Acoustic Behavior of a Reactivated, Commercially Available Ultrasound Contrast Agent




Background


Commercially available microbubbles such as Definity contain octafluoropropane encapsulated in a lipid shell. This perfluorocarbon can be compressed into liquid nanodroplets at room temperatures and activated with transthoracic diagnostic ultrasound. The aim of this study was to determine the size range and acoustic characteristics of Definity nanodroplets (DNDs) compared with Definity microbubbles (DMBs).


Methods


An in vitro flow system was used with a diagnostic ultrasound transducer (S5-1, iE33). DMBs were prepared using package insert instructions. DNDs were prepared by cooling DMBs in a −10°C to −15°C isopropyl alcohol bath before hand-pressurizing the solution. The formed DNDs were sized, diluted to 1% solutions, and infused continuously into a phosphate-buffered saline solution running within Silastic tubing. Acoustic intensity (AI) was compared with equivalent dilutions of DMBs at different mechanical indices (MIs) ranging from 0.2 to 1.4 ( n = 6 comparisons at each MI) using real-time 56-Hz and triggered 2-Hz frame rates (FRs). A 3-cm-thick tissue-mimicking phantom was used to simulate transthoracic attenuation. In vivo transthoracic studies were performed in four normal pigs infused with 10% intravenous infusions of DMBs or DNDs at real-time and triggered end-systolic FRs to compare differences in myocardial and left ventricular cavity AI.


Results


DNDs were smaller than DMBs and ranged in size from 50 to 1,000 nm. In vitro studies revealed that at an MI of 0.2 and an FR of 56 Hz, DMBs had high AI (37 ± 2 dB), but AI dropped to 25 ± 2 dB at an MI of 1.0 ( P < .001, analysis of variance). In comparison, DNDs had virtually no AI at MIs of 0.2 to 0.6 at both triggered and 56-Hz FRs (1 ± 0 dB), but AI increased to 34 ± 2 dB at an MI of 1.4 using an FR of 56 Hz ( P < .001 vs MI of 0.2). AI also persisted longer at 56 Hz with DNDs when using higher MIs. In vivo studies demonstrated higher myocardial AI for DNDs at higher MIs when using real-time FR, most likely from microvascular nanodroplet activation.


Conclusion


These data indicate significant differences in acoustic responses of the commercially available DMBs when administered as an equivalent number of DNDs. The DND formulation may render them more useful for high-MI real-time imaging and other targeted transthoracic diagnostic applications.


Highlights





  • A commercially available lipid-encapsulated perfluoropropane MB can be condensed into submicrometer droplets and reactivated with diagnostic ultrasound.



  • The droplets range in size from 50 to 1,000 nm, and all size ranges can be activated with diagnostic ultrasound at MIs of ≥0.7.



  • Targeted in vivo activation of the droplets produces myocardial contrast in real time at higher MIs.



The use of microbubbles (MB) for contrast enhancement during diagnostic ultrasound imaging has become common over the past few decades. MBs have been used as a contrast agent in a wide variety of medical applications, including solid-organ tumor detection and enhanced cardiac imaging. Current US Food and Drug Administration–approved MBs have also been used for off-label applications such as free intravascular tracers for myocardial perfusion analysis. MBs are typically 1 to 5 μm in diameter and are composed of a gaseous center encapsulated by a protein, lipid, or polymer shell. This size prohibits crossing into extravascular spaces.


Phase-change ultrasound contrast agents are nanometer-sized droplets, or nanodroplets (ND), that also contain fluorocarbons in their liquid form. These fluorocarbons stay in a liquid form well above their boiling points. Their size (50–1,000 nm) results in increased Laplace pressure that prevents boiling and also increases the acoustic energy necessary to vaporize the NDs. Because of this significant limitation, it has been difficult to formulate a phase-change agent that is stable at physiologic temperature but can be activated into stable MBs after vaporization.


