Magnetic Resonance Imaging and Computed Tomography




Echocardiography remains the most widely used diagnostic imaging modality in the assessment of patients with congenitally malformed hearts, owing to its portability and excellent temporal resolution, and the contrast provided by the interface between the blood and the tissues. 1 Ultrasound, however, does depend on acoustic windows, which are often limited in older patients or following surgery. Both computed tomography and magnetic resonance imaging allow imaging at all angles, and largely irrespective of surrounding structures or air. Consequently, and because they deliver multi-slice or true three-dimensional images, magnetic resonance ( Table 18D-1 ) and computed tomography help clarify the three-dimensional morphology of the cardiovascular system and its topographic relationships to the extracardiac structures, such as the trachea and bronchuses.



TABLE 18D-1

COMMON INDICATIONS AT A DEDICATED PAEDIATRIC SERVICE FOR CARDIAC MAGNETIC RESONANCE IMAGING

























Post-operative tetralogy of Fallot
Stenosis of the right and left pulmonary arteries
Native and repaired coarctation of the aorta
Aortic valvar insufficiency
Marfan syndrome
Partially anomalous pulmonary venous connection
Functionally univentricular circulations after the bidirectional cavopulmonary connection, or with Fontan-type anatomy
Arrhythmogenic right ventricular cardiomyopathy
Dilated and hypertrophic cardiomyopathy
Cardiac mass
Takayasu’s arteritis

Not given in order of frequency.



The history of magnetic resonance imaging is dotted with Nobel prizes awarded in physics, chemistry, and medicine. In 1946, Felix Bloch 2 and Edward Purcell 3 , in the same journal issue but independently of one another, described the behaviour of the nucleuses of certain elements of the periodic system when introduced into a magnetic field. Imaging using magnetic resonance, however, was not possible until Paul Lauterbur applied gradients spatially to encode the signal, producing the first magnetic resonance image of two tubes of water in 1971. This work was published in 1973, 4 and, in the following year, he published the first images of a living animal, a clam. 5 When Richard Ernst applied the Fourier transformation to the technique pioneered by Lauterbur, magnetic resonance imaging, as we know it today, was born. 6 The first magnetic resonance images of the heart date back to the early 1980s. 7 Major breakthroughs include electrocardiographic gating in 1984, 8 phase velocity mapping in the same year to measure the flow of blood, 9 and gadolinium-enhanced magnetic resonance angiography in 1988. 10


Since then, improvements of the hardware, particularly stronger and faster gradients, as well as higher resolution coils, paralleled by exponential increments in the capacity to process data, have laid the foundation for an array of sequences and applications. Between the early days of cardiovascular magnetic resonance imaging in the 1980s and the present, the technique has undergone an evolution from a non-invasive anatomical tool to a powerful window towards haemodynamics and myocardial mechanics.


The technological advancement in computed tomography is moving towards faster scanning, as well as a reduction in irradiation. Both improvements have helped to establish computed tomography in the work-up of pathology of the vessels and the airways in children. We can now obtain images during spontaneous breathing, and at an acceptable level of exposure to irradiation.


In this section of the chapter devoted to imaging, we offer an overview of the information now provided by magnetic resonance and computed tomography in children with congenital and acquired cardiac disease, discussing the limitations and risks of the techniques. We hope to guide cardiologists in using the appropriate cross sectional modality for imaging their patients, and in understanding the results.


MAGNETIC RESONANCE IMAGING


Physical Principle


Magnetic resonance imaging utilises the varying magnetic properties of hydrogen nucleuses, or protons, in different tissues in the body. 11,12 When excited by a radio-frequency pulse, the protons, like small stab magnets, briefly change orientation before returning to their initial state. During this relaxation, the protons give off energy. This process is termed resonance, or echoing. It can be measured, and then translated mathematically into an image of grey values. The way these echoes are created and manipulated determines the shading of the resulting image, a process termed weighting. Manipulation of the echo can be achieved by applying one or more additional radio-frequency pulses following the initial pulse that change the spin of the protons. This is called spin echo. Another way to influence the echo is by alternating the orientation of the magnetic field of the scanner during proton relaxation, resulting in a gradient echo sequence.


