Immediate Post-Stent Evaluation with Optical Coherence Tomography



Fig. 16.1
Optical coherence tomographic images of in-stent restenosis 1 year after stent implantation . On left panel, the percentage of area stenosis was 60.1%, showing in-stent restenosis. The cross-sectional area of stent and neointima was, respectively, 3.1 mm2 and 0.4 mm2, suggesting that stent underexpansion was a causative mechanism for in-stent restenosis. However, on right panel, neointimal hyperplasia was a restenotic mechanism. The cross-sectional area of stent and neointima was 7.2 mm2 and 6.3 mm2, respectively





16.2 Stent Apposition


The position of individual struts from the vessel wall can be assessed qualitatively (apposed or malapposed) or quantitatively (distance or area) with OCT. Immediately after stent implantation, individual stent struts are in touch with the vessel wall (apposed) or not (malapposed). Apposed struts are further classified into embedded or protruding. Struts are defined as embedded if more than half of the strut thickness is below the level of luminal surface [6] and as protruding if the adluminal strut surface is just above the vessel wall [6]. Considering the thickness of stent strut or abluminal polymer, struts that are detached from the vessel wall are defined as malapposed [7]. Accordingly, the distances for defining malapposed struts are different among types of drug-eluting stents (Table 16.1). Representative OCT images of apposed and malapposed struts are shown in Fig. 16.2.


Table 16.1
Definitions of malapposed strut according to types of drug-eluting stents [8, 9]








































Trade name

Company

Distance between vessel wall and strut

Cypher

Cordis

≥160 μm

Taxus

Boston scientific

≥130 μm

Endeavor resolute

Medtronic

≥110 μm

Resolute integrity

Medtronic

≥110 μm

Xience

Abbott vascular

≥100 μm

Nobori

Terumo

≥130 μm

Biomatrix

Biosensors

≥130 μm


A430616_1_En_16_Fig2_HTML.gif


Fig. 16.2
Representative optical coherence tomographic images of apposed (left panel) and malapposed struts (right panel) immediately after implantation. Arrows indicate embedded struts and arrowheads protruding struts. All struts were malapposed on top right panel, whereas partial struts (asterisks) were malapposed on bottom right panel

The malapposed struts are commonly observed immediately after stent implantation. The incidence of acute stent malapposition was approximately 40–60% (Table 16.2), and about 70% of acute stent malapposition spontaneously disappeared at follow-up of 1 year [8, 10]. Determinants of the spontaneous resolution were malapposed distance or area, indicating that tiny malapposition can be resolved at follow-up [8, 10, 11]. The coverage of malapposed struts is delayed compared with apposed struts [12]. Recent two registries investigating mechanisms of stent thrombosis showed that stent malapposition was the most frequent finding (about one third of all cases) that caused stent thrombosis [13, 14]. In these studies, a significant or observed maximal distance of malapposition was >200–300 μm (Fig. 16.3) [13, 14]. Accordingly, it may be important to interpret malapposition as a quantitative, rather than binary phenomenon (present or absent), and to define the threshold of malapposition detachment that may benefit from optimization during stent implantation [15]. However, simple presence of acute stent malapposition >200 μm itself was not associated with worse outcomes [4], and results from prospective trials have not been addressed yet. On the contrary to malapposed struts, embedded struts at post-intervention were highly covered at follow-up of 6 months (median percentage of uncovered struts, 0% in embedded and 26.8% in malapposed, p < 0.001) [16]. Using contour plot analysis [17], the associations between strut apposition immediately after intervention and strut coverage at follow-up can be identified at a glance. On the contour plot, x-axis represented circumferential arc length of individual stent strut, and y-axis represented stent length. In (x, y) format, the locations of stent struts were delineated by their pixel coordinates [17]. In addition, individual stent struts could be marked regarding the status of apposition or coverage. Figure 16.4 is an example of serial contour plot analysis at post-intervention and follow-up.


Table 16.2
Frequency of acute malapposition , detected by optical coherence tomography


































Authors

No. of study population

Stent type

Frequency of acute malapposition

Kawamori et al. [10]

40

DES

65%

Im et al. [8]

356

DES

62%

Soeda et al. [5]

1001

BMS and DES

39%

Prati et al. [4]

1002

BMS, DES, and BVS

49%


BMS bare metal stent; BVS bioresorbable vascular scaffold; DES drug-eluting stent


A430616_1_En_16_Fig3_HTML.jpg


Fig. 16.3
A case of very late stent thrombosis related to malapposed struts. A 69-year-old man visited the emergency department for severe angina. He underwent biolimus-eluting stent (3.5 mm × 18 mm) implantation 24 months ago. Electrocardiogram showed ST-segment elevation on inferior leads, suggesting acute myocardial infarction. Emergent angiography showed intraluminal haziness (arrow) within stent of right coronary artery. After thrombosuction, optical coherence tomography revealed that irregular-shaped thrombi (arrowheads) were attached to malapposed struts. The maximal distance between malapposed struts and vessel wall was 560 μm

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Jan 19, 2018 | Posted by in CARDIOLOGY | Comments Off on Immediate Post-Stent Evaluation with Optical Coherence Tomography

Full access? Get Clinical Tree

Get Clinical Tree app for offline access