Electrocardiographic correlates of mechanical dyssynchrony in recipients of cardiac resynchronization therapy devices




Summary


Background


The relationship between electrical and mechanical indices of cardiac dyssynchronization in systolic heart failure (HF) remains poorly understood.


Objectives


We examined retrospectively this relationship by using the daily practice tools in cardiology in recipients of cardiac resynchronization therapy (CRT) systems.


Methods


We studied 119 consecutive patients in sinus rhythm and QRS ≥ 120 ms (mean: 160 ± 17 ms) undergoing CRT device implantation. P wave duration, PR, e PR (end of P wave to QRS onset), QT, RR–QT, JT and QRS axis and morphology were putative predictors of atrioventricular (diastolic filling time [DFT]/RR), interventricular mechanical dyssynchrony (IVMD) and left intraventricular mechanical dyssynchrony (left ventricular pre-ejection interval [PEI] and other measures) assessed by transthoracic echocardiography (TTE). Correlations between TTE and electrocardiographic measurements were examined by linear regression.


Results


Statistically significant but relatively weak correlations were found between heart rate ( r = −0.5), JT ( r = 0.3), QT ( r = 0.3), RR–QT intervals ( r = 0.5) and DFT/RR, though not with PR and QRS intervals. Weak correlations were found between: (a) QRS ( r = 0.3) and QT interval ( r = 0.3) and (b) IVMD > 40 ms; and between (a) e PR ( r = −0.2), QRS ( r = 0.4), QT interval ( r = 0.3) and (b) LVPEI, though not with other indices of intraventricular dyssynchrony.


Conclusions


The correlations between electrical and the evaluated mechanical indices of cardiac dyssynchrony were generally weak in heart failure candidates for CRT. These data may help to explain the discordance between electrocardiographic and echocardiographic criteria of ventricular dyssynchrony in predicting the effect of CRT.


Résumé


Introduction


Les corrélations électromécaniques sont peu connues chez les patients présentant une insuffisance cardiaque avec dysfonction ventriculaire gauche. L’objectif de cette étude est d’essayer de mieux comprendre les relations entre l’activation électrique et l’asynchronisme mécanique dans cette population.


Patients et méthodes


Cent dix-neuf patients insuffisants cardiaques ayant une indication classique de resynchronisation ont été inclus dans cette étude rétrospective. Les asynchronismes atrioventriculaire (DFT/RR), interventriculaire (IVMD) et intraventriculaire (délai préejectionnel VG [LVPEI] et d’autres mesures) ont été évalués en échographie transthoracique. La fréquence cardiaque, la durée de l’onde p, les intervalles PR, P′R (entre la fin de l’onde p et le début du QRS), RR–QT, JT, QT, QRS, l’axe et la morphologie des QRS ont été définis comme des critères prédictifs possibles de l’asynchronisme mécanique. Les corrélations entre les paramètres échographiques et les mesures électriques ont été analysées sous forme de régressions linéaires.


Résultats


On observe une corrélation significative entre la fréquence cardiaque ( r = 0,50), le JT ( r = 0,40), le QT ( r = 0,30), l’intervalle RR–QT ( r = 0,0) et le ratio DFT/RR ; cette relation n’est pas observée pour les intervalles PR et QRS. Une corrélation significative mais faible est observée entre les intervalles (a) QRS ( r = 0,24) et QT ( r = 0,24) et (b) IVMD > 40 ms, et entre les intervalles (a) ePR ( r = 0,24), QRS ( r = 0,30), QT ( r = 0,24) et (b) LVPEI. On ne retrouve pas de corrélations significatives avec les autres paramètres d’asynchronisme intraventriculaire gauche.


Conclusion


Les corrélations électromécaniques sont globalement faibles dans cette population. Ces observations peuvent nous amener à nous poser, d’une part, la question de la validité des critères échographiques utilisés actuellement pour caractériser l’asynchronisme mécanique et, d’autre part, peuvent laisser penser que l’effet bénéfique de la resynchronisation est multifactoriel et ne résulte pas seulement de la correction des anomalies mécaniques.


Background


Cardiac resynchronization is an important means of managing heart failure for patients presenting with a wide QRS complex and a left ventricular ejection fraction (LVEF) < 35%, who remain in New York Heart Association functional classes II–IV despite an optimal pharmaceutical regimen. Cardiac resynchronization therapy (CRT) alleviates symptoms and lowers major heart failure morbidity, all-cause mortality and the risk of sudden death . Electrical dyssynchrony on surface electrocardiogram (ECG), manifest in the QRS morphology (left bundle branch block [LBBB] pattern) and duration, is a strong predictor of clinical outcome after CRT . Current guidelines recommend basing patient selection on electrical dyssynchrony criteria .


