and Alwyn Scott2
(1)
School of Computer Science, University of Manchester, Manchester, UK
(2)
Cardiology High Dependency Unit, Papworth Hospital NHS Foundation Trust, Cambridge, UK
Keywords
WaveformsIntervalsSegmentsArtefactCalibrationRateRhythmBackground
Interpreting the ECG is a task that requires an underpinning knowledge of the way the ECG is organised, what is being displayed and what the normal ranges and values of the various waveforms, intervals and segments should be. In addition to this initial knowledge base practitioners should spend as much time as they can looking at real ECGs in context, and preferably discussing these findings with more experienced colleagues. Like learning to play the piano or developing foreign language proficiency it requires many hours of practice built on top of the basic knowledge gained from books, lectures and other sources.
There are many ways to interpret an ECG. The authors recommend using a systematic approach that includes the following aspects:
Basic quality control checks
Rhythm determination
Rate calculation
Examination of the morphology and duration of the various waves, intervals and segments
Determining of the QRS axis
Scanning for any additional features or abnormalities
We begin by looking at how the ECG is organised on the paper and in leads and move on to look at the normal values for the various components of the ECG waveform, including different methods of calculating the rate and determining the rhythm.
ECG Paper
12-lead ECGs are usually displayed on special gridded paper (Fig. 2.1). The 12 leads I, II, III, aVR, aVL, aVF and V1–V6 are displayed on the paper under their respective headings. The limb leads are found on the left hand side and the chest leads V/C 1–6 on the right. The leads are separated by lead divider markers which resemble an elongated punctuation colon (:). The gridded area of the paper is split into larger boxes with smaller boxes inside them. Each large box measures 5 mm2 (containing 25 smaller 1 mm2 boxes). Time is measured along the x-axis (horizontal) in seconds. Each large box represents 0.2 s of time, with each smaller box measuring 0.04 s. Each lead represents around 3 s of time. Most 12-lead ECGs also include a rhythm strip, shown below the other leads. This strip is one of the existing leads displayed above, shown for around 12 s, usually lead II or V1. This allows interpreters to look for patterns and features that might not otherwise be visible in shorter time periods. Some ECGs have more than one rhythm strip included. The y-axis (vertical) represents the amplitude of the ECG waveforms, measured in millivolts. One large square represents 0.5 mV, with a small square measuring 0.1 mV.
Fig. 2.1
ECG paper details
Basic Checks
There are several details that should be checked prior to analysing the waveforms and formulating an interpretation. These checks include:
Ensuring the ECG is free from artifact and recorded at sufficient quality to enable a subsequent interpretation.
Checking R wave progression in the chest leads and the deflection of lead aVR
Checking the calibration markers/calibration signal boxes to ensure the ECG is recorded using the standard settings.
Artefact
Is any artificial disturbance that negatively impacts on the quality of the ECG. There are many forms of interference that can affect the quality. There are however several types of commonly encountered interference that can be easily recognised and prevented or reduced in most cases. Table 2.1 shows the most common forms of interference, including possible causes and solutions.
Table 2.1
Common forms of artifact with possible causes and solutions
Muscle/Somatic tremor | Baseline wander | 60-cycle/AC mains interference |
---|---|---|
Cause: Usually caused by patients moving during a recording. | Cause: May be caused by movement of the ECG cables, perspiration or respiratory swing seen in some diseases, such as COPD. | Cause: Improper grounding of electrical equipment causing interference or a fractured wire within an ECG cable. |
Possible solution(s): | Possible solution(s): | Possible solution(s): |
Ask the patient to remain still during recording. If caused by shivering, ensure patient warm before attempting recording. | Ensure cables are not dangling over the edge of the bed or allowed to move during recording. | Remove the equipment/electrical device causing the interference. In the case of medical equipment turn off any non-essential devices that may be causing the interference. |
In the case of pathological tremor such as Parkinson’s disease, patients can be asked to place their arms under their legs. Modified lead positions can also be used to reduce interference (i.e. shoulders and upper legs). | Cleanse the skin with alcohol wipes to remove oils, debris and perspiration (check for patient sensitivity or religious objections first). Alternatively soap and water may be used. | Ensure patients own electrical equipment is checked by appropriately trained personnel. This is also often carried out for safety reasons. |
Ask patient to hold breath if possible, alternatively record ECG in a more upright position (remember to document this on the ECG i.e. ECG recorded in sitting position). | If the fault is not caused by other electrical equipment the ECG machine should be tested and evaluated by the relevant department/personnel. |
Calibration
Standard ECGs in the UK, USA and many other parts of the world are recorded at a speed of 25 mm/s and a voltage of 10 mm/mV. This is usually displayed on the ECG somewhere (often at the bottom). This information is also displayed graphically in the form of calibration markers, sometimes called calibration signal boxes. These markers look like rectangles (Fig. 2.2) and are usually seen on the left hand side of the ECG preceding the leads.
