inherent limitations. The aims of this study were to evaluate the feasibility and reproducibility of real-time 3D measurements of the aortic annulus utilizing a novel off-label use of commercially available software and to use this method to predict PVR after the placement of a balloon-expandable THV.


Methods


Patients and Procedure


Data from 58 (nonconsecutive) patients treated between May 2007 and November 2011 with Edwards SAPIEN THVs (Edwards Lifesciences, Irvine, CA) as part of the Placement of Aortic Transcatheter Valves and Placement of Aortic Transcatheter Valves 2 trials were retrospectively analyzed. After implantation of the THV, reballooning or postdilatation of the implanted valve was performed at the discretion of the operator, typically on the basis of the assessment of PVR severity.


Echocardiographic Data


All TEE studies were performed using an iE33 xMATRIX system using the X7-2t TEE transducer (Philips Medical Systems, Andover, MA). Intraprocedural two-dimensional echocardiographic (2DE) and 3D echocardiographic (3DE) images were selected for evaluation. Two-dimensional echocardiographic sagittal diameter (SagDiam) measurements were made from the long-axis view of the aorta, with biplane imaging used to ensure that the largest diameter was measured ( Figure 1 ). Frequently, a lower TEE window was required to avoid acoustic shadowing of the anterior annulus by the bulky calcified aortic cusps.




Figure 1


Measurement of 2DE SagDiam of the aortic annulus using simultaneous multiplane imaging. (A) Short-axis view and (B) simultaneous orthogonal view obtained by positioning the dashed yellow line to bisect the maximum dimension of the aortic root in this imaging plane. The orthogonal long-axis view (B) should bisect the maximum diameter of the annulus. The red double arrow shows the level of the cusp hinge points for measurement of the sagittal plane annulus diameter.


Color Doppler short-axis imaging of the region just apical to the proximal edge of the THV was used for the evaluation of PVR ( Figure 2 ). Simultaneous multiplane imaging was used to enable consistent imaging of this plane using the long-axis view (120°–140°) as the primary plane with the orthogonal short-axis view placed in the left ventricular outflow tract just beneath the skirt of the THV. The short-axis areas of the jets were measured and the sum of the jet areas was used as the total PVR area. Central regurgitation resulting from malcoaptation of leaflets created by the guidewire was distinguished from PVR by locating the catheter on real-time images and correlating it with the associated color Doppler signal, changing wire position to better identify the regurgitant jet, or removing the wire completely.




Figure 2


Measurement of PVR area using simultaneous multiplane imaging. (A and B) Examples of simultaneous multiplane imaging of a patient with mild residual PVR following TAVR. (C and D) Examples of simultaneous multiplane imaging of a patient with moderate residual PVR following TAVR; this patient had a post-dilation performed. The yellow dashed arrow in (A and C) shows the level of imaging in the orthogonal planes seen in (B and D) . PVR areas seen on a short-axis view in (B and D) were measured immediately after deployment, just below the THV stent.


From the long-axis two-dimensional TEE views of aortic valve, user-defined 3D TEE volumes of the aortic valve complex were acquired (single-beat acquisition). Commercially available software (QLAB MVQ version 8.1; Philips Medical Systems) was used to measure the aortic annulus using the following protocol ( Figure 3 , Video 1 ; available at www.onlinejase.com ):



  • 1.

    A short-axis view of the annulus is positioned in the “blue” plane using a double oblique technique.


  • 2.

    A frame in early systole is selected.


  • 3.

    The orthogonal long-axis green and red planes are then centered on the annulus in the blue (short-axis) plane.


  • 4.

    In the green and then the red planes, the blue plane is then positioned at the most caudal attachments of the aortic valve leaflets (hinge points).


  • 5.

    The orthogonal planes are repeatedly rotated in the blue plane, and the level of the annulus is confirmed in the green and red planes.


  • 6.

    Once the level and location of the annulus have been confirmed, the green and red planes are returned to position on the blue plane that will bisect the minimal diameter (MinDiam) and maximal diameter (MaxDiam), respectively.


  • 7.

    A pair of orthogonal points on the annulus is chosen. The MVQ package automatically approximates the annulus using these four points.


  • 8.

    Using an eight-slice protocol, the annular plane is confirmed for eight pairs of orthogonal segments in the green and red planes.


  • 9.

    The annular measurement must exclude ectopic calcifications, with the points measured outside the calcifications so as not to underestimate the annular area. In some cases, real-time imaging may help distinguish side-lobe and reverberation artifacts from real structures.




Figure 3


Multiplanar reconstruction with off-label use of vendor-specific analysis package for assessment of annular dimensions and area.


Using this protocol, the following annular measurements were made: MinDiam, MaxDiam, perimeter, and area. From these measurements, the following pre-TAVR annular calculations were made:




  • Mean annular diameter (MeanDiam) = (MinDiam + MaxDiam)/2



  • Average annular diameter derived from annular perimeter (AveAnnDiam P ) = perimeter/π



  • Average annular diameter derived from annular area (AveAnnDiam A ) = (area/π) 1/2 × 2



  • Eccentricity index = (MaxDiam − MinDiam)/MaxDiam



Undersizing the THV relative to the annular size has been identified in numerous studies to be the cause of PVR. Using the formula for cover index as 100 × [(THV diameter − TEE annular diameter)/THV diameter], cover indexes were calculated using the following TEE annular diameters 2DE SagDiam, 3DE MinDiam, 3DE MaxDiam, 3DE MeanDiam, 3DE AveAnnDiam P , and 3DE AveAnnDiam A .


Statistical Analysis


All analyses were performed using SPSS version 19.0 (IBM, Armonk, NY). P values < .05 were considered statistically significant, and all P values are two sided. Categorical variables were compared using χ 2 or Fisher’s exact tests and are expressed as percentages. Chi-square tests were used to calculate trend P values. Continuous variables were compared using unpaired Student’s t tests and are expressed as mean ± SD. For comparisons of the effect of postdilatation on PVR, patients were divided into two groups: those who received postdilatation and those who did not. Postdilatation was performed in cases in which PVR as determined by transesophageal echocardiography immediately after THV implantation was qualitatively more than mild, as determined by a single echocardiographer (R.T.H.). Postdilatation was performed with the same balloon used for THV implantation with the addition of up to 2 cm 3 of volume. Because a spontaneous regression rate has been noted in prior study from this lab, a PVR area cutoff of >29 mm 2 was chosen to indicate more than mild PVR (the mean + 1 SD for no postdilatation). Using this cutoff, receiver operating characteristic curves were generated for the cover indexes calculated using the annular dimensions listed above.


Intraclass correlation coefficients (ICCs) were used to assess interobserver (R.T.H. and O.K.) and intraobserver (R.T.H.) variability.




Results


Annular Dimensions


Table 1 lists the annular measurements obtained by two-dimensional and 3D echocardiography.



Table 1

Annular measurements and correlation of measurement means




























































Variable Measurement (mm) P vs 2DE annulus P vs MinDiam P vs MaxDiam P vs MeanDiam P vs AveAnnDiam P
2DE SagDiam 22.66 ± 1.89
3DE MinDiam 21.46 ± 2.01 <.0001
3DE MaxDiam 24.21 ± 2.66 <.0001 <.0001
3DE MeanDiam 22.83 ± 2.22 .360 <.0001 <.0001
3DE AveAnnDiam P 23.14 ± 2.31 .017 <.0001 <.0001 <.0001
3DE AveAnnDiam A 22.82 ± 2.25 .383 <.0001 <.0001 .782 <.0001