Recently, it has been shown that even perfluoropropane (the fluorocarbon gas that is used to formulate commercially available contrast agents such as Definity) can remain in the liquid form at body temperatures, even though its boiling point is −36.7°C. We have been able to condense Definity MBs (DMBs) into Definity NDs (DNDs) using a modification of protocols published by Matsunaga et al. and Sheeran et al. Some investigators have shown the reactivated MBs formed from lipid-encapsulated NDs may have different acoustic responses that affect their destruction rates and stability. A variety of potential explanations exist for the altered acoustic behavior of the reactivated MBs, including altered shell characteristics upon reactivation, larger size distribution, and more resistance to acoustic destruction. We hypothesized that DNDs could be reactivated using a technique observed with other perfluoropropane NDs. Furthermore, we postulated that their behavior in the presence of the ultrasound imaging protocols used for contrast imaging (real time and triggered) would be different from DMBs because of the requirement that they first must be activated before being detected with imaging. In this study, we compared the acoustic responses of DNDs with those of standard DMBs and the potential for specifically activating the <220-nm NDs that may cross defective vascular endothelial barriers and be used for extravascular applications.


Methods


MBs


To create MBs for both imaging and formation of NDs, a vial of Definity was mechanically agitated (Vialmix shaker; Bristol-Myers Squibb, New York, NY) for 45 sec, creating a maximum MB concentration (within the vial) of 1.2 × 10 10 microspheres/mL, with diameters ranging from 1.1 to 3.3 μm according to the package insert. Because the measurement techniques described in the package insert may have failed to measure the number of submicrometer MBs produced, these formulated MBs were sized in our laboratory with analysis systems capable of determining the quantity of submicrometer MBs (see below).


ND Formulation


DNDs were formulated according to a protocol adapted from Matsunaga et al. and Sheeran et al. : 2-mL vials of DMBs were agitated using a Vialmix for 45 sec; a 10% solution of DMBs was created with 0.9% saline as the diluent; the solution was submerged in a 70% isopropanol bath at a temperature between −10°C and −15°C for 3 min; and manual pressure was applied until the solution cleared, indicating condensation. The solution was then warmed to room temperature before administration. A second filtered condensed suspension containing NDs <220 nm was prepared by passing a 10% DND formulation through a 220-nm filter (Merck Millipore, Billerica, MA).


ND and MB Sizing


Because of the polydisperse distribution of the NDs formed following condensation, the DNDs were sized using two different machines: a NanoSight NS300 (Malvern Instruments, Malvern, United Kingdom), which is capable of sizing particles as small as 10 nm in diameter, and a Nano Zetasizer (Malvern Instruments), which can measure particles up to 10 μm in diameter. Initial sizing of the filtered and unfiltered DNDs ( n = 3 each) was done using the Nano Zetasizer to ensure that the larger NDs were being removed from the samples. The filtered DNDs ( n = 6) were also sized using the NanoSight NS300 to improve resolution of NDs with diameters <1 μm and to determine the change in diameter of the filtered DNDs over a 24-hour period at 25°C. DMBs ( n = 3) were analyzed with the Nano Zetasizer to determine the entire range of sizes produced in the Vialmix sample. The sizes produced with manual pressure were compared with those produced with known applied pressures of 6, 9, and 12 atm using a high-pressure inflation device for angioplasty balloon catheters (Merit Medical Systems, Jordan, UT).


In Vitro Experimental Setup


An in vitro flow system was created that has been described previously. This system consists of a pulsatile flow pump (Masterflex; Cole Parmer, Vernon Hills, IL) that propels fluid through a 2-mm-diameter Silastic tube flow system. The solution used for all in vitro studies was phosphate-buffered saline solution running at 20 mL/min at 37°C. An imaging probe (S5-1; Philips Medical Systems, Andover, MA) was placed over the flow system; the probe had a 3-cm tissue-mimicking phantom (Computerized Imaging Reference Systems, Norfolk, VA) placed between it and the tubing to mimic transthoracic attenuation. A port proximal to the imaging chamber allowed either DMBs or DNDs to be infused into the Silastic tubing at dilutions that can be adjusted.