Magnetic resonance imaging operates between the poles of the duration of the scan, its temporal and spatial resolutions, and the intensity of the signal, or its ratio to the produced noise. As a general rule, improving one of these features typically worsens the other three. The settings used, therefore, have to be composed to maximise the benefit from a particular sequence in each individual patient. A comprehensive approach to the physics is beyond the scope of this chapter, and the interested reader is directed elsewhere. 11


Practical Considerations and Patient Safety


Cardiovascular magnetic resonance is a complex modality. Awareness of the underlying physical principles and computations is a prerequisite to its successful application to children, in whom a cookbook approach frequently fails. The field of congenital cardiac disease, on the other hand, bears its own intricacy that must be understood to best design the examination. Training in both aspects of paediatric cardiovascular magnetic resonance is essential to maximise the gain from this technique. An intimate cooperation between radiologists and cardiologists is advocated to master and help advance the fast evolving field.


It is our practice to sedate or anaesthetise most patients younger than 6 years of age. This, however, is only a rule of thumb, and varies according to the maturity of the patient, as well as to anaesthetic support, the monitoring available, and the presence of a physician. Careful preparation and guidance of the child prior to and during the scan have a beneficial effect on the quality of the images obtained. Prior to any study, absolute and relative contraindications must be addressed. 13,14 Pacemakers and other electronic implants, such as pumps for infusing drugs, are currently absolute contra-indications to magnetic resonance imaging. Controversy surrounds magnetic resonance imaging at 1.5 Tesla following implantation of metallic objects. 15 Patients with non-ferromagnetic foreign bodies can probably undergo testing immediately after implantation. For an implant or device that is weakly ferromagnetic, such as most stents, coils, and artificial valves, it is customary to wait until at least 6 weeks after implantation in order to prevent dislodging. 14 A strong magnetic field can induce an electrical current in temporary or permanent pacing wires, causing thermal injuries to surrounding structures. 16 If they are within or close to the field of view, all objects are desirably not only magnetic resonance imaging safe but also compatible, in order not to impair the information that can be obtained from the study ( Fig. 18D-1 ). 14 If possible, dental braces should be removed. The classification into safe and compatible is currently being replaced by the terminology safe and conditional, introduced by the American Society for Testing and Materials International. According to the new terminology, a patient with an implant labeled conditional may undergo magnetic resonance testing if certain requirements pertaining to the scanner and sequences used are met. Both classifi cations currently co-exist and can be found on the device packaging. Gadolinium-based contrast mediums are safe, with a risk of anaphylactic reactions in the range of 0.001% to 0.01%. 14 Patients with severe renal insufficiency, expressed by a glomerular filtration rate of less than 30 mL/min/1.73 m 2 , face a risk of developing nephrogenic systemic fibrosis after exposure to gadolinium, and should not receive contrast. 17 Nephrogenic systemic fibrosis was initially observed in, and thought solely to affect, the skin, but internal organs such as liver, lungs, and heart may be involved. An extensive review of the safety aspects of magnetic resonance can be found elsewhere. 14 With the exception of real-time imaging and contrast-enhanced angiography, most sequences require electrocardiographic gating to compensate for cardiac motion. Respiratory motion is countered either by breath-holding, tracking the motion of the diaphragm, or acquiring multiple sets of data to average out the effects of the inconsistent position of the thorax.




Figure 18D-1


Artifact from a vascular stent in the arterial duct following a hybrid procedure consisting of bilateral banding of the pulmonary arteries and placement of a ductal stent for hypoplastic left heart structures. DAO, descending aorta; LV, left ventricle; PT, pulmonary trunk; RPA, right pulmonary artery; TA, transverse arch.