In the past 10 years, several echocardiographic indices of mechanical dyssynchrony have been proposed to prospectively identify responders to therapy. Despite the promising results of observational studies from single centres, most echocardiographic measurements made in large multicentre non-randomized or randomized trials, including analyses by core laboratories, have failed to predict the effect of CRT. In the recent EchoCRT study, the therapy failed to reduce the rates of death from any cause and first hospitalization for management of heart failure in patients presenting with a QRS ≤ 130 ms but echocardiographic signs of left ventricular (LV) dyssynchrony . This discordance between electrical and mechanical dyssynchrony in patients with heart failure remains unexplained, although the attempts made thus far to find electromechanical correlations have been suboptimal.


The aim of the present study was to revisit and try to better understand the relationship between electrical activation and mechanical dyssynchrony in the left heart of heart failure patients who are candidates for CRT, by using the daily practice tools in clinical cardiology (i.e. 12-lead surface ECG and Doppler echocardiography). CRT response was not considered in this study.




Methods


Consecutive patients scheduled to undergo implantation of CRT systems at the Rennes University Medical Centre between March 2009 and March 2012 were retrospectively included in this study. The inclusion criteria were: New York Heart Association functional classes II–IV despite optimal medical therapy; LVEF ≤ 35%; stable sinus rhythm; QRS duration ≥ 120 ms on 12-lead ECG; and no previous pacemaker or cardioverter defibrillator implantation. The heart disease was considered ischaemic if > 50% stenosis was observed in at least one major epicardial coronary artery or if the patient had a history of myocardial infarction or coronary revascularization.


All patients granted their informed consent to participate in the study, which was reviewed and approved by our institutional ethics review committee.


Electrocardiography


Before CRT implantation, standard 12-lead ECGs were recorded at 25 mm/s paper speed and calibrated at 1.0 mV/cm before recording of the echocardiogram. The method used for ECG analysis has been reported . Heart rate, P wave duration, PR interval, e PR interval (end of P wave to onset of QRS), QRS duration, QT interval, JT interval (end of QRS to onset of T wave) and RR cycle minus QT interval (as a measure of electrical diastole) were measured, and the QRS morphology and axis were analysed. The frontal plane QRS axis was considered normal when between −30° and +90°, left deviated when beyond −30°, and right deviated when beyond +90°. LBBB was defined as a QRS duration ≥ 120 ms, with a broad R wave in leads I, aV L , V 5 and V 6 and an R peak time > 60 ms in leads V 5 and V 6 , according to the practice guidelines issued by major professional societies . Other intraventricular conduction disturbances were classified as right bundle branch block or non-specific intraventricular conduction delays. The intra- and interobserver reproducibility of the measurements were ascertained by comparing the analysis of 20 randomly selected ECGs by two experts unaware of each other’s interpretation.


Transthoracic echocardiography


All patients underwent resting two-dimensional Doppler and speckle-tracking transthoracic echocardiography before CRT device implantation, using Vivid 7 or Vivid E9 ultrasound instrumentation (General Electric Medical Systems, Horten, Norway), according to a standardized protocol for image acquisition. LV end-systolic and end-diastolic diameters were measured in the parasternal long-axis view with M-mode; LV volumes indexed to body surface area and LVEF were measured in apical four- and two-chamber views using the biplane Simpson’s method .


The septal and lateral mitral annular peak systolic velocities were measured from the apical four-chamber view using tissue Doppler imaging. LV strain was analysed by speckle-tracking echocardiography using the four-chamber and mid-LV short-axis views. Images were acquired at end-expiration and analysed off line, using the dedicated automated imaging function of the EchoPAC BT12 ® software package (GE Healthcare, Chalfont St Giles, UK). We used the echocardiographic indices of mechanical dyssynchrony previously published by Gorcsan et al. , and measured according to the recommendations of the American Society of Echocardiography and the Heart Rhythm Society.


Definitions


The diastolic filling time (DFT)/RR interval ratio was used to characterize atrioventricular (AV) dyssynchrony in the left heart. AV dyssynchrony was defined as DFT/RR < 40% .


Interventricular mechanical dyssynchrony (IVMD), calculated as the time difference between right ventricular (RV) and LV ejection at the onset of pulsed Doppler flow velocities in the LV and RV outflow tracts, respectively, was used to characterize interventricular dyssynchrony. Interventricular dyssynchrony was defined as IVMD > 40 ms .