Fig. 2.2
A calibration marker
When the ECG is set up to record at the standard 25 mm/s and 10 mm/mV, the calibration markers should measure 1 cm in height (2 large boxes) by 0.5 cm in width (1 large box). The authors recommend that practitioners recording and interpreting ECGs always check that the ECG was recorded in the standard calibration before attempting interpretation. Speed and amplitude settings can be altered on most ECG machines, it is therefore possible that someone may either deliberately or accidentally alter these settings. If the recording speed was altered to 50 mm/s, it would have the effect of elongating the waveforms, and can make it appear that the patient has an extremely low heart rate and a long QT interval, even though their other observations could otherwise be normal. Sometimes altering these settings is done deliberately. A patient may have a very rapid heart rate making it difficult to see P waves. This can sometimes be overcome by changing the recording speed and elongating the waveforms to see features that would otherwise be missed. Figure 2.3 shows a standard calibration marker and one that is set to 50 mm/s. As shown in the image the second marker is twice as wide as normal, encompassing two large boxes.
Fig. 2.3
Calibration markers showing differences in recording speeds. 25 mm/s (left), 50 mm/s (right)
Similar changes can be made to the amplitude of the ECG. Certain conditions, such as left ventricular hypertrophy can cause the height/depth of the waveforms to be very large and they may overlap each other, making the ECG difficult to view. In these cases some practitioners may request the ECG is recorded at half normal voltage. Conversely if a normal ECG is viewed at half voltage the waveforms can appear very small, which can lead to a false interpretation, as it can display false features of conditions like pericardial effusion, which features small voltage deflections. Figure 2.4 shows a calibration marker at half voltage.
Fig. 2.4
Calibration marker showing ½ voltage (5 mm/mV)
Some machines will also allow the voltage to be reduced just for the limb or chest leads, leaving the other leads at normal voltage. This is represented by a calibration marker with a step in it. If the step is on the left it represents a reduction in the limb lead voltage only, whereas a step on the right hand side represents a reduction in the voltage of the chest leads only (Fig. 2.5).
Fig. 2.5
Reduced limb lead voltage only (left), reduced chest lead voltage only (right)
Finally the voltage can be decreased and the recording speed changed simultaneously. This is represented by a small and wide calibration marker (5 mm in height and 10 mm in width). Practitioners should be familiar with both normal and adjusted calibration markers, what they represent and what impact changing these settings may have on the interpretation of the ECG. Even though the calibration markers will reflect any changes it is still good practice to document on the ECG any changes to the normal settings made when recording to draw other practitioners attention to these changes explicitly.
R Wave Progression/Lead aVR
R wave progression refers to the deflection changes that occur in the chest leads (V1–V6) as they move from a predominantly negative to a predominantly positive defection (Fig. 2.6.).
Fig. 2.6
Normal R wave progression in leads V1 to V6
If there are any sudden changes in deflection it is possible that one or more of the chest leads have been placed in the wrong position.
Incorrect lead placement in the limb leads can be identified by a positively deflected lead aVR, especially if the patients previous ECG(s) have a negatively deflected aVR. Lead aVR is nearly always negatively deflected. If it is positive the limb leads may have been applied the wrong way around. This can mimic a condition called dextrocardia, where heart is situated on the right hand side of the chest as oppose to the left. With dextrocardia there is often right axis deviation, positively deflected complexes in lead aVR, negatively deflected complexes in lead I and an absence of R wave progression in the precordial leads. The principle difference between true and technical dextrocardia is that in the later there are no changes seen in the precordial leads.