Three different states of Definity ( n = 6 for each state) were acoustically compared: the MBs, the NDs, and the filtered NDs. DMBs and DNDs were both infused at 5.8 × 10 8 MBs or NDs/min, while filtered NDs containing just the <220-nm NDs were infused at 3.9 × 10 10 NDsmin. A higher concentration of filtered NDs was infused to produce equivalent acoustic intensity (AI) as that achieved with DMBs and unfiltered DNDs. Each formulation was continuously infused at a rate of 1.0 mL/min into the proximal port connected to the 2-mm Silastic tubing containing PBS flowing at 20 mL/min. Using real-time imaging (56-Hz frame rate [FR]) and a triggered FR (2.22 Hz), the diagnostic transthoracic ultrasound transducer (1.3-MHz center frequency, 3.4-MHz received frequency, 3-μs pulse duration) was used in ultraharmonic mode to insonate the infusion at incremental mechanical indices (MIs) (0.2, 0.4, 0.6, 0.8, 1.0, 1.2, and 1.45) to (1) determine the threshold required to activate the NDs and (2) determine the AI of the solutions at different regions along the scan plane within the field of insonation ( Figure 1 ). The elevation plane of the transducer was placed in parallel with the Silastic tubing, and AI was measured as the formulations crossed this imaging field. Exposure time to the insonation field at the 20 mL/min flow rate was 5 sec, resulting in cumulative FR exposures of 280 frames at the 56-Hz FR and 11 frames at the 2-Hz FR by the time the MBs or NBs reached region 4 of the imaging plane ( Figure 1 ). Because flow in the tubing was 0.33 mL/sec, and the ultrasound-exposed tubing volume was 1.75 mL, the 500-msec time interval between frames at 2 Hz allowed only partial replenishment of NDs and MBs between frames. AI in the different horizontally placed regions 1 to 4 demonstrated in Figure 1 was analyzed with QLAB software (Philips Medical Systems).




Figure 1


AI of the contrast agents was measured at four different points along the ultrasound imaging space (regions 1–4), with region 1 most proximal to the entry of the contrast agent to the ultrasound field and region 4 most distal to the entry of the contrast agent to the ultrasound field. In this example, the MI is 1.0. (A) DMBs at 2-Hz FR; (B) DNDs at 2-Hz FR; (C) DMBs at 56-Hz FR; (D) DNDs at 56-Hz FR.


In Vivo Experimental Setup


All studies performed were approved by the institutional animal care and use committee at the University of Nebraska Medical Center. Four normal pigs (closed-chest model) with normal left ventricular (LV) systolic function (confirmed by two expert cardiologists) received an intravenous infusion of 10% DMBs while a Philips S5-1 traducer (Philips Medical Systems) was used to perform real-time imaging (35-Hz FR) and 1:1 triggered end-systolic imaging in ultraharmonic mode (1.3-MHz center frequency, 3.4-MHz received frequency) at MIs of 0.2, 0.4, 0.7, 1.0, and 1.2. This same procedure was followed with an intravenous infusion of 10% DNDs.


Background-subtracted AI was compared within the LV cavity, the anteroseptal myocardium, and the anterolateral myocardium at end-systole at each MI and between the two imaging modes ( Figure 2 ). An established version of quantitative AI measurement software (QLAB version 9) was used to compare the AIs in different regions for both in vitro and in vivo measurements. For in vitro studies, the different regions were chosen along a longitudinal plane to assess for activation and destruction ( Figure 1 ). For in vivo studies, different regions were chosen to determine whether different degrees of acoustic activation resulted in different myocardial contrast enhancement within normal zones. The background intensity was defined as the decibels present in each region of interest before contrast administration.




Figure 2


The intensity of the contrast agents was measured in three different areas of the heart: (A) anteroseptal myocardium, (B) LV cavity, and (C) anterolateral myocardium.


To determine whether DNDs could be detected within an infarct zone, two rats (weighing 0.33 and 0.31 kg) underwent surgical left anterior descending coronary artery ligation for 45 min followed by reperfusion. At 72 hours, both rats underwent transthoracic imaging using a Siemens 15L8 transducer (Siemens Acuson Sequoia, 7-MHz fundamental frequency; Siemens Healthcare, Erlangen, Germany) equipped with contrast pulse sequencing (CPS), which transmits pulse sequences of alternating polarity and amplitude for contrast-sensitive imaging. After baseline two-dimensional images of the LV short axis were obtained, the transducer was fixed in one location using a cross clamp attached to a ring stand. Two hundred microliter injections of DNDs were administered during both low-MI (<0.7) and high-MI (1.9) imaging at one frame every four cardiac cycles during and at periodic intervals after the injection, examining for droplet activation within the infarct and normal zones. Imaging was suspended in between each interval to allow droplet accumulation to occur within the infarct zones. Following these injections, the rats were sacrificed, and serial cross sections of the left ventricle were stained with triphenyltetrazolium chloride (TTC) to confirm infarct location.