Techniques for Imaging


Each examination begins with a series of static scout images, or localizers, of the thorax and adjacent body parts in three orthogonal body planes ( Fig. 18D-2 ). With most recent scanners, the scout images with an acceptable spatial resolution can be obtained within 1 or 2 minutes, making most anatomical information readily available. All subsequent sequences for detailed examination are planned using this scout, as well as any subsequently obtained images, as a reference. As a principle of prescription, any imaging plane is defined unequivocally either by three points, so-called three-point planning, or by how it dissects two separate images previously obtained, the so-called double-oblique technique.




Figure 18D-2


Bright blood view of the left ventricular outflow tract in a patient with combined aortic valvar disease. The aortic valve is doming and stenotic, with post-stenotic dilation of the proximal ascending aorta. The turbulent jet causes a characteristic dephasing artifact. AO, aorta; LV, left ventricle.


Cine Imaging


These moving images of the beating heart are the workhorse of current magnetic resonance imaging in patients with congenitally malformed hearts ( Figs. 18D-3 and 18D-4 ; see also Fig. 18D-2 ). Blood is signal-intense, and contrasts well with the grey myocardium. Turbulent flow causes loss of signal from dephasing, and can be identified as dark streaks within the bright blood pool (see Fig. 18D-2 ). As an important difference to echocardiography, these moving images, with few exceptions, are not acquired in real time, but are assembled over many cardiac cycles, so that a temporal correlation of cardiac events with the coinciding beat-to-beat electrocardiogram is not possible. A representative work flow for cine imaging is shown in Figure 18D-3 . The strategies to achieve the basic views vary between users. A stack of 10 to 12 short-axis cine images (see Figs. 18D-3 and 18D-4 ) are used to calculate ventricular volumes and myocardial masses (see Fig. 18D-4 B). As opposed to echocardiography, this method does not rely on geometric assumptions, and is widely accepted as the gold standard tool for ventricular volumetry.




Figure 18D-3


Work flow for anatomical and cine imaging. Each scan plane is prescribed starting from two differently angled reference images, using the double oblique technique, for the four-chamber view in this figure. Following coronal ( A ), axial ( B ), and sagittal ( C ) localizer images, cine imaging begins with a vertical long-axis or pseudo-two-chamber view ( D ). From here, a set of short-axis imaging planes ( E ) is prescribed parallel to the atrioventricular groove, using the previously acquired vertical long-axis view and the axial localizer as reference images. A four-chamber view ( F ) cuts through the mitral and tricuspid valves and through the ventricular apex. If needed, a three-chamber view ( H ) of the left atrium (LA), left ventricle (LV), and proximal ascending aorta can be obtained by placing the imaging plane through the aortic and mitral valves (AV, MV) in the basal short-axis plane ( G ). Imaging perpendicular to the three-chamber view opens up the left ventricular outflow tract ( I ). AO, aorta; PT, pulmonary trunk; RA, right atrium; RV, right ventricle; TV, tricuspid valve.



Figure 18D-4


Stack of short-axis images, parallel to the atrioventricular groove. Contours ( A ) are drawn around the left ventricular endocardium (red), epicardium (green), and right ventricular endocardium (yellow). This is repeated at end-diastole and end-systole in each slice of the short-axis stack. Using the contours in A , a three-dimensional model ( B ) is reconstructed. The red contours and green mesh indicate the endocardium and epicardium, respectively, of the left ventricle. The right ventricular endocardium is in yellow. The computer calculates the volumes, ejection fractions, and cardiac output. The mass is typically given only for the left ventricle, as the right ventricular trabeculations preclude accurate measurements.


A modification of cine imaging is myocardial tagging ( Fig. 18D-5 ): Here, selective saturation pre-pulses spoil all spins within multiple planes perpendicular to the plane of imaging, thus superimposing a grid of black lines across the field of view. The black lines persist through systole, and are deformed by myocardial contraction. From the degree of deformation, it is possible to calculate radial and circumferential strain, as well as ventricular twisting. 18–21 The assessment of cardiac function is fuelled by the expectation of better understanding of myocardial mechanics, and by the hope of detecting and treating myocardial disease in its early stages. The sensitivity of analysis of regional wall motion can be increased by protocols inducing stress, using dobutamine or magnetic resonance compatible ergometers.