LV pre-ejection interval (LVPEI), the delay between the onset of QRS and the beginning of LV ejection flow by Doppler imaging, was used to characterize left intraventricular dyssynchrony. Intraventricular dyssynchrony was defined as LVPEI > 140 ms .


Intraventricular longitudinal dyssynchrony was defined as the maximum delay between opposing septum-to-posterior wall, in colour-coded tissue Doppler imaging in the apical long-axis view, or by the maximum delay between opposing septum-to-posterior wall in speckle-tracking longitudinal strain imaging, in the apical long-axis view .


Intraventricular radial dyssynchrony was defined as the delay between opposing anteroseptal-to-posterior wall in the mid-LV short-axis view in speckle-tracking radial strain imaging .


The intrinsic intra- and interobserver reproducibilities of intraventricular dyssynchrony measured by longitudinal strain were ascertained from corresponding repeated measurements, using intraclass correlations. The intra- and interobserver reproducibilities were evaluated by a second measurement of 20 randomly selected transthoracic echocardiograms. The intra- and interobserver reproducibilities of other echocardiography measurements are already known .


Statistical analysis


A descriptive analysis of pertinent patient characteristics is expressed as means ± standard deviations or counts and percentages, unless specified otherwise. Correlations between indices of mechanical dyssynchrony and electrocardiographic measurements used Spearman’s correlation coefficients (and 95% confidence intervals [CIs] through Fisher’s Z transformation) and linear regression estimates. ‘Good’ correlation was considered when the r coefficient was ≥ 0.50. Categorical variables were compared between groups using Fisher’s exact test, while continuous variables were compared using the non-parametric Kruskall–Wallis test. A multivariable regression analysis included covariates emerging at a P ≤ 0.05 statistical level in the univariate analysis; we estimated a squared semi-partial coefficient, which represents the proportion of variance in y that is explained by x 1 only. Functional forms of continuous covariates were assessed graphically and statistically, using non-parametric regression and the PROC GAM smoothing technique (SAS Institute, Cary, NC, USA). A two-sided P value < 0.05 was considered statistically significant. The data were analysed with the SAS ® software package, version 9.3 (SAS Institute, Cary, NC, USA).




Results


The study sample consisted of 119 patients; their baseline demographic, clinical, electrocardiographic and echocardiographic characteristics are presented in Table 1 . The disease aetiology was ischaemic in one-third of patients. Over 90% of patients were treated with a beta-blocker and an angiotensin-converting enzyme inhibitor or angiotensin II receptor blocker at the highest tolerated doses.



Table 1

Baseline characteristics of 119 study participants.










































































































Characteristic
Age (years) 64 ± 10
Men 81 (68)
NYHA functional class 2.8 ± 0.4
Ischaemic cardiomyopathy 39 (32)
Distance covered in 6-minute walk (m) ( n = 62) 400 ± 96
N-terminal B-type natriuretic peptide (pg/mL) ( n = 89) 2154 ± 2300 [107–13,000]
Drug regimen
Beta-blocker ( n = 107) 103 (94.5)
ACE inhibitor or ARB ( n = 107) 101 (93)
Electrocardiogram
Heart rate (beats per minute) 66 ± 12
P wave duration (ms) 111 ± 28
PR interval (ms) 202 ± 4
e PR interval (ms) 90 ± 50
QRS duration (ms) 160 ± 17 [130–200]
QT interval (ms) 450 ± 37
JT interval (ms) 289 ± 36
RR–QT interval (ms) 470 ± 188
LBBB 101 (85)
QRS axis within normal range 61 (51)
Transthoracic echocardiogram
LVEF (%) 26.6 ± 0.6
LV end-diastolic diameter, mm 67.8 ± 8.2
DFT/RR 0.44 ± 0.11
DFT/RR < 0.4 41 (35)
Interventricular mechanical delay (ms) 42 ± 24
Interventricular mechanical delay > 40 ms 54 (45)
LVPEI (ms) 137 ± 35
LVPEI > 140 ms 58 (47)
Septal-lateral delay by Doppler tissue imaging (ms) 97 ± 85
Septal-lateral delay by two-dimensional strain (ms) 240 ± 129
Anteroseptal-posterior delay by two-dimensional strain (ms) 158 ± 131
Overlap 15 (13)
Septal flash 40 (34)

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Jul 11, 2017 | Posted by in CARDIOLOGY | Comments Off on Electrocardiographic correlates of mechanical dyssynchrony in recipients of cardiac resynchronization therapy devices

Full access? Get Clinical Tree

Get Clinical Tree app for offline access