Statistical Analysis


Analysis of variance was used to compute differences in AI in the different horizontally placed regions in Figure 1 within the flow system at the different FRs and different MIs, while an unpaired Student’s t test was used to compute differences in AI between DMBs and DNDs at the same MI. All data are presented as mean ± SD. Because of the multiple controlled comparisons in the in vitro studies, a P value < .01 was considered to indicate statistical significance for comparisons between AIs produced by equivalent concentrations of DNDs versus DMBs. For in vivo comparisons of NDs with MBs, a P value < .05 was considered to indicate significance. AI comparisons of filtered DNDs with other formulations were not performed, because higher concentrations of the smaller NDs were used in the in vitro studies, and the purpose of using the filtered NDs was to determine whether contrast could be produced from these smaller sized droplets.




Results


ND Sample Sizing


DNDs had a polydisperse distribution, with diameters ranging from approximately 50 to 1,000 nm and the largest concentration in the 250-nm range ( Figure 3 A). There were three peaks, with the first two most likely representing the previously described bimodal distribution of droplet formation with perfluoropropane. The smaller third peak could represent either droplet coalescence or MB formation. This size range was similar to that seen with 6 and 9 atm of pressure using the balloon insufflator (median, 294 nm [range, 175–500 nm] for 6 atm and 261 nm [range, 150–450 nm] for 9 atm). When using 12-atm pressure, median size was 386 nm (range, 250–600 nm). After applying the 220-nm filter, the mean size of the NDs from the six different samples ranged from approximately 94 ± 22 to 150 ± 46 nm ( Figure 3 C). The MBs had a size range of 600 to 5,800 nm ( Figure 3 B), with the largest concentration in the 800- to 1,000-nm range ( Figure 3 B).




Figure 3


Mean diameter ranges of the (A) DNDs, (B) DMBs, and (C) filtered DNDs acquired using the Nano Zetasizer.


In Vitro Studies


During 56-Hz-FR imaging, DMBs exhibited higher AIs at low MIs, which diminished as MI increased ( P < .001). As seen in Figure 4 A, the AI of region 3 was 37 ± 2 dB at an MI of 0.2, which increased to 41 ± 3 dB at an MI of 0.4, following which AI progressively decreased to 23 ± 2 dB when an MI of 1.45 was reached. The AI for DMBs was maximal in regions 1 and 2 at higher MIs ( P < .001). At the 2-Hz FR ( Figure 4 B), AI increased at an MI of 0.4 (compared with 0.2) and then demonstrated little variance as the MI was increased and among different regions of the ultrasound field ( P = .40). There was no significant difference in AI in the different regions at any of the MIs tested when using the 2-Hz FR ( P = .42).




Figure 4


AIs of DMBs in (A) real-time and (B) triggered modes. P < .01 when analysis of variance was used to compare the AIs of the four different regions at each MI.


DNDs produced visually evident contrast only at higher MIs at both the 2- and 56-Hz FRs ( Figures 5 A and 5B). There was a consistent threshold of activation at an MI of 0.8 with both the 2- and 56-Hz FRs. At the 56-Hz FR, regions 2 and 3 had the highest AI at the higher MIs ( P < .001), with region 3 having a significantly higher AI than DMB at MIs of 1.2 and 1.4 ( P = .04 and P < .001, respectively; Figure 5 A). At the triggered 2-Hz FR ( Figure 5 B), the NDs exhibited similar AI to DMBs in all regions when at the highest two MI settings (1.2 and 1.45).




Figure 5


AIs for DNDs in (A) real-time and (B) triggered modes. Acoustic activation of the NDs required higher acoustic pressures (MI approximately 0.8 and higher); however, at these pressures, destruction of the formed MBs also occurs. P < .01 when analysis of variance was used to compare the AIs of the four different regions at each MI.


At the 56-Hz FR, the AI produced by filtered DNDs followed a similar pattern to the unfiltered DNDs, exhibiting negligible AI at low MIs and significantly increased AI at high MIs ( Figure 6 A). Figure 6 A depicts the changes in region 3 for the filtered NDs, which had an AI of 2 ± 1 dB at an MI of 0.2, remained at a negligible value until 0.8, and then sharply increased to 29 ± 5 dB at an MI of 1.45 ( P < .001). This was also seen at the 2-Hz FR ( Figure 6 B).


Apr 15, 2018 | Posted by in CARDIOLOGY | Comments Off on Acoustic Behavior of a Reactivated, Commercially Available Ultrasound Contrast Agent

Full access? Get Clinical Tree

Get Clinical Tree app for offline access