Physics of Ultrasound and Physics: Behind the Knobs



Physics of Ultrasound and Physics: Behind the Knobs


Peter Rozman1

Michael G. Licina2

James D. Thomas2

Robert M. Savage2


1OUTLINE AUTHOR

2ORIGINAL CHAPTER AUTHORS






I. INTRODUCTION



  • Echocardiography is the use of sound to produce an image of the heart and/or surrounding structures.


  • Sound is a mechanical vibration in a physical medium which stimulates hearing and travels in the form of a propagating wave that can be expressed graphically as a sine wave (Fig. 1-1).


  • Amplitude (A), which is measured in decibels (dB), is the maximal compression of particles above the baseline and equates to loudness in a sound wave.


  • Decibels are logarithmic units based on a ratio of the measured value (MV) to the reference value (RV), so dB = 20 log (MV/RV), which allows for compression of a large range of values into a small range.


  • Intensity (I) is the level of sound energy in an area of tissue and is proportional to the amplitude of the sound wave squared.


  • Wavelength (λ) is the distance between two adjacent areas of maximal compression, while frequency (f or Hz) is the number of wavelengths per unit time.


  • The propagation velocity (c) equals the wavelength times the frequency and is relatively constant in sound waves in the human heart (1.5 × 103 m/s).


  • Wavelength is important in image resolution, the ability to distinguish two points in space, because image resolution is no greater than 1 or 2 wavelengths and the depth of penetration of the ultrasound is directly proportional to it.


  • Acoustic impedance, the process of sound traveling through a medium, equals the density of the medium times the velocity of the sound and is responsible for reflection when sound beams travel between two tissues.


  • Reflection occurs when a sound wave reaches a boundary between surfaces with different acoustic impedances, and specular reflections occur on smooth surfaces.


  • Refraction is the change in direction of a sound wave as it travels between these surfaces and occurs when propagation speeds differ and the angle between the wave and the surface is oblique, causing imaging artifacts.


  • Scattering occurs when waves reflect off small, irregularly shaped objects.






FIGURE 1.1




  • Attenuation is the loss of the ultrasound wave and is directly related to the distance traveled.


  • Absorption is the conversion of ultrasound wave energy to another form of energy.


  • Sound can be classified as subsonic or infrasonic, audible sound, and ultrasound.


  • Ultrasound has a frequency greater than 20,000 Hz, can be directed in a beam, obeys wave properties, and is reflected by small objects.


II. GENERATION OF ULTRASOUND BEAM: PIEZOELECTRIC EFFECT



  • When a cut plate of quartz is subjected to mechanical stress, it will develop an electrical charge on its surface (Fig. 1-2), which is known as the pressure electric, or piezoelectric, effect (Fig. 1-3).


  • If a crystal is subjected to an alternating electric current, it will generate ultrasonic sound waves, forming the basis of ultrasonography.


  • The molecules in certain crystals are highly polarized, which explains their vibration when electrical charges are placed across the crystal (Fig. 1-3), just as the molecules will emit an electrical field when hit with ultrasound waves.1






FIGURE 1.2







FIGURE 1.3


III. IMAGING WITH ULTRASOUND



  • Due to the predictable time-distance relationship for ultrasound within the body (1,540 m/s), distances are easily determined given echo return times (Fig. 1-4).


  • The depth in cm is given by d = 77t, where t is the echo return time in ms (Fig. 1-4).2,3


  • The pulsed repetition frequency (PRF, in kHz) is approximated as 77/d (Fig. 1-4).


A. Wave interaction with tissues and organs



  • The amount of reflection that occurs at the tissue boundaries and interfaces is determined by the relative change in acoustic impedance.


  • Lateral dimensions greater than one wavelength act as specular reflectors, with an optimum angle occurring when the sound beam is perpendicular to the transducer.






FIGURE 1.4




  • Poor echo images occur due to dropout (the angle is greater than or less than 90 degrees) and scattering (lateral dimensions less than a wavelength).


  • Refraction allows enhanced images through acoustic lenses that focus the beam, but can also cause a “double image” artifact.


  • Conversion of ultrasound energy to heat, a form of absorption, results in attenuation, which has a direct relationship to frequency.


  • Due to the variances in returning signal strengths, a logarithmic compression, also called “time-gain compensation,” followed by differential amplification is necessary (Fig. 1-5).


IV. TRANSDUCERS



  • A transducer consists of the piezoelectric element, electrodes, case with insulation, and backing material.


  • A transducer is in receiver mode approximately 99% of the time,2,5,6 and can detect a wavelength signal that is less than 1% of the original.4,7


  • The electrodes stimulate and conduct electric current in and from the piezoelectric element.


  • The matching layer (faceplate), the layer between the esophagus and the piezoelectric element, causes the majority of the ultrasound to be transmitted instead of reflected, and possesses an acoustic lens to focus the beam.


  • The case and insulation protects from electric noise and prevents electric shock, while the backing material improves the picture quality by shortening the pulse duration and spatial pulse length.






FIGURE 1.5




  • Pulse wave Doppler uses a single piezoelectric crystal, while CW Doppler uses two.


V. WAVE FRONT CHARACTERISTICS



  • An unfocused beam travels in a column in its near field, or Fresnel zone, of which the length is directly proportional to the transducer’s diameter and inversely related to its wavelength (FN = D2/4λ) (Fig. 1-6).


  • In its far field, or Fraunholfer zone, the beam diverges in a manner that is directly proportional to the wavelength and inversely proportional to the transducer’s diameter (divergence angle = 1.22λ/D)4-6 (Fig. 1-6).


  • Focused zones can be manipulated by making the transducer lens concave or electronically focusing the beam, improving near field image resolution but worsening far field resolution (Fig. 1-6).


  • Axial (oriented along the length of the beam), lateral (side-to-side), or elevational (along beam thickness) resolutions exist, where axial is the most precise.


VI. REVIEW OF CONCEPTS



  • Wave energy exists as kinetic energy in the form of particle motion and as potential energy in the form of tissue compression and rarefaction.


  • Waves are characterized by the inverse relationship between wavelength and frequency.


  • A high frequency transducer will provide better resolution, while a low frequency transducer will provide better penetration.


  • Imaging ultrasound pulses must be extremely short, while Doppler pulses must be longer.






FIGURE 1.6



A. Resolution



  • Axial (or longitudinal, radial, range, depth) resolution distinguishes between two structures that are close to each other and front to back and depends on transducer frequency, bandwidth, and short pulse length.


  • Lateral resolution is the shortest distance that can exist between two structures while still producing two distinct echoes, equals beam diameter, and will vary with depth.


  • Temporal resolution is the ability to accurately locate moving structures at an instant in time and is affected by the number of pulses per scan line, the imaging depth, the sector size, and the line density.


VII. ULTRASOUND INSTRUMENTS AND IMAGING MODALITIES


A. A-, B-, or M-mode echocardiography



  • A- or B-mode echoes can determine distances between the transducer and the reflective surface, and the intensity of the reflected ultrasound.


  • A-mode echoes note intensity by the height of the oscilloscope’s electric signal, while B-mode echoes note intensity by varying brightness at the reflection point.


  • M-mode provides a time reference for B-mode, and a single crystal is used in both modes (Fig. 1-7).


  • PRF, or the frequency at which a transmission may be repeated, is typically up to 3,800 times/s for M-mode echoes.4,7,8






FIGURE 1.7



B. Two-dimensional echocardiography

May 26, 2016 | Posted by in CARDIOLOGY | Comments Off on Physics of Ultrasound and Physics: Behind the Knobs

Full access? Get Clinical Tree

Get Clinical Tree app for offline access