Figure 18D-5


This 17-year-old boy has myocardial infarction after Kawasaki disease as a toddler. Myocardial tagging is shown in a midventricular short-axis slice at the beginning ( A ) and end ( B ) of systole. The tagging lines change from straight to inward convex, following the myocardial motion during contraction ( black arrowheads ). The inferolateral wall of the left ventricle (LV) is akinetic ( white arrowheads ), indicated by the grid lines remaining straight. The black lines are strongest at the beginning of systole and fade during the cardiac cycle. The akinetic wall segment of the left ventricle corresponds to a zone of hyperenhancement late after gadolinium, indicating myocardial infarction ( C ; arrowheads ). RV, right ventricle.


Measurements of Flow


Phase contrast imaging is another key technique in the assessment of congenitally malformed hearts by magnetic resonance. It is used to quantify the velocity and volumes of flow ( Fig. 18D-6 ). 22–24 The ability to assess the flow in all major vessels in any orientation makes this technique ideal for the assessment of congenitally malformed hearts. Data is acquired during spontaneous breathing or with the patient ventilated, as breath-holding manoeuvres can affect the flow of blood. The vessel is imaged perpendicular to its long axis, so-called through-plane imaging. In Figure 18D-7 ,we show the prescription of the planes to image the great vessels as well as the atrioventricular valves, using the double oblique technique. Phase contrast sequences produce a cine modulus image (see Fig. 18D-6 A) of the anatomy, as well as a velocity-encoded image which contains the information pertaining to flow (see Fig. 18D-6 B). Phase contrast imaging is used to quantify valvar regurgitation. Flow can accurately be measured in the right and left pulmonary arteries, as well as in systemic vessels. By subtracting pulmonary arterial influx from venous efflux, the amount of systemic-to-pulmonary arterial collateral flow can be calculated in patients with a functionally univentricular circulation, or in those with obstructed right ventricular outflow tracts. 25–27 Flow across a patent arterial duct can be directly measured. Intracardiac shunts can be calculated by comparing flows in the ascending aorta and pulmonary trunk. 28,29 Unlike ultrasound Doppler, phase contrast imaging allows analysis of the pattern of flow in all vessels shown in the field of view, helping clarify timing of flows relative to one another. Certain conditions exhibit characteristic patterns: for example, the configuration of the velocity of flow in patients with pulmonary hypertension ( Fig. 18D-8 ) is characterised by an early systolic peak with decreased maximum velocity; one or more secondary peaks; early cessation of systolic forward flow, with a nadir arriving before that of the ascending aortic flow; and undulation of velocities during diastole. 30 Flow trajectories and turbulence can be assessed by imaging along the vessel or jet of flow, known as in-plane imaging, and, experimentally, it is possible to produce four-dimensional phase contrast imaging ( Fig. 18D-9 ). A number of conditions compound the accuracy of phase contrast magnetic resonance, including insufficient temporal resolution, a too high or too low limit of encoding, as well as non-laminar target flow. 23 Large translational movements of the great arteries through the fixed imaging plane, as well as turbulence and swirling in frequently dilated arterial roots hamper the quantification of regurgitation across arterial valves. 31,32




Figure 18D-6


Magnitude ( A ) and velocity ( B ) images of a phase contrast study of the ascending aorta in a patient with aortic valvar regurgitation. The resulting curves of flow volume over time ( C ) show flow reversal in the ascending (area shaded in red) and descending (area shaded in yellow) aorta. Flow in the superior caval vein is shown in blue. In patients without extracardiac shunts, the sum of flow volumes in the superior caval vein (SCV) and in the descending aorta (DAO) are expected to equal that in the ascending aorta (AAO) with minor discrepancies for measurement inaccuracy and blood flow into the bronchial and vertebral arteries. PT, pulmonary trunk.