Basic quality control checks should be carried out by the recorder of the ECG whilst the leads are still in situ. This makes it easier to identify issues such as incorrect lead placement, identification of artifact and any issues with recording settings. Not only is it easier to correct problems at this stage before interpretation is attempted but it also reduces the potential for subsequent interpretation errors.
Determine the Rhythm
The normal heart rhythm is very rarely exactly regular. The human heart rate has a fractal quality. A fractal is a mathematical term to describe patterns that look similar at progressively smaller scales. Other examples from nature include snowflakes and fern leaves. If you were to zoom in on a snowflake again and again the pattern you would see would resemble the larger pattern. An example of this self similarity can be seen in Fig. 2.7. The Sierpinski’s triangle or gasket as it is sometimes referred is a basic example of self similar sets.
Fig. 2.7
Sierpinski’s triangle
The heart rate itself has a fractal variance making it ever so slightly variable. This is extremely important because if this was not the case the stress on the heart muscle each beat would occur at exactly the same point. This slight variance allows wear and tear of the heart to be dramatically reduced.
For pragmatic clinical purposes however rhythms tend to be defined as either regular or irregular. There are a couple of practical ways rhythms can be checked for irregularity. One of the easiest is to take a piece of paper and place it just under the tip of the R waves on the rhythm strip. Next mark a line directly under two consecutive R waves on the piece of paper. Now move the paper along the rhythm strip and check that the two lines on the paper line up with the preceding R waves (Fig. 2.8). Some practitioners do this with a set of calipers (like a compass for drawing circles but with two needles). The calipers are set to mark the distance between two consecutive R waves then swung between the preceding ones (Fig. 2.8).
Fig. 2.8
The paper method (left), caliper method (right)
When checking the rhythm practitioners should consider the following; is the rhythm regular or irregular? if irregular is there a pattern to the irregularity or not? and are PQRST waves present for each beat?
Sinus Arrhythmia
Is a normal variation that produces an irregularity in the heart rate. This is caused by the heart rate increasing as the individual breaths in and slowing down when they breath out. Sinus arrhythmia does not cause any symptoms and PQRST waves are all present. Sinus arrhythmia is discussed in more detail in Chap. 6.
Rate Calculation
The normal adult heart rate is between 60 and 100 BPM. When lying still in bed the normal rate is between 60 and 80 BPM. Normal childrens heart rates can be much higher than adults, for more information on children see Chap. 9, which discusses the pediatric ECG in more detail. With adults anything below 60 BPM is classified as a bradycardia. Anything above 100 BPM is defined as a tachycardia (Table 2.2). Although the ECG can be used to determine the heart rate, a lot of useful information can also be determined from manual palpation of the patients pulse. To feel a pulse slight compression should be made to an artery located against a bone. The common pulse points are shown in Fig. 2.9. Manual inspection of the pulse should not be underestimated. For example; pulseless electrical activity (PEA) is an arrest rhythm where electrical activity is present on the ECG, because the conduction system is working normally, however there is no effective mechanical pumping action taking place and no cardiac output.
Table 2.2
Heart rate classifications
Bradycardia | <60 BPM |
Normocardia | 60–100 BPM |
Tachycardia | >100 BPM |
Fig. 2.9
Palpable pulse points
There are several different methods for determining the heart rate on the ECG. These calculations vary in accuracy and complexity. Some of the methods only work when the heart rate is regular. Several of these methods are described here in the hope that the reader will select the one(s) that are of the most use to their individual circumstances, and that they will develop an awareness of the existence of other methods. The same ECG is used to demonstrate the different methods.
Automated Rate Determination
Most modern ECG machines have an automated computerised set of measurements and interpretation statements. These measures often include the various intervals and durations. They usually also include the heart rate, seen in Fig. 2.10 as Vent rate: 59 BPM. This states the ventricular rate is 59 beats per minute.