Figure 18D-7


Work flow for phase contrast imaging. The prescribed imaging planes are perpendicular to the three-dimensional course of the vessel as long as they transect the vessels on two separate images. Using the localizer images in three orthogonal planes, the cuts for the ascending aorta ( A ) at the level of the right pulmonary artery, the inferior caval vein ( B ) at the diaphragm, and the pulmonary trunk ( C ) are designed. The plane for the ascending aorta typically transects the descending aorta and superior caval vein in a near-perpendicular fashion. The plane for the descending aorta at the level of the diaphragm includes the inferior caval vein between the liver and the atrium. The right ( D ) and left ( E ) pulmonary arteries are prescribed off the ascending aorta phase contrast magnitude image and the coronal and sagittal localizers, respectively. Inflow through the atrioventricular valves ( F ) is imaged in a plane perpendicular to the atrioventricular groove in a vertical long-axis view at end-systole and an axial localizer image.



Figure 18D-8


Time-velocity flow profile in the pulmonary trunk in a patient with primary pulmonary hypertension. See text for details.



Figure 18D-9


Flow streamlines in a normal thoracic aorta, reconstructed from four-dimensional, three-directional phase contrast velocity mapping.

(Courtesy of A. Frydrychowicz, Freiburg, Germany.)


Contrast-enhanced Magnetic Resonance Angiography


Gadolinium-based contrast agents can be used to increase the signal from the blood and reduce the time required to acquire images, thus allowing three-dimensional angiography with a submillimeter spatial resolution ( Figs. 18D-10 to 18D-12 ; see also Fig. 18D-9 ). When repeated in short intervals, four-dimensional angiography can be performed, with time added as the fourth dimension to the three-dimensional anatomy. The practical temporal resolution of four-dimensional or time-resolved magnetic resonance angiography for the pulmonary vasculature is between 0.5 and 1.0 seconds, depending on the imaging volume and the spatial resolution. The interpretation of an angiogram should always begin with a careful review of the source data. Image processing can highlight the information contained in the images, clarify topographic relationships, and facilitate demonstration of the anatomy to cardiologists and surgeons. A three-dimensional data set can be displayed either as a projection or as a volume rendered image ( Fig. 18D-13 ; see also Figs. 18D-10 through 18D-12 ). Maximum intensity projections (see Figs. 18D-12A and 18D-13A ), the most commonly used type of projection, resemble static images from fluoroscopic angiography. To create these images, the computer retains the brightest voxels along a virtual beam of projection. The orientation of this projection, as well as the thickness of the projected slice, is freely adjustable during post-processing. Computationally intensive volume rendered images (see Figs. 18D-10, 18D-11, 18D-12B, and 18D-13B ) lend a plastic appearance to the anatomy, optionally including colour and a virtual source of light.




Figure 18D-10


Contrast-enhanced magnetic resonance angiography in a patient with severe native coarctation ( asterisk ) of the aorta (AO) and massive collateralisation via the internal mammary, intercostal, and vertebral arteries. IMA, internal mammary artery.



Figure 18D-11


Discordant ventriculo-arterial connections in a patient with transposition after an atrial switch procedure. The trabeculated right ventricle (RV) connects to the aorta (AO; shown in red), and the smooth-walled left ventricle (LV) gives rise to the pulmonary trunk (PT; in blue). The complex venoatrial anatomy including the baffl e connecting the superior caval vein (SCV) to the mitral valve, and separating the systemic venous return from the morphological right atrium ( asterisk ) is clearly shown.

(Courtesy of S. Sarikouch, Hannover, Germany.)



Figure 18D-12


Maximum intensity projection ( A ) and volume-rendered ( B ) images of an isolated stenosis ( asterisk ) at the origin of the left pulmonary artery (LPA). PT, pulmonary trunk; RA, right atrium; RPA, right pulmonary artery; RV, right ventricle.

Apr 6, 2019 | Posted by in CARDIOLOGY | Comments Off on Magnetic Resonance Imaging and Computed Tomography

Full access? Get Clinical Tree

Get Clinical Tree app